Premiere XC-2000 Centrifuge with Timer and Speed Control User Manual
- June 6, 2024
- Premiere
Table of Contents
Centrifuge
Model XC-2000
Operating Manual
Before using this product, please read this manual carefully to learn about
the available features to obtain the best results of this equipment.
Warning: To prevent fire or shock hazard, do not expose this unit to rain
or moisture. Unless labeled otherwise, this equipment should be used with AC
110V 60Hz in the USA or Canada.
220V 50Hz available as Model No. XC-2000-220
Caution: No user-serviceable parts inside. Refer servicing to qualified
service personnel
Safety
Before using the centrifuge, please read this instruction manual carefully. Incorrect or inappropriate use may result in equipment damage or personal injury!
This centrifuge is designed for clinical medicine, biology, genetic engineering, immunology, etc. The density of the separated sample cannot exceed 1.2g/cm³ under the maximum speed, when the density of the sample exceeds 1.2g/cm³, the maximum speed of the rotor must be reduced accordingly.
When the centrifuge is running (rotor is spinning), Please keep an area of
about 12 inches around the centrifuge clear while operating it and ensure
nothing is blocking the vents.
To avoid injury to personnel and or separated samples inside the unit; do not
use the following materials in the centrifuge:
- Flammable and explosive materials
- Strong chemical materials
- Toxic or radioactive materials
Contents
Number|
Name
| Qty|
Note
---|---|---|---
1|
Centrifuge
| 1 set|
2|
Rotor (included)
Capacity =< 10-15mL x 6 tube holders
| 1 2 pc|
(six included inside unit & six extra)
3|
Black silicone tube cushions
| 6 pc|
4|
Fuse
| 1 pc|
4 AMP
5|
Operating manual
| 1 pc|
Upon receipt, please examine the package, if there’s anything wrong, please kindly contact your dealer promptly.
Accessories
If you need optional auxiliary parts, please contact our company or authorized dealers.
Item
01| Description| Number of Pieces| Image
---|---|---|---
XC-06| Replacement tube
holders| Set of 6 pcs|
Specifications
Function/ Parameter | Technical Data |
---|---|
Environment of use | ♦ Clean, dry and flat surface |
♦ Stable work surface
♦ Altitude:
♦ Relative humidity:≤80%
Ambient Temperature| ♦ +5°C∼-40°C
Applicable Voltage| ♦ AC110V 60Hz (or 220V 50Hz-220 Model Only)
Setting Time Range| ♦ 1∼60min
Maximum RPM| ♦ 4000 r/min
Maximum relative centrifugal
force (RCF)| ♦ 1000 rpm- 112 RCF
♦ 1500 rpm- 252 RCF
♦ 2000 rpm- 447 RCF
♦ 2500 rpm- 699 RCF
♦ 3000 rpm- 1006 RCF
♦ 3500 rpm- 1370 RCF
♦ 4000 rpm- 1790 RCF
Rotor length| ♦ 5 cm
Radius| ♦ 10 cm
Maximum capacity| 15mL x 6
Noise (maximum speed)| ♦ ≤70dB
Dimensions (machine)| ♦ 11 x 11.5 x 10 in. (28 x 29 x 25 cm.)
Net Weight| ♦ 9 lbs
Dimensions (package)| ♦ 12 x 12 x 11 in. (32 x 32 x 28 cm.)
Gross Weight| ♦ 11 lbs
Notes of Operation
- Ensure the rotor is tightened before running the centrifuge.
- Do not open the lid while the rotor is spinning. Allow centrifuge to come to a complete stop before attempting to open the lid and remove specimen tubes.
- Use only with accessories provided or parts designed specifically for use with this model.
- Do not use centrifuge when the door is open.
- Make sure the specimen tubes are equal and balanced. (Unbalanced tubes will cause the machine to shake and not produce a smooth spinning motion).
- Before use check rotor. Cease use if there is obvious corrosion or damage signs on the rotor.
Control Panel
Control Panel
Time LED: Time in min
+ -: increase/decrease time/rpm’s
Switch: alternate display between time/speed
On/Off: run/stop
Setup
- Open box and remove centrifuge from packaging
- Place centrifuge onto flat surface
- Open lid and remove foam packaging inside
- Plug Power cord from back of unit into power supply, unit will power up and your centrifuge is ready for use.
Operations
Time Adjustment
- Press “Switch” to alternate to time setting. Push the “+” or “-” button to adjust the time setting. To increase the run time push the “+” button until desired time reached or maximum of 60 minutes. (See section 6.1). To decrease the run time push the “-” button until desired time reached or minimum of 1 min. (See section 6.1). If the numbers above “TIME LED” are not Lit, make sure AC power cord is completely plugged into socket.
Speed Adjustment
- Push the “SWITCH” button one time to adjust the speed setting. The 4 digit number beside the word “TIME/SPEED LED” should change to rpm’s. (See section 6.1). If the numbers above “TIME/SPEED LED” do not change to rpm’s, then continue to push the “SWITCH” button until this happens.
- This centrifuge default setting for speed is in RPM’s. Please select desired speed (shown in RPM’s) by using the “+” or “-” arrow buttons until the display panel reads your desired settings.
Spinning Centrifuge
2. Insert the centrifuge tubes in a manner that will balance the specimen tubes’ weight. Always insert an even number of tubes on opposite sides of the rotor with approximately the same amount of specimen in each tube.
Now that the desired time is selected, you can now run your samples.
- Make sure lid is closed and locked.
- Press the “On/Off” button, and centrifuge will begin spinning.
- After the centrifuge has run for the programmed amount of time unit will begin to decelerate and come to a complete stop. Please note that the lid should never be opened until centrifuge has come to a complete stop.
Note: This centrifuge lid is designed to automatically stop when the lid is open. If you have any problems operating the centrifuge; Please contact your dealer or authorized provider.
Upon receipt of your new centrifuge
we recommend you record the following information for future reference:
Date purchased/received:___
Dealer: __
Dealer’s phone number: ____
Warranty
The manufacturer warrants this instrument to be free from defects in material
and workmanship under normal use for 1 year from the date of purchase. It does
not cover damage resulting from abuse or misuse, repairs or alterations
performed by unauthorized repair technicians, or damage occurring in transit.
If you have questions concerning this product or warranty, contact the dealer
from whom it was purchased. For warranty service, product should be well
packed to avoid damage in transit, preferably in original box and packing.
Include your complete return address and telephone number as well as a
description of the difficulty, date and place of purchase, and ship to the
address below. If under warranty it will be repaired or replaced at no charge
and returned. If misuse, alterations, accident or abnormal conditions of
operation caused failure, an estimate for repairs will be provided for your
approval prior to work being performed.
E2 v.2 © 2015