Kangaroo™ Joey Enteral Feeding Pump
(Feed & Hydration Mode)
User Guide
Pump Set Loading Procedure.
|
---|---
1. Locate power button on lower right hand corner of control panel.
Press to turn pump on.| 2. Confirm if you want to KEEP or CLEAR any prior
pump settings.
|
3. With pump display facing you open blue door at top of pump.| 4. Grasp
finger tab on valve and insert valve into valve slot on the left.
|
5. Grasp black ring retainer, and gently wrap tubing anti-clockwise around
rotor. Pull up retainer and insert into slot.| 6. Close blue door. Display
shoul read SET LOADEO.
Set Priming and Pump Programming Procedure.
|
---|---
1. Press PRIME PUMP followed by AUTO PRIME to automatically prime
or HOLD TO PRIME to manually prime both feed and flush lines.
Press DONE when complete.| 2. Press ADJUST FEED followed by FEED RATE to
adjust hourly
delivery rate.
|
3. Press buttons on left to program hourly feed rate. Press ENTER to confirm
rate.| 4. Press FEED VTBD to program feed Volume To Be Delivered if desired.
|
5. Press buttons on left to program volume of feed to be delivered. Press
ENTER to confirm rate, followed by DONE to return to main screen.| 6. Press
ADJUST FLUSH to program flush volume and interval.
|
7. Press FLUSH VOLUME to program flush volume. Press FLUSH INTERVAL to
program interval between each flush.| 8. Press buttons on left to program
volume of flush to be delivered. Press ENTER to confirm volume.
|
9. Press buttons on left to program flush interval. Press ENTER to confirm
interval, followed by DONE to return to main screen.| 10. Press RUN to
commence feeding regime. In order to achieve proper accuracy, the initial
fluid line in the feeding set container must be 15 cm above the top of the
feeding pump when initiating the feeding cycle.
* IMPORTANT NOTE. Kangaroo™ Joey Pump will automatically flush an additional 25mL of water after completing the programmed delivery of formula, as long as at least one flush has occured during formula delivery.
POWER DOWN PROCEDURE Locate power button on lower right hand corner of control pane. Press and hold power button until the pump display turns off.
Alarm and Error Code Information.
| |
---|---|---
HOLD ERROR
Pump left in Hold for more than 10 minutes.
Press CONTINUE followed by RUN to start pump running or power down pump.|
ROTOR ERROR
Problem with pump set tubing around the rotor. Ensure silicone tubing is
correctly loaded
around rotor.| FLOW ERROR
Obstruction in line downstream from pump.
Check for kinked or blocked tubing below pump and rectify.
| |
FEED ERROR
Feed container empty or obstruction in the line
upstream from pump.
Check for kinked tubing or obstruction in line above pump or attach
new/refilled feed container.| FLUSH ERROR
Water container empty or obstruction in the line upstream from pump. Check for
kinked tubing or obstruction in flush line above pump or refill flush
container.| FEEDING COMPLETE
Programmed feed has been delivered.
Power down pump.
| |
PUMP SET DISLODGED
Black retainer ring (MISTIC) is not properly loaded in
MISTIC pocket. Ensure pump set is correctly loaded.| BATTERY LOW
Battery must be recharged immediately.
Plug power cord into AC outlet.| SYSTEM ERROR
General error – see pump manual.
Send pump for service.
For complete operation information please refer to Kangaroo™
Joey enteral feeding pump operation and service manual cardinalhealth.com.au
Cardinal Health Australia
Level 2, 5 Eden Park Drive, North Ryde NSW 2113
© 2019 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal
Health LOGO and ESSENTIAL TO CARE are trademarks of Cardinal Health and may be
registered in the US and/or in other countries.
All other marks are the property of their respective owners. 1432-03-19
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>