PENTAIR F3000-B2B FreshPoint 3-Stage Under Counter Water Filter System Instruction Manual
- August 16, 2024
- Pentair
Table of Contents
- PENTAIR F3000-B2B FreshPoint 3-Stage Under Counter Water Filter System
- SPECIFICATIONS
- PARTS INCLUDED
- PRECAUTIONS
- TOOLS AND MATERIALS REQUIRED
- INSTALLATION
- FILTER CARTRIDGE REPLACEMENT
- TROUBLESHOOTING
- CARTRIDGE TIMER PROCEDURE
- REPLACEMENT PARTS
- PERFORMANCE DATA
- CARTRIDGE SEQUENCE
- FAQ
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
PENTAIR F3000-B2B FreshPoint 3-Stage Under Counter Water Filter System
Specifications
- Dimensions: 317mm x 285mm x 133mm
- Weight:
- Model F3000-B2B : 6 lbs (2 kg)
- Model F3000-B2M : 6 lbs (2 kg)
SPECIFICATIONS
- Temperature Range :……………………………..40-100°F (4.4-37.8°C)
- Pressure Range :……………………………..40-100 psi (2.75-6.89 bar)
- Service Flow Rate @ 60 psi (4.1 bar):
- Model: F3000-B2B………………………………….0.6 gpm (2.27 L/min)
- Model: F3000-B2M………………………………….0.6 gpm (2.27 L/min)
- Rated Service Life:
- Model: F3000-B2B…………………………………..675 gallons (2555 L)
- Model: F3000-B2M…………………………………..675 gallons (2555 L)
- Dimensions :…………………………………………. 12.5″ x 11.25″ x 5.25″ (317mm x 285mm x 133mm)
- Weight:
- Model: F3000-B2B (system only)…………………………….6 lbs (2 kg)
- Model: F3000-B2M (system only)……………………………6 lbs (2 kg)
PARTS INCLUDED
- Filter System with Cartridges
- Installation Hardware Kit
- Lead-free drinking water faucet
- 3/8″ (white and blue) plastic tubing
PRECAUTIONS
WARNING : Do not use water that is microbiologically unsafe or of unknown
quality without adequate disinfection before or after the system. Systems
certified for cyst reduction may be used on disinfected waters that may
contain filterable cysts. Chemical and mechanical reduction unit.
CAUTION The System must be protected against freezing, which can cause
cracking of the filter and water leakage.
NOTE:
- For cold water use only.
- Make certain that installation complies with all state and local laws and regulations.
- The system must be maintained according to the manufacturer’s recommendations, including the replacement of filter cartridges. The contaminants or other substances removed or reduced by the selected cartridge are not necessarily in your water. Ask your local water municipality for a copy of their water analysis, or have your water tested by a reputable water testing lab.
- After prolonged periods of non-use (such as during a vacation), it is recommended that the system be flushed thoroughly. Let water run for 5–6 minutes before using.
- The filter cartridges used with this system have a limited service life. Changes in taste, odor, and/or flow of the water being filtered indicate the cartridge should be replaced.
CALIFORNIA PROPOSITION 65 WARNING
WARNING: This product contains chemicals known to the State of California to cause cancer birth defects or other reproductive harm.
TOOLS AND MATERIALS REQUIRED
-
Safety glasses
-
Adjustable wrench
-
Tube cutter or utility knife
-
Hand or electric drill (cordless recommended)
-
File
-
1/4″ drill bit
If the sink does not have a hole for a separate faucet: -
Center punch
-
3/4″ drill bit or hole saw
-
3/32″ drill bit
INSTALLATION
NOTE:
- For standard under-sink installation on 1/2″ 14 NPS threads or 3/8″ x 3/8″ compression
- Please read all instructions and precautions before installing and using your Filtration System
- Numbered diagrams correspond with numbered steps
Installing the Water Supply Adapter
The supply adapter fits 1/2″-14 NPS supply threads or
3/8″ x 3/8″ compression. If local codes permit, it may be used to connect the
system to the cold water supply line. If local codes do not permit the use of
the supplied adapter, alternate connectors can be obtained from your local
supplier.
- A. Turn off the cold water supply line. If the cold water line does not have a shut-off valve under the sink, you should install one.
- B. Turn on the cold water faucet and allow all water to drain from the line.
- C. Disconnect the riser cold water supply valve.
- D. Ensure the sealing gasket is fully seated into the feed adapter valve female thread.
- E. Install the feed adapter valve onto the supply valve. Hand tighten only.
- F. Connect the riser to the feed adapter valve.
NOTE: See Figures 1G-1H for configuring the feed adapter to 3/8″ x 3/8″ compression connections.
Selecting the Faucet Location
- NOTE: The drinking water faucet should be positioned with function, convenience, and appearance in mind. An adequate flat area is required to allow the faucet base to rest securely. The faucet fits through a 3/4″ hole. Most sinks have pre-drilled 1-3⁄8″ or 1-1⁄2″ diameter holes that may be used for faucet installation. If these pre-drilled holes cannot be used or are in an inconvenient location, it will be necessary to drill a 3/4″ hole in the sink to accommodate the faucet.
- CAUTION This procedure may generate dust which can cause severe irritation if inhaled or comes in contact with the eyes. The use of safety glasses and respirators for this procedure is recommended.
- CAUTION DO NOT ATTEMPT TO DRILL THROUGH AN ALL-PORCELAIN SINK. If you have an all-porcelain sink, mount the faucet in a pre-drilled sprayer hole or drill through the countertop next to the sink.
- CAUTION When drilling through a countertop, make sure the area below the drilled area is free of wiring and piping. Make certain you have ample room to make the proper connections to the bottom of the faucet.
- CAUTION Do not drill through a countertop that is more than 1″ thick.
- CAUTION Do not attempt to drill through a tiled, marble, granite, or similar countertop. Consult a plumber or the countertop manufacturer for advice or assistance.
- A. Line the bottom of the sink with newspaper to prevent metal shavings, parts, or tools from falling down the drain.
- B. Place masking tape over the area to be drilled to prevent scratches if the drill bit slips.
- C. Mark the hole with a center punch. Use a 1/4″ drill bit to drill a pilot hole.
- D. Use a 3/4″ drill bit and drill a hole completely through the sink.
- E. Smooth rough edges with a file.
Mounting the Faucet
-
A. Apply 3-5 wraps of plumber tape to the faucet stem.
-
B. Assemble the faucet as shown in Figure 3.
-
C. Screw the quick connector onto the threaded end of the faucet stem.
-
D. Place a mark 5/8″ from one end of the blue tubing. Wet the end of the 3/8″ tube and push into the bottom of the connector. The 5/8″ mark should be flush with the collar of the connector.
NOTE: To remove the tube, push on the fittings’ collar and pull the tube out. -
E. Holding the faucet, feed the tubing through the hole in the sink. Position the faucet handle at the desired location.
-
F. Center the faucet slip a slotted disc between the bottom of the counter or sink and lock the washer. Tighten the stem nut with a wrench until it is tight.
-
G. Firmly insert the spout into the faucet base.
Mounting the System
-
A. Select a location under the sink, or other suitable area where the system will be installed.
NOTE: The system carton can be used to determine the operational footprint required for installation of the system and replacement cartridges. (Figure 4a) -
B. Mount the system vertically. Ensure mounting is level.
Place a pencil mark in the upper slots of the bracket mounting holes. (Figure 4b). Alternate mounting option: If mounting the system near the floor, place the pencil marks above each of the mounting holes on the bracket to ensure enough space resides below the cartridge and the floor for cartridge replacement (Figure 4c). Use the 3/32″ drill bit to create pilot holes for mounting.
WARNING: The system should be mounted to a firm, solid surface that can support the weight of the system.
Connecting the Faucet to the System
-
A. Determine the length of plastic blue tubing needed to connect to the outlet (right) side of the filter from the faucet. Be sure to allow enough tubing to prevent kinking and cut the tubing squarely. Use a marker to mark one end of the tubing 5/8″ from the end (Figure 5). Wet the end of the 3/8″ tube and push into the outlet (right) connection of the system up to the mark.
CAUTION Do not bend or crimp tube inserting. -
B. Gently pull back on the tube to ensure it is connected properly.
Connecting the Supply Adapter
- A. Determine the length of plastic white tubing needed to connect the inlet (left) side of the filter with the feed adapter installed on the cold water supply line. Be sure to allow enough tubing to prevent kinking and cut the tubing squarely. Place a mark 5/8″ from one end of the tubing. Wet the end of the tubing. Insert the end of the tubing into the 3/8″ fitting of the inlet supply adapter. The 5/8″ mark should be flush with the collar of the fitting located on the inlet supply adapter (Figure 6a).
- B. Mark the remaining end of the tubing. Use a marker to mark 5/8″ from the end. Wet the end of the tube. Insert the end into the feed connection on the inlet side (left) of the system. The tube should go up to the mark. (Figure 6b).
Placing the System into Operation
- A. Slowly turn on the cold water supply. Ensure that the feed adapter valve is also open.
- B. Open the faucet and flush the filter for 10 minutes. Discard water.
- C. Check for leaks at all fittings before leaving installation. If leaks appear, see Troubleshooting.
NOTE: A drinking water cartridge may contain carbon fines (very fine
black powder). If carbon fines appear in the water, discard the water. More
flushing is required if carbon fines are present.
NOTE: Initially, filtered water may appear cloudy. If cloudiness in a
glass of water disappears from the bottom; fine air bubbles are present. This
air within the water will disappear within a few weeks after installation.
FILTER CARTRIDGE REPLACEMENT
NOTE: The life of the filter cartridges depends on the water volume used and the quality of the feed water. It is recommended that the filter cartridges be replaced every 6-12 months, or when there is a noticeable change in taste, odor, or flow of filtered water.
Ensure the correct cartridges are purchased for the system.
Model F3000-B2B and F3000-B2M use the F2B2-RC2 Replacement Cartridge set and
F1S5-RC Cartridge.
Cartridge Replacement
- A. Relieve pressure by turning off the water supply to the system and opening a faucet until the water flow stops. Place a bucket or towel under the system to catch any water drips (Figure 8a).
- B. Remove the cover and lift the locking bar upward until the filter cartridge disengages from the filter head assembly (Figure 8b).
- C. Pull the cartridge away from the filter head assembly (Figure 8c).
- D. Ensure the locking bar is fully in the up position (Figure 8c).
- E. Align the posts on the filter cartridge with the ports in the filter head assembly. Slide the cartridge filter towards into the filter head assembly engaging with the locking bar causing it to drop forward and down. (Figure 8d).
- F. Pull down the locking bar until it snaps into place (Figure 8e).
- G. Install the cover.
NOTE: See the Placing the System into Operation section for the remaining steps.
TROUBLESHOOTING
Leaks between filter head assembly and filter cartridge
- Relieve pressure by turning off the water supply to the system and opening the faucet until the water flow stops. Place a bucket or towel under the system to catch any water drips.
- Remove the cartridge and inspect the O-rings to make sure they are seated and clean.
- Install filter cartridge. Place the system into operation and check for leaks. If leaks persist, turn off the water supply and contact Technical Support at 1-800-279-9404.
Leaks from tubing fittings
- Relieve pressure by turning off the water supply to the system and opening the faucet until the water flow stops. Place a bucket or towel under the system to catch any water drips.
- Depress the collet on the system or inlet supply adapter tubing fittings and pull the tubing from the fitting. Inspect the surface of the tubing for scratches or debris. Clean or cut back tubing to access clean surface.
- Wet the end of the inlet tubing and press into the inlet fitting of the system. Ensure the tubing is fully pushed past the fitting O-rings. Place the system into operation and check for leaks. If leaks persist, turn off the water supply and contact Technical Support at 1-800-279-9404.
CARTRIDGE TIMER PROCEDURE
FILTER CARTRIDGE TIMER (MONITORED SYSTEMS ONLY)
The filter cartridge timer can be installed to the systems decorative cover by
peeling away the adhesive backing on the metal plate included with the
cartridge timer. The timer can also be attached to a metal magnetic surface
using the magnet that is factory-installed on the cartridge timer.
INSTALLATION AND ACTIVATION
Once the installation location has been selected, activate the timer by
pulling the plastic tab out from the side of the timer. Press and release the
key to verify the battery is operational. The light will blink green three
times. (Figure 9a)
NOTE : The timer operates on 12-month schedule. The timer uses a coin
cell type 2023 battery.
The Filter Cartridge Timer is Operational
Operation
- The timer will begin to blink red once every three minutes after 11 months and will blink red three times every three minutes after 12 months. (Figure 9b)
NOTE: The timer status of the cartridge life can also be viewed immediately by pressing and releasing the key.
Reset TimerReset Timer
- To reset the timer after filter cartridge replacement, press the key and hold for 5 seconds. The timer is now reset to 12 months. (Figure 9c)
Battery Replacement
- Replace the battery every 12 months. To replace the battery, locate the slot on the side of the timer body and carefully remove the front of the timer. The battery is now accessible. (Figure 9d)
- Slide the new battery under the battery retention bracket with the positive + side of the battery facing up ttowardthe battery retention bracket. Align the tab on the timer back to the slot on the timer front and press and snap both timer halves together. Battery replacement will not reset the 12-month timer. If reset is required, press and hold the key for 5 seconds.
REPLACEMENT PARTS
ITEM NUMBER | PART NUMBER | DESCRIPTION | QUANTITY REQUIRED |
---|---|---|---|
1 | 655126-96 | Cartridge Set, F2B2-RC2 | 1 |
2 | 655116-96 | Cartridge, F1S5-RC | 1 |
3
| 4004604| Filter Head, Model F3000-B2B| 1
4004606| Filter Head, Model F3000-B2M| 1
4| 4003596| Installation Kit 2 & 3 Stage| 1
5| 244960*| Faucet| 1
6| 4004917| Cover, 3 Stage| 1
7| 4003641| Cartridge Timer| 1
The image may be different than the actual faucet
For replacement parts, contact your nearest Pentair distributor or call 800.279.9404
PERFORMANCE DATA
Important Notice : Read performance data and compare the capabilities of the system with your actual water treatment requirements. It is recommended that the supply water be tested, before installing a water treatment system, to determine your water treatment needs.
Test Conditions
Flow Rate
- F3000-B2B: = 0.60 gpm (2.27 L/min)
- F3000-B2M : = 0.60 gpm (2.27 L/min)
Filter Capacity
- F3000-B2B = 675 gallons (2555 L)
- F3000-B2M = 675 gallons (2555 L)
- Inlet Pressure = 60 psi (4.1 bar)
- Temperature = 68°F +/- 5°F (20°C+/- 2.5°F)
Testing was performed under standard laboratory conditions, actual performance may vary.
Operating Requirements for F3000-B2B and F3000-B2M Systems:
- Pressure = 40-100 psi (2.75-6.89 bar)
- Temperature = 40-100°F (4.4–37.8°C)
CARTRIDGE SEQUENCE
F3000-B2B, F3000-B2M SYSTEMS INSTALLED WITH F2B2-RC2 FILTER CARTRIDGE SET AND F1S5-RC CARTRIDGE
Substance
| Influent Challenge Concentration| Max. Permissible Product
Water Concentration| Reduction Requirements| Minimum
Reduction| Average Reduction
---|---|---|---|---|---
Standard 42
Chlorine Taste & Odor| 2.0 mg/L±10%| | ≥50%| | 95.9%
Particulates (0.5-
<1uM) Class 1*
| at least 10,000 particulates/mL| | >85%| | 97.9%
Standard 53
Cysts | Minimum 50,000/L| | 99.95%| 99.97%| 99.99%
Atrazine| 0.009 mg/L ± 10%| 0.003 mg/L| | 90.5%| 93.7%
Lead (pH 6.5)| 0.15 mg/L ± 10%| 0.010 mg/L| | 99.3%| 99.9%
Lead (pH 8.5)| 0.15 mg/L ± 10%| 0.010 mg/L| | 99.3%| 99.6%
Lindane| 0.002 mg/L ± 10%| 0.0002 mg/L| | 94.8%| 97.4%
Chloroform (VOC surrogate chemical)| **
0.300mg/L ± 10%
| ****
0.015 mg/L
| | ****
96.5%
| ****
98.8%
Perfluorooctanoic Acid (PFOA) and Perfluorooctane Sulfonate (PFOS)
| ****
0.0015 mg/L ± 10%
| ****
0.00007 mg/L
| | ****
96.27%
| ****
98.06%
- Flow Rate = 0.6 gpm (2.2 L/min); Capacity = 675 gallons (2555 L) or 12 months
- Testing was performed under standard laboratory conditions, actual performance may vary.
- Reduces particles as small as 0.5-1 micron in size by mechanical means
- NSF/ANSI Standard 53 is certified to reduce cysts such as Cryptosporidium and Giardia by mechanical means.
EPA Est. No. 082989-CHN-001
- The Models F3000-B2B and F3000-B2M are Tested and Certified by NSF International against NSF/ANSI Standard 42, 53, and CSA B483.1 for the reduction of substances specified on the Performance Data Sheet.
- The Models F3000-B2B and F3000-B2M are certified by IAPMO R&T against NSF/ANSI 53 for the reduction of perfluorooctanoic acid (PFOA) and perfluorooctane sulfonate (PFOS).
F3000-B2B, F3000-B2M Filtration Systems Instruction Manual
- For Pentair Product Warranties visit:
- Pentair para las garantías de los productos visite: Pour Pentair garanties produit visitez le site :
13845 Bishops Dr. | Suite 200 | Brookfield, WI 53005 | United States
P:262.238.4400 | Customer Service : 800.279.9404 | tech-
support@pentair.com |
pentair.com
All indicated Pentair trademarks and logos are property of Pentair. Third-
party registered trademarks and logos are the property of their respective
owners.
©2023 Pentair. All Rights Reserved.
4003611 Rev F JE23
FAQ
- Q: Can I use this filtration system with unfiltered water?
- A: No, do not use the system with microbiologically unsafe water. Always ensure water quality before use.
- Q: How often should I replace the filters?
- A: Filter replacement frequency may vary based on usage and water quality. Refer to the user manual for specific guidance on filter replacement intervals.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>