SUMIKO Blue Point #3 High Output Moving Coil Phonograph Cartridge User Manual
- June 5, 2024
- SUMIKO
Table of Contents
SUMIKO Blue Point #3 High Output Moving Coil Phonograph Cartridge
Design Specifics
The Blue Point #3 builds upon the strengths of, and lessons learned from one of the best-selling high-end cartridge in the world, The Blue Point Special. No aspect of the cartridge’s design was left to chance. To minimize unwanted external resonance storage the cartridge body was modified to stiffen the housing and create a low mechanical impedance mounting system. Combined with the optimization of mass, this provides the perfect environment for the carefully chosen working elements of the cartridge.
Suspension Design
Virtually all conventional moving coils get by with a sus- pension element
made from a butyl material. For the Blue Point, we chose to use a synthetic
rubber material so the specified parameters of the suspension are held to
design tolerance for a longer period of time. These steps help ensure the
cartridge will perform as designed for a longer period of time lasting
considerably longer than other designs. This provides a combination of
superior dynamics and excellent long term tracking ability.
Why an Elliptical Stylus Shape?
Your BP#3 is fitted with a specially ground, low mass stylus, a design type
where the inherent character is always musical. With care and precise
adjustment, tremendous amounts of detail can be extracted. Unlike some
designs, even if it is not adjusted within a gnat’s whisker of perfection,
your cartridge will still deliver an intensely musical experienc.e The
aluminum cantilever chosen for the BP#3 provides excellent energy transfer
from the stylus to the coils.
High Output moving coil design philosophy
The Blue Point #3 High Output version outputs 2.5mV (@1kHz), a level high
enough to be useable with standard MM (Moving Magnet) phono preamps. The
recommended loading is 47k Ohms, again, the same as the standard for MM
cartridges, ensuring compatibility with a wide range of phono preamps. Sumiko
pioneered the concept of high out moving coil cartridges and we were issued
U.S. patent #4675859 for what we call intensified field focus
The Goal
The purpose of fine adjustment of a phono cartridge is simple: you are trying to optimally align the playing sur- face of the stylus with the groove wall in the record, doing so in a way that the stylus is securely seated against the groove wall with neither too much, nor too little force, in all directions. All of these adjustments would be child’s play if the stylus was the size of a baseball, and the groove was the size of a rain gutter. The tricky part comes in the fact that the stylus is so tiny that no one can possibly see the scanning surface of the stylus (the part that actually touches the record groove) without the aid of a microscope. Add in the fact that the stylus is moving at the equivalent of about three hundred miles per hour and is being asked to maneuver faster than a Formula One race car and the problem becomes quite complex. Luckily, with a little patience and training, there is a grand equalizer: your ears. So, let’s get to it!
Before You Begin Setting Up the Cartridge
Preliminary User Note: Before we begin, it may prove helpful to obtain an
understanding of correct set-up, why you are doing it, and what you hope to
accomplish. This is not intended as a theoretical treatise, but merely as
simple, practical tips on how to go about getting the most from your new
cartridge.
WARNING: Phono cartridges are inherently delicate things. Work SLOWLY,
and in a methodical fashion with lots of light and enough space to work
comfortably. Rushing to make an adjustment is likely to have the unhappy
result of, at best, performing a lousy adjustment, or at worst, damaging the
cartridge. Don’t forget the wis- dom in the old adage, “There never seems
enough time to perform a task properly the first time, but there is always
enough time to do it again.” Slow down and enjoy the process! All of the
adjustments we speak of in the following pages are small incremental
adjustments. As a general frame of reference, virtually any change you make
will be less than one-eighth of an inch. Some, such as vertical tracking angle
adjustments, may end up being a few thousandths of an inch!
Basic Mounting Instructions
There are two basic functions to accomplish: physically bolting the cartridge
to the tonearm and making the necessary four electrical connections to
complete the audio pathway. Note: Make sure the volume control is turned all
the way down or an alternative to the phono input is selected when you are
making any electrical connection.
Handle with extra care! The Blue Point #3 comes with a stylus guard cover
which should be left in place during the initial mounting, electrical
connections and initial alignment.
We prefer first to bolt the cartridge to the headshell, using the supplied
stainless-steel mounting bolts. Next, attach the audio leads. The order in
which you perform these tasks is functionally irrelevant, but if you should
slip while trying to affix one of those tiny wires, at least the cartridge
will not drop to the ground! Using a pair of small needle-nose pliers or
tweezers, fasten the color-coded audio leads to the appropriate color-coded
pins on the cartridge. If the headshell space is limited, as it is on some
tonearms, loosen the mounting screws and move the cartridge forward to give
yourself more room to fit the wires. If you still don’t have enough room,
remove the cartridge entirely and then reattach the cartridge when you’re done
with the wiring.
A Tip to Remember
One helpful little tip to remember; “Christmas colors go together to form the
right channel, and hot colors are ‘hot.”‘ By this we mean that red and white
(red hot and white hot) are the positive leads, so white is left channel hot
and red is right channel hot. To complete the Christmas color scheme, green
attaches to right ground. The only wire left is blue which is, by process of
elimination, left channel ground. We may have thoroughly confused you by this,
try it once or twice and it should make more sense.
And the Basic Setup Terms are …
Before we press on too far ahead, let’s identify the basic concepts and terms
that define the setup procedure. Basically, there are four different alignment
angles to be concerned with: 1. HTA, 2. Offset Angle, 3. VTA, and 4. Azimuth.
There is also the pre-loaded weight of the arm and cartridge combination,
which is Tracking Force. Additionally, there is the compensation required to
counteract the inward pull developed by tracking a spirally-wound groove. This
is called Anti-Skate. A discussion of these parameters will follow later on in
this manual. Careful attention paid to the adjustment of all of these
parameters will result in optimal performance.
Setting Tracking Force
Now that you’ve got the cartridge mounted and wired, the next thing you’ll
want to do is set the basic tracking force, or the amount of down force that
will be applied by the stylus on the record. This setting not only determines
the amount of stylus force in the record groove but also loads the suspension
system and aligns the coils to the yoke. This also determines the tracking
ability and properly aligns the coils to the magnetic field of the magnets,
contributing to the linearity of the cartridge. Initially, the cartridge
should be set to 2.0 grams of tracking force.
To set the tracking force, first balance the tonearm. As you move the
counterweight backwards, the cartridge end of the arm will lighten (reducing
stylus tracking weight), and moving the counterweight forward will increase
the vertical tracking force (the weight on the stylus). The first objective is
to “statically balance” the arm, which is to achieve the proper weight balance
such that the arm “floats” level. Practice a bit and you’ll find it’s fairly
easy to achieve. Remember to exercise caution so that the stylus and
cantilever are kept clear of any obstructions and are prevented from striking
anything that could cause damage. After achieving static balance, adjust the
counter-weight to apply 2.0 grams tracking force by bringing the weight in
toward the cartridge.
A Brief Discussion on Tracking Force
Tracking force is primarily responsible for pre-loading the cartridge’s
suspension system so that it works optimally. It is useful to think of a
cartridge’s suspension as being similar to an automobile suspension. Too stiff
of a shock absorber setting on too light a car will result in poor dynamic
performance and a crude, uncomfortable ride. Similarly, a relatively stiff
performance suspension on a phono cartridge, set at too low a tracking force,
will result in performance that is bass light, and harsh sounding in the high
frequencies. While the mechanical characteristics of the tonearm (how inert it
is, how well it allows energy to pass through it, etc.) also bear on this, the
tracking force is something we can exert control over.
HTA (Horizontal Tracking Angle) and Offset Angle
We’ve combined these two categories because they are essentially interrelated.
Other than a brief description of what you are adjusting for, we will leave
these adjustment parameters to the maker of whatever tonearm you intend to
use. Many tonearm manufacturers supply alignment templates with their tonearms
and their tonearm alignment geometry may be maximized around this template.
Briefly, HTA or Horizontal Tracking Angle is the fore/aft adjustment of the
cartridge within the headshell (though some tonearms, like SME, use a fixed
cartridge position in the headshell and require the entire arm be moved). By
moving the cartridge position forward or backward in the headshell, the angle
that the cartridge will describe over the entire playing surface of a record
can be altered and optimized. Within the headshell the angle of turn-in
required is a function of the amount for- ward or backward the cartridge is
moved. Suffice it to say that obtaining a good alignment gauge and following
its instructions is the best way to go about making these adjustments. Note:
When the HTA alignment process is completed, remember to re-check the tracking
force of the cartridge as any movement of the cartridge in the headshell
forwards or backwards will change the tracking force value.
Anti-skate Adjustment
It is now time to adjust for anti-skate.. On all tonearms, the anti-skate is
located towards the rear of the tonearm, in the vicinity of the bearing
housing. The adjustment device typically takes one of the following forms. 1.
A dangling weight hanging off the side of the arm, with some adjustment
mechanism, such as slots cut into a fixed post. The further away from the
tonearm, the string is attached on that post, the greater the anti-skate
applied. 2. Magnetic anti-skate calibrated via, a dial with numbers printed on
the dial face indicating 1, 2, 3, grams and fractions thereof. Anti-skate, as
defined earlier, is a force applied to approximately counteract the inward
pull created by the record’s decreasing radius spiral groove. We say
approximately because the record groove is constantly changing, and no
existing anti-skate is up to the task of correcting for all the non-linear
forces present. The velocity of the groove changes with frequency, amplitude
and position of the stylus, relative to the center of the record. As the
velocity changes, so do the amount of inward pull being exerted on the
stylus/suspension system. There is no way to completely and accurately correct
for this dynamic situation. While it is common to suggest setting the anti-
skate to exactly match the tracking force indicated (e.g., 2 grams tracking
force equals 2 grams indicated anti-skate force), our research suggests that
an alternative approach is called for since the compliance of the suspension
is not at all taken into account in this approach. Modern moving coil
cartridges are lower in compliance and are not as affected by the vector force
applied by the record’s groove. In the case of the Blue Point, we suggest
about 2/3 as much anti-skate as the tracking force indicated, so in the case
of the cartridge, about 1.25 – 1.3 grams anti-skate indicated. If more anti-
skate is applied, the sound may begin to lose delicacy, and a sense of
constriction will set in. Test records will not yield particularly good
results because they are testing for worst-case scenarios. Using our method in
real-world conditions, the user will obtain higher average results. Going back
to the automobile analogy, running a higher anti-skate setting is like driving
on snow tires year-round on the unlikely chance you might run into a
snowstorm. Additionally, some will recommend the use of mono records to set
anti-skate. The problem with this method is that a mono record is cut quite
differently from a stereo record. Additionally, some recommend a blank record,
however, this does not duplicate actual stereo record tracking either. Setting
the anti-skate should be done in a real-world situation and not a steady-state
or laboratory experiment simulation.
Fine Tuning, VTA and Azimuth are the Tricky Parts
To this point, all the adjustments we’ve had you do have been relatively
straightforward, intended to obtain and set a specific parameter. From this
point on, though, the work becomes more qualitative and good judgment enters
into it, as you will be called on to make assessments entirely subjective in
nature. Relax and go slowly. It is unlikely that you will totally foul things
up. With a little patience and finesse, you can obtain a more musically
satisfying experience from your turntable.
The Line Drawings
As the saying goes, “a picture is worth a thousand words,” so we chose to add
some simple illustrations to this section of the manual for use as a guide,
illustrating what your cartridge will take like when both properly and
improperly adjusted for VTA and azimuth on your tonearm. Although there is no
way to show you exactly what a perfectly adjusted cartridge will look like,
the line drawings can serve to give you a mental template for your particular
setup. Found on pages 19 and 22, these line drawings will illustrate gross
problems found in VTA and azimuth adjustments. Directly after the text on
adjusting VTA and Azimuth, the illustrations on page 23 allow you to see what
your cartridge should generally look like when properly set up. For instance,
if your particular cartridge sounds best with a more exaggerated negative
rake, so be it. If the visual aspect of your particular cartridge looks as if
the azimuth is tilted very slightly to one side, but this is the attitude that
allows for the best sound, this too is okay. So, use the guide as just that, a
guide. Explore the possibilities of what you can extract from your cartridge,
and you will be rewarded with great music.
VTA Adjustment
Vertical Tracking Angle (more precisely, Stylus Rake Angle) is the front-to- rear rake of the stylus within the groove and is controlled by raising or lowering the pivot end of the tonearm. Look for a small screw, possibly two, located low on the mounting collar where the tonearm meets the turntable. You should always refer to your owner’s manual to familiarize yourself with all the functions of your arm, so now would be a good time to do that for this function. VTA adjustment’s primary effect is upon the time domain behavior of the musical presentation. Often, we hear and read about tonal balance differences attributed to changing VTA. But in the changing of the time signature, affects the arrival of high frequencies relative to the low frequencies. The more one rakes a cartridge back, the more the high-frequency content is “slowed” resulting in, to a point, subjectively richer, rounder sound. Inversely, raising the VTA will have the effect of sharpening up or focusing the sound; too high a setting and stridency results. When you get very close on VTA adjustment, note that adding a touch more or less tracking weight (assuming your initial tracking weight is 2.0 grams) is a simple way of fine-tuning the VTA. Adding a tiny bit of weight is effectively lowering the arm height and reducing the weight will have the apparent effect of raising the back of the arm. Be careful, since adding force also changes the mechanical damping of the system as well as the relationship of the coils to the magnets. Still, for all but the most talented among us, adjusting the weight is, at the very least, a useful interim tool in fine-tuning VTA. Always, always, always start with the arm lower than necessary and slowly raise its position. Trying to start above the eventual point and lowering the arm will not work, we guarantee it. Please refer to the line drawings on the facing page. On page 19, illustration No. 3 shows a cartridge with 2° of negative cartridge rake, which is a good starting point for you to work with. Remember, the time domain will not lie, and the ear is quite sensitive to time arrival cues, whereas trying to guess whether a piece of program material has too much or too little bass, for example, is an exercise in madness. When VTA is right, the sound will take on a properly large image scale (not bloated, simply full in size), with an easy “breath” to the music that emulates the natural ebb and flow of live music. When wrong, the sound can be either too dull at one extreme, or pinched and small sounding at the other. Remember, you are working within a very small window of acceptability. The total height differential you are working with is perhaps 1/4″ and that includes the “falsing zone”, that area above and below the correct position that you will wind up finding by trial and error.
Illustration No. 1 shows a neutral attitude of rake. The cartridge will
almost never look like this on the turntable. There will always be some
negative rake angle to correct the cartridge setup.
Illustration No. 2 shows a positive rake of 2°. This too will never be
the correct adjustment in a properly set-up cartridge. This attitude will
result in a thin sound and could result in record damage.
Illustration No. 3 shows what a negative rake angle looks like. This
illustration shows a negative 2° attitude. This degree of rake angle is too
great for the cartridge but serves to show what your car- tridge should
generally look like when starting the VTA adjustment process.
Azimuth Adjustment
This is probably the most misunderstood adjustment in analog audio. We believe
this stems from the fact that its primary effect can be relatively subtle, but
when it is not precisely correct, its effect on other parameters such as VTA
can be dramatic. As you can see from the diagram, azimuth describes the left-
right angular orientation when viewed from the front of the cartridge.
Necessarily,y this affects the balance of left-versus-right information, but
not as dramatically as you might expect. Because of the tiny dimensions
involved, its effect will be most noticeable in the lower treble (basic high
frequency performance). Additionally, the groove-tracing ability will be
affected by incorrect azimuth adjustment. Because the stylus does not sit
perfectly vertical in the grooves (see VTA), azimuth misalignment will also
result in the stylus “scrubbing” in the grooves, resulting in mistracking and
poor high-frequency performance. Your cue is to listen for stridency,
sibilants, or an unusual metallic clang on instruments that do not normally
produce a sibilant or clang. The sound of an unnatural “task task” on cymbals
and other leading-edge transients are telltale signs of misaligned azimuth. A
very slight adjustment will correct for this problem. Only in grossly
misadjusted arms will you hear an actual channel imbalance whereby the left
channel actually sounds as though it is playing louder than the right, or vice
versa.
Note: When you have achieved the correct azimuth, the sound will take on
a tidy, integrated character that simply sounds “right.” Some people have
reported good results by playing mono records and listening for focus of image
as a determinant for azimuth. We are reluctant to recommend this method since
the groove geometries of mono records are different from those of stereo
records. We believe that a real-world, dynamic set-up is the way to achieving
a satisfying cartridge alignment and using mono records to set up for stereo
records is not applicable.
Addendum: Further experimentation will reveal that small changes made to
azimuth may necessitate an additional slight change to VTA. This area is the
trickiest part of fine-tuning. That is, the interactive nature of and VTA
changes can be misleading. A legitimate improvement to one area can lead to a
subjective degradation in another simply because an initial error has been
exposed. As always, it is important to listen rationally and build up enough
of an experience base so that you can easily determine whether the changes you
made are a true improvement or simply an irrelevant alteration.
Illustration No. 4 shows the attitude of the cartridge when the azimuth
is set correctly. The bottom of the cartridge may be used to set azimuth by
eye. However, the final setting will be done by ear.
Illustrations Nos. 5 and 6 show azimuth errors of 2°, lilted severely to
the right and left. These angles may result in volume imbalances, biased
toward the right or left channels. More importantly, this azimuth error will
not allow proper tracing of the grooves resulting in mistracking. An edgy,
hyper-detailed high-frequency sound will result if the azimuth is set wrong
.
Please refer to the manual section titled Azimuth Adjustment (p.20) for
further suggestions. These illustrations show what the cartridge should look
like when properly mounted and adjusted
Illustration No. 7 shows azimuth at 0° of error in side-to-side tilt.
Illustration No. 8 shows a negative rake angle, (VTA), of 1.2°. This
angle will result in a very fast, dynamic, and robust sound.
Illustration No.9 shows a steeper negative rake angle than Illustration
No. 8 where the angle has been increased by only 0.3, barely perceptible to
the naked eye. The negative angle of 1.5° will result in a more rounded low-
frequency dynamic structure, less clarity in the midrange, and a reduced high-
frequency component.
Wrapping Up the Installation
Now that you have gone through the basic set-up procedure of your cartridge
you can expect the solutions that are outlined in this manual to aid you in
optimizing your cartridge as it breaks in over time. You will find, that for
the first 10 – 20 hours of playing time, the cartridge will go through changes
in character, sound quality, and perceived output. As the suspension system
settles in and becomes mechanically “free”, the VTA position you originally
found to be satisfactory will become less and less optimal and aberrations in
the presentation will appear. Typically, the VTA will have to be changed so
that the back of the tonearm is raised, which is why we advise you never to
start with the VTA at a point that requires that you work your way downward.
As the suspension settles in and the VTA becomes more and more wrong, the
dynamic structure of the music will become compressed, leading to the
perception of reduced output. We recommend that you do not fiddle with the VTA
too much for the first twenty hours of play time as the frequency and degree
of changes required to keep the cartridge per- forming optimally during this
time might drive you crazy. So, sit back and relax, accept the performance
through the initial break-in period and listen to lots of records.
Warranty
This product is warranted to be free of all defects in material and
workmanship for one year from the date of original purchase by the original
owner. A purchase receipt or other proof of original purchase will be required
before warranty service is rendered. This warranty is not transferable and
does not apply to any defects caused by negligence, accidents, misuse,
modification, disassembly, or repair by other than the manufacturer, or by
other than normal use for which this product was intended. Within the period
of this warranty, Sumiko will repair or replace at our service center located
in
Minnesota any part proving defective in material or workmanship. Costs, except
for collateral expenses related to replacing or repairing a defective part
under warranty will be assumed by Sumiko, except for the cost of transporting
and insuring the product to our service center. The buyer must notify Sumiko
of any defect, malfunction, or nonconformity promptly upon discovery. Within
30 business days after receiving the defective product from the buyer, Sumiko
will repair or replace the defective part. We neither assume nor authorize any
representative or other person to assume for us any other liability in
connection with the sale or shipment of our products. We reserve the right to
make changes or improvements in our products without incurring an obligation
to similarly alter products previously purchased. The buyer has the right to
bring any action at law or equity to resolve disputes concerning or to enforce
the provisions of this warranty.
Specifications
- Generator
- Suspension Frequency Response Output Voltage
- Moving Coil
- Synthetic rubber 13 Hz – 35kHz 2.5 mV
- Channel Separation >
- 30dB, @ 1 kHz
- Channel Balance
- <1 dB@ 1 kHz
- Optimum Tracking Force 2.0 Grams
- Tracking Force Range Dynamic Compliance
- Internal Impedance Load Impedance Cantilever
- Stylus Type
- Output Terminals Weight
- Mounting System
- 1.8 – 2.2 Grams
- 12 x 6cm/dyne
- 135 Ohms
- 47k Ohms
- Aluminum
- 0.3 x 0.7mil /Elliptical 1.2 mm gold plated (EIA) 6 Grams
- Standard 0.5”
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>