YIMI LIfe YM201 0.96 Inch Finger Clip Pulse Oximeter User Manual
- June 6, 2024
- YIMI LIfe
Table of Contents
YIMI LIfe YM201 0.96 Inch Finger Clip Pulse Oximeter User Manual
Product Introduction and Operation Guide
Front View
- Figure 1 Front View of YM201/YM301
Operation Method
- Open the battery cover, and put the two AAA batteries into the battery compartment in correct polarities, then replace the cover;
- Press the bottom of the equipment and open the probe, then insert one finger into the probe;
- Press the button to turn the equipment on, and the measure interface will appear;
- After about 8 seconds, the measurement result can be read directly from the display screen;
- Before reading the parameters, make sure that stable numbers of the pulse oximeter interface has sustained more than 4 second;
- The equipment will turned off automatically within 8 seconds when the finger left the probe.
Battery Installation
- Put the two AAA batteries into battery compartment in correct polarities (Figure2).
- Push the batter y cover horizontally along the arrow shown as right.
WARNINGS:
- Battery polarities should be correctly installed, otherwise, damage may be caused to the equipment.
- Please remove the batteries if the equipment will not use for a long time.
Figure 2 Battery Installation
Lanyard installation
- Pass the thinner end of the lanyard through the hanging hole;
- Pass the thicker end of the lanyard through the thinner end and tighten the lanyard(Figure3).
Lanyard Installation
Attention for Operation
- Before use check and confirm that the people or finger size were applicable;
- Before use check and confirm that the environment should be non- combustible material, as well as to avoid high or low temperature and humidity, but also need to pay attention to the following:
- To avoid glare and direct sunlight exposure;
- To avoid radiation infrared or ultraviolet radiation;
- Avoid contact with the organic solvent, mist, dust, corrosive gases;
- The equipment should not be used at a location or limb tied with arterial canal or blood pressure cuff or receiving intravenous injection;
- The equipment may not work normally on microcirculation barrier patients, Warm or rub the finger, or re-position the equipment could improve the measurement.
- The ray between photo detector and light emitting diode should across patient’s arteriole.
- The patient should not use enamel or other makeup;
- Avoid to insert a wet finger into the probe.
Notes:
- The user should fully insert the finger into the probe;
- It is recommended to let the LED light shine directly on the nail(Figure 4);
- Don’t shake the finger and try to keep still during the measurement.
Finger Placement Diagram
Function and menu operation
After turning on the oximeter, press and hold the power button for about 2 seconds. The oximeter will call up the parameter setting interface and set it by pressing the button. Defined here, long- press indicates that the button hold time reaches 1-2s, short-press indicates that the button hold time is less than 0.5s.
On parameter interface 1
-
Move”*” to the corresponding option, and hold the button to set Alm or Beep to
on or off. -
When Alm is set to on and the measured SpO2 or PR Values go beyond the upper
limit or lower limit, the oximeter gives off an alert sound. -
When Alm is set to off and the measured values go beyond the limit, the Oximeter
will not give any alert sound. -
When Beep is set to on, a ticking sound synchronized with the pulse is emitted during the measurement, and when Beep is set to off, no sound is output.
-
While the “*” symbol stays on the Restore option, hold the button to restore
factory settings. -
Press the button to select a Brightness level ranging from 1 to 5. The greater the value, the greater the brightness of the screen.
- Interface 1
- Settings
- Alm setup *
- Alm on
- Beep off
- Demo off
- Restore ok
- Brightness 1
- Exit
On parameter interface 2
- Press the button to switch between options. On this interface, you can set the upper limit and lower limit of SpO2 Alm and PR Alm.
- While the “*” symbol stays on the +/- option, hold the button to set the option to + or -. In + mode, select the corresponding option and hold the button to increment the upper or lower limit; in – mode, hold the button to decrement the upper or lower limit.
- Move “*” to the Exit option, and hold the button to return to the monitoring interface.
- Interface 2
- Settings
- Sounds setup *
- SpO2 Alm Hi 100
- SpO2 Alm Lo 94
- PR Alm Hi 120
- PR Alm Lo 50
- +/- +
- Exit
Specification
Classification
Type of protection against electric shock….Ⅱ (Internally powered equipment)
Degree of protection against electric shock………………….Type BF-Applied part
Operating mode ……………………………………………………………….Spot checking
Degree of protection against hazards of explosion………………………………….P22
Power Requirements
Specification of alkaline batteries…………………………………….Two AAA (LR03)
Operating current……………………………………………………………………….25-50mA
Physical Specifications
WidthHeightDepth……………………………………………………….57×30×31 mm
Weight………………………………………………………………………..28g (Bare machine)
Measurement Specifications
Spo2 declared accuracy …………………………………………..70%~100%: ±2digits 0% ~ 69%:
unspecified
SpO2 Display Range…………………………………………………………………30%~99%
SpO2 Resolution…………………………………………………………………………………1%
PR declared accuracy ………………………………………………25-250bpm: ±3digits
PR Resolution………………………………………………………………………………….1bpm
Environmental Specifications Temperature
-
Temperature
Operating…………………………………………………+50~+104°F / +10~ +40°C
Storage/Transportation…………………………………..-4~+140°F / -20~+60°C -
Humidity
Operating………………………………………………………..15~95%, noncondensing
Storage/Transportation……………………………………..10~95%, noncondensing -
Atmosphere Pressure
Operating……………………………………………………………………………70~106kpa
Storage/Transportation………………………………………………………50~107.4kpa
Display
Display Color …………………………………………………………………….. YM201: 0.96″,Yellow&Blue
YM301: 1.3″,Blue
Display content: SpO2%, Pulse Rate, PI%, Bar Graph Battery Indicator, Pulse
Wave.
Notes:
-
- The claim for oxygen saturation accuracy should be supported by clinical studies covering the entire claimed range, The fraction of inspired oxygen (FiO2) delivered to test subjects is varied to achieve a series of targeted steady-state saturation periods over the specified SpO2 accuracy range (e.g. 70 % to 100 % ),then the SpO2 accuracy is calculated by comparing SpO2 readings of the pulse oximeter to the values of SpO2 determined with a Co- Oximeter.
- The clinical trial included 11 subjects, including 6 males and 5 females, with an age range of 18 to 46 years, the subjects skin color included dark black, medium black, light color and white.
Maintenance, Cleaning, Disinfection
Maintenance
The equipment’s design life expectancy is about 2 years, keep your equipment and accessories free of dust and dirt, and follow these rules:
-
Please clean the equipment before use according to chapter 6.2; Remove the batteries inside the battery cassette if the equipment will not be operated for a long time;
-
Replace the batteries in time when the battery voltage indicate lamps were empty;
-
It is recommended that the equipment should be kept in a dry environment with no
corrosive gases and good ventilation anytime. The moisture and high-light environments will affect its lifetime and even might damage the equipment. -
It is best to preserve the product in a place where the temperature is between -20 to 60℃ and the relative humidity is less than 95%. E. The packed equipment can be transported by ordinary conveyance. The equipment not be transported mixed with toxic, harmful, corrosive materials.
WARNINGS:
- No modification of this equipment is allowed.
Cleaning
Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place, the equipment should be cleaned more frequently. Before cleaning the equipment, consult your hospital’s regulations for cleaning the equipment. Recommended cleaning agents are:
- Mild soap (diluted).
- Ethanol (70%).
To clean your equipment, follow these rules:
- Shut down the pulse oximeter;
- Clean the display screen using a soft, clean cloth dampened with a glass cleaner;
- Clean the exterior surface of the equipment and probe using a soft cloth dampened with the cleaner;
- Wipe off all the cleaning solution with a dry cloth after cleaning if necessary;
- Dry your equipment in a ventilated, cool place. To avoid damage to the equipment, follow these rules:
CAUTIONS:
- Always dilute according the manufacturer’s instructions or use lowest possible concentration.
- Do not immerse part of the equipment in the liquid.
- Do not pour liquid onto the equipment or accessories.
- Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or acetone-based cleaners).
- If you spill liquid onto the equipment, contact us or your service personnel.
Disinfection
Clean the pulse oximeter before disinfecting it. The recommend disinfectant is ethanol 70%. Disinfection step are the same as cleaning.
CAUTION
- Never use ETO or formaldehyde for disinfection.
Disposal
Dispose of the pulse oximeter in accordance with local environment and waste disposal laws and regulations.
Accessories
- One lanyard.
- Two AAA batteries(Optional).
- One user manual.
- One certificate card.
Troubleshooting
Trouble
| Possible Reason|
Solution
---|---|---
The equipment can’t be turned on.| The battery is drained away or almost
drained away.| Please replace batteries.
The battery installation is incorrect.| Install the battery over again.
The device works abnormally.| Please contact the product distributor.
The Spo2 and PR are not displayed normally| The finger size is too big or
small.| Select the suitable size finger to measure.
Excessive ambient light.| Aviod the excessive ambient light irradiation.
User’s blood perfusion is very low.| Warm the finger and try again.
The display is off suddenly.| The equipment is set to shut down automatically
in 8 seconds when there is no correct physiological Signals.| Normal.
The battery is almost drained away.| Replace batteries.
The Spo2 and Pulse Rate are not displayed stably.| The finger is not inserted
deep enough.| Replace the finger and try again.
The finger is shaking or the body is moving.| Try to keep still.
Not used in the work environment required by this manual.| Please use in
normal working environment.
The device works abnormally.| Please contact the product distributor.
Appendix A EMC
The equipment complies with the requirement of standard EN 60601-1-2:2014
- “Electromagnetic Compatibility – Medical Electrical Equipment”.
1| Guidance and manufacturer´s declaration – electromagnetic emission
---|---
2| The model YM201/YM301 is intended for use in the electromagnetic environment specified below. The customer or the user of the model YM201/YM301 should assure that it is used in such an environment.
3| Emissions test| Compliance| Electromagnetic environment – guidance
4| RF emissions CISPR 11| Group 1| The Model YM201/YM301 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
5| RF emissions CISPR 11| Class B| The Model YM201/YM301 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
6| Harmonic emissions IEC 61000-3-2| Not applicable
7| Voltage fluctuations / flicker emissions IEC 61000-3-3| Not applicable
Guidance and manufacturer´s declaration – electromagnetic immunity
The Model YM201/YM301 are intended for use in the electromagnetic environment specified below. The customer or the user of the Model YM201/YM301 should assure that it is used in such an environment.
Immunity test
| IEC 60601 test level| Compliance level|
Electromagnetic environment – guidance
| |
± 8 kV contact
| Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
| ± 8 kV contact|
Electrostatic discharge (ESD) IEC 61000-4-2| ±2 kV, ±4 kV, ±8 kV,
±15 kV air
| ±2 kV, ±4 kV,
±8 kV, ±15 kV
air
Electrostatic transient / burst IEC 61000-4-4| ± 2 kV for power supply lines
100 kHz repetition frequency ± 1 kV for input/output lines| N/A| N/A
| ± 0.5 kV,| ±| 1| kV| |
Surge| | | | | |
| differential| | | | N/A| N/A
IEC 61000-4-5| | | | | |
| mode line-line| | | | |
Voltage| 0 % UT (100 % dip in UT )| |
dips, short| for 0.5 cycle at 0°, 45°, 90°,| |
interruptions and| 135°,180°, 225°, 270°, and| N/A| N/A
voltage variations| 315°| |
on power supply| | |
input lines
IEC 61000-4-11| 0 % UT (100 % dip in UT )
for 1 cycle at 0°
70 % UT (30 % dip in UT )
for 25/30 cycles at 0°
0 % UT (100 % dip in UT )
for 250/300 cycle at 0°
| |
---|---|---|---
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8| 30 A/m, 50/60Hz| 30A/m,
50/60Hz
| Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
NOTE: UT is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The YM201/YM301 is intended for use in the electromagnetic environment specified below. The customer or the user of the YM201/YM301 should assure that it is used in such an environment.
Immunity test
| IEC 60601 test level| Compliance level|
Electromagnetic environment – guidance
Conducted| 3 Vrms 150| N/A| Portable| and| mobile| RF| communications
RF IEC 61000-4-6| kHz to 80 MHz 6 Vrms 150 kHz to 80 MHz outside
ISM bandsa| | equipment should be used no closer to any part of the Models
YM201/YM301, including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
---|---|---|---
Radiated RF| 10 V/m| | Recommended separation distance
80MHz to 800MHz
800MHz to 2.7GHz
where P is the maximum output power ratin the transmitter in watts (W) according to transmitter manufacturer and d is recommended separation distance in metres (m Field strengths from fixed RF transmitters determined by an electromagnetic site surve should be less than the compliance level in frequency range b
Interference may occur in the vicinity equipment marked with the following symbol:
| | 10 V/m
IEC| 80 MHz to|
61000-4-3| 2.7 GHz|
| | |
---|---|---|---
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
1. The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
2. The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,7 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges.
3. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the YM201is used exceeds the applicable RF compliance level above, the YM201should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the YM201/YM301.
4. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
MANUFACTURER
Shenzhen Yimi Life Technology Co.,Ltd
Add: 305, Building A, Tengbo Industrial Park, Changshangjiang street,
Longbei Village, Pingshan District,518118,Shenzhen, China
Tel: +86 755-86573112
Web: www.yimilife.com
EC REPRESENTATIVE
Shanghai International Holding corp. Gmbh(Europe)
Address: Eiffestrasse 80, 20537, Hamburg, Germany
Email: shholding@hotmail.com
Certificates
PN: Finger Pulse Oximeter
Date:_____