LIFEPAK 3312981-000_B Code Management Module Instruction Manual Product Information
- June 1, 2024
- LIFEPAK
Table of Contents
3312981-000_B Code Management Module
“`html
Product Information
Specifications
- Manufacturer: Fluke Biomedical
- Model: CodeManagement Module
- Part Number: 3312981-000_B
Product Usage Instructions
Performance Inspection Procedures (PIP)
The Performance Inspection Procedures (PIP) are manual test
procedures used for evaluating the CodeManagement Module’s
operational performance. These tests are essential to ensure that
the module operates within the required specifications.
When to Perform PIP
Perform the PIP as part of a regularly scheduled preventive
maintenance routine and after any repair, replacement, or
calibration procedure.
Equipment Required for PIP
-
Safety Analyzer: Fluke Biomedical ESA612 (or
equivalent) -
Digital Pressure Meter: Fluke Biomedical
DPM2Plus (or equivalent) -
Alligator Clip: Fluke P/N AC220 (or
equivalent) -
Cable Assemblies:
- 3 Lead ECG: 11110-000029
- 5 Lead ECG: 11110-000066
- QUIKCOMBO Therapy Cable: 11110-000040
- QUIK-COMBO to Snap Termination Cable: 3009139-001
Personnel Requirements
Service personnel conducting the PIP must be familiar with the
operation of the LIFEPAK 20e defibrillator and CodeManagement
Module. They must meet specific requirements related to biomedical
electronics.
Test Equipment Verification
All test equipment used for the PIP must have a current
calibration label from a certified calibration facility.
Workstation Power
The workstation power must be connected to a grounded power
source for safe operation.
Frequently Asked Questions (FAQ)
Q: Why is it important to perform the PIP?
A: The PIP ensures that the CodeManagement Module operates
within the required specifications and helps detect any
malfunctions or out-of-tolerance conditions that need
correction.
Q: What should service personnel be familiar with?
A: Service personnel must be familiar with the operation of the
LIFEPAK 20e defibrillator and CodeManagement Module, and meet
specific requirements related to biomedical electronics.
“`
CodeManagement Module
Section Contents
Performance Inspection Procedures
6
The Performance Inspection Procedures (PIP) are a set of manual test
procedures used for an operational closed-case evaluation of the
CodeManagement Module.This section describes the test procedures you will
perform to determine if the CodeManagement Module is operating within the
required specifications. Investigate and correct any malfunctions or out-
oftolerance test conditions detected during the PIP.
The PIP comprises safety and performance tests recommended by AHA/ASHE (American Hospital Association/American Society for Hospital Engineering) Maintenance Management for Medical Equipment and International Electrotechnical Commission (IEC) Technical Report 1288-2, Maintenance of Cardiac Defibrillators-Monitors.
Perform the PIP as part of a regularly scheduled preventive maintenance routine. Also, perform the PIP after any repair, replacement, or calibration procedure.
PIP/TCP Scope and Applicability PIP/TCP Resource Requirements
PIP/TCP Test Equipment Requirements
PIP Instructions
Previous Page
TCP CO2 Calibration
PIP Checklist
Table of Contents
3312981-000_B
Back
Index
Next Page
CodeManagement Module
PIP/TCP Scope and Applicability
Performance Inspection Procedures
The PIP applies to the CodeManagement Module exclusively. To complete the PIP,
you must perform the combination of manual tests outlined in the PIP
Instructions section of this electronic service manual. All PIP tests
applicable to the CodeManagement Module configuration under test must be
performed.
Refer to the PIP/TCP Resource Requirements for a listing of the necessary
qualifications for PIP equipment, test equipment verification, workstation
power, and personnel.
Refer to the PIP/TCP Test Equipment Requirements for a listing of test
equipment, including specifications, required to complete the PIP/TCP.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
Performance Inspection Procedures
PIP/TCP Resource Requirements
This section describes the requirements for PIP/TCP equipment, test equipment verification, workstation power, and personnel.
Equipment
To perform the PIP, you must use the equipment listed in the PIP/TCP Test Equipment Requirements table. Although the table lists specific test equipment by manufacturer, test equipment with equivalent specifications may be substituted.
Test Equipment Verification
All test equipment used to perform the PIP must have a current calibration label, issued by a certified calibration facility.
Workstation Power
The ac line power to the workstation used must be connected to a grounded power source.
Personnel
Previous Page
Service personnel who perform the PIP must be thoroughly familiar with the
operation of the LIFEPAK 20e defibrillator and CodeManagement Module and must
meet at least one of the following requirements (or the equivalent):
Associate of Applied Science, in the field of electronic technology.
Certificate of Technical Training, in the field of electronic technology.
Equivalent electronics experience with a preferred emphasis in
biomedical electronics.
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
Performance Inspection Procedures
PIP/TCP Test Equipment Requirements
The following is a list of test equipment required to conduct the PIP/TCP.
Equipment/Cable
Specifications
Manufacturer/Part Number/ Catalog Number
Safety Analyzer
110 or 220 Vac line voltage Current range: 0-1999 mA
Current accuracy: 5% of reading or 1 digit (whichever is greater) Resistance
range: 0-1.000 Ohm
Fluke Biomedical ESA612 (or equivalent)
ESA612 adapter box
Provides additional ECG snap connections.
Fluke Biomedical model 1210
Digital Pressure Meter
+/- 1% of full scale. Used as a vacuum manometer.
Fluke Biomedical DPM2Plus (or equivalent)
Alligator Clip
Used to access to LIFEPAK 20e defibrillator/monitor Fluke P/N AC220 (or equivalent) ground stud for electrical safety testing.
Cable Assembly, 3 Lead, ECG
Standard accessory with the 3-Lead LIFEPAK 20e defibrillator/monitor.
11110-000029
Cable Assembly, 5 Lead, ECG LIFEPAK 20e defibrillator optional accessory.
11110-000066
Cable Assy-Therapy, QUIKCOMBO
11110-000040
Cable, QUIK-COMBO to Snap Termination
For use in electrical safety testing with QUIK-COMBO 3009139-001 therapy cable.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
Performance Inspection Procedures
PIP/TCP Test Equipment Requirements
Equipment/Cable
Specifications
Cable Assy, LP20 SPO2 Test
For use in electrical safety testing.
Paddle Assy-Detachable, LP20
LIFEPAK 20e defibrillator/monitor optional accessory.
Adapter – Test, Paddles Leakage, For use in electrical safety testing with standard
LP12 & LP20
paddles.
QUIK-COMBO Leakage Cable
For use in electrical safety testing with standard paddles.
Cable, Test, ECG Snap to Banana Plug
For use in electrical safety testing.
RS-232 (DB9 M/F) Cable
USB to Serial Adapter (as required)
Computer
OS, Windows XP, 7
Software, CodeManagement Module Computer Assisted Test
Wireless network (WPA/WPA2 enterprise or personal)
Required to test Wireless functionality.
LIFENET Device Agent Component (LDA)
Used to check and calibrate CO2.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Manufacturer/Part Number/ Catalog Number
3201832 (Multiple REF) 3200936 (Multiple REF)
3206631
3207066
3305684
Commercial Commercial
Commercial 3313963
Commercial
21340-000836
Back
Index
Next Page
CodeManagement Module
Performance Inspection Procedures
PIP/TCP Test Equipment Requirements
Equipment/Cable
Specifications
Manufacturer/Part Number/ Catalog Number
Paper clip, Large
LIFEPAK 20e defibrillator/monitor with CodeManagement Module compatible
software Tubing Assy – CO2 Calibration Filterline H Set Adult/Pediatric
Chemical – Gas, Mixture, Calibration, Aerosol Tubing Assy – Service, CO2 Male
Luer
Self powered 1.1 Compliant USB device AC power cable, hospital grade (USA)
Stopwatch
Used to access the CodeManagement Module reset button.
For use in CO2 calibration and calibration check. For use in CO2 calibration
and calibration check. For use in CO2 calibration and calibration check.
For use in CO2 leak testing. Used to connect CO2 tubing assy to CO2 exhaust
port. Used as part of the computer assisted test software, 3313963, to test
the USB interface. International users – local hospital grade cable is an
acceptable substitution.
Commercial
Physio-Control, Inc.
21330-000239 11996-000068 21300-001572
21330-000238 Value Plastics MTLL250
Belkin 4-Port Hub F5U021V (or equivalent) Physio-Control 11140-000015
Commercial
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP General Instructions
This section lists the general instructions for performing the Performance
Inspection Procedure (PIP).
· All required PIP tests applicable to the CodeManagement Module under test
must be performed.
· The PIP Checklist is provided as an optional tool for the recording of test
results.
· Throughout the body of this PIP, the LIFEPAK 20e defibrillator/monitor is
referred to as the defibrillator.
· Throughout the body of this PIP, the CodeManagement Module is referred to as
the Device Under Test or DUT.
· Unless specified in the test step, the DUT should have a battery installed
and be connected to AC power.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
Warnings
WARNING!
Shock hazard. The LIFEPAK 20e defibrillator/monitor discharges up to 360 J of
electrical energy through the defibrillator cable. You must safely discharge
this electrical energy. Do not attempt to perform this procedure unless you
are thoroughly familiar with the operation of the LIFEPAK 20e
defibrillator/monitor and CodeManagement Module.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP Exterior Physical Inspection
To perform an exterior physical inspection: 1. Inspect the DUT exterior for
the following:
Damage Excessive wear Improper mechanical function Damaged connectors
2. Pick up and turn over the DUT and listen for loose or rattling hardware.
Locate any loose or rattling hardware and tighten or replace it.
3. Inspect the rubber feet on the underside of the DUT. Reinstall or replace
rubber feet as necessary.
4. Inspect carrying strap and storage bags for damage (if equipped).
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP- Power Port and AC Mains LED
To test the Power Port and AC Mains LED: 1. Verify the DUT battery is installed. 2. Disconnect the AC power cable from the DUT to the power source. 3. Verify the DUT AC Mains LED is OFF. 4. Connect the AC power cable from the DUT to the power source. 5. Verify the DUT AC Mains LED is ON. 6. Verify the defibrillator AC Mains LED is ON. 7. Turn ON the defibrillator and verify the defibrillator turns ON. 8. Turn the defibrillator OFF.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP- Reset Button
To test the Reset Button: 1. Press and hold the reset button for a minimum of 1 second and release. Note: Use the straightened end of a large paper clip to access the reset button.
Note: Upon initial release of the reset button, all LEDs will glow steadily.
After ~30-60 seconds, the AC mains LED will flash and then glow steadily.
2. Verify the DUT AC mains LED flashes, then glows steadily.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP- CO2 Leakage Test
Note: Perform the CO2 tests if the device is equipped with the CO2 option.
Otherwise, skip to PIP – Computer Assisted Tests.
To test the CO2 monitoring system for leaks:
Note: Make sure the defibrillator is turned OFF and no tubing is connected to
the device.
1. Open the hose clamp on the CO2 leak test kit and depress the syringe
fully.
2. Connect the tubing to the front panel CO2 connector. 3. Use P/N MTLL250 to
connect the tubing to the back panel CO2 gas outlet. 4. Secure the fittings
that connect to the device to avoid leakage.
vacuum manometer
Previous Page
to vacuum manometer
CO2 leak test kit REF 21330-000238
Table of Contents
3312981-000_B
Section Contents
use P/N MTLL250 to connect tubing to back-panel gas outlet
to front-panel CO2
connector
DUT
(Continued on next page)
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP- CO2 Leakage Test (continued)
5. Pull the syringe plunger out to induce a vacuum into the system. When the
vacuum manometer indicates approximately -250 mmHg +/- 50 mmHg, close the
tubing clamp firmly.
6. Wait a minimum of three (3) seconds after closing the tubing clamp, then
obtain an intitial vacuum reading.
7. Verify that after 30 seconds, the change in vacuum reading is less than 15
mmHg (20 mBars).
8. Disconnect the tubing connection from the front panel CO2 connector to
release the vacuum.
9. Disconnect the tubing connection from the back panel gas outlet and
release the tubing clamp.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP- CO2 Calibration Check
To check CO2 calibration: 1. Connect the RS-232 (DB9 M/F) cable from the DUT to the test computer. 2. Launch the LIFENET Device Agent (LDA) from the test computer. 3. Follow instructions as prompted by the LDA to check CO2 calibration.
CO2 calibration kit REF 21330-000239
Cut CO2 filter line REF 11996-000068 to ~18 inches in length
Remove luer Fitting
to front-panel DUT CO2 connector
Calibration gas REF 21300-001572
DUT
4. Verify that the measured gas concentration reads 5.0% ± 0.3%. If the
measured value is incorrect, perform the TCP CO2 Calibration.
5. Disconnect the CO2 filter line from the DUT.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
Performance Inspection Procedures
PIP Instructions
PIP – Computer Assisted Tests
Previous Page
The CodeManagement Module Computer Assisted Test software automatically tests
the serial communication to the defibrillator, the USB interface, battery
power, battery I2C, Wireless and Wireless switch functionality.
1. Ensure that an operational wireless network (WPA/WPA2 enterprise or
personal) signal is being broadcasted.
2. Launch the CodeManagement Module Test software P/N 3313963 from the test computer and follow instructions as prompted.
Note: Use a USB to Serial adapter as required to connect the DUT to the test computer.
Note: If the DUT does not communicate with the test computer, remove both the battery and AC power from the DUT. Wait ~ 1 minute. Reinstall both the battery and AC power. Try the test again as necessary.
3. Verify all tests PASS.
Note:
Table of Contents
If any difficulty is observed during communication between the PC and
the CodeManagement Module while using a USB to Serial adapter, in
particular during the Wireless communications test, it is possible that the
Serial to USB adapter is configured incorrectly, or is incompatible with
the PC and/or its operating system. Replace the Serial to USB adapter
with an alternate model as necessary.
3312981-000_B
(Continued on next page)
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP – Computer Assisted Tests (continued)
Note: A wireless communications test failure could be the result of the
wireless signal. Retry the test as necessary.
4. Turn off the defibrillator and disconnect the RS-232 cable from the DUT.
5. Press the reset button on the side of the DUT to reboot the DUT. Note:
Ensure the DUT completes the reboot before continuing (~1 min).
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
Electrical Safety Tests
Perform leakage current testing in accordance to the following electrical
safety standards:
· IEC (International Electrotechnical Commission) 62353 · IEC (International
Electrotechnical Commission) 60601-1 and 60601-2-4
Warning
SHOCK HAZARD. Failure to properly perform these tests could result in a
failure to detect excessive leakage current. Make sure you are familiar with
your test equipment and these test performance procedures.
Previous Page
Leakage Current flow induced by the application of high voltage to a
material or object with high dielectric strength.
Earth Ground Third wire ground.
Normal Condition (N.C.) AC voltage is applied in either normal or reversed
polarity (that is, measurements made with the POLARITY switch in both NORMAL
[NC] and REVERSED [RM] positions). The earth ground is intact during these
measurements.
Single Fault Condition (S.F.C.) AC voltage is applied in either normal or
reversed polarity (that is, measurements made with the POLARITY switch in both
NORMAL [NC] and REVERSED [RM] positions). The earth ground is NOT intact
during these measurements.
Note: Ensure the RS-232 cable is not connected to the DUT during electrical
safety testing.
Note: Safety analyzer setup instructions are specific to the Fluke Biomedical
ESA612.
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
Performance Inspection Procedures
PIP Instructions
PIP- Protective Earth Resistance LIFEPAK 20e Defibrillator Ground Stud
To test the Protective Earth Resistance from the defibrillator ground stud: 1.
Turn the defibrillator OFF.
2. Set up the test as shown below.
Note: Use Alligator Clip P/N AC220 (or equivalent) to access the defibrillator
ground stud through the DUT bridge opening for the DB15 port.
Previous Page
3. Set the safety analyzer controls to measure Ohms by selecting the Ohm button on the safety analyzer.
4. Verify the measured protective earth resistance is less than 0.3 Ohms (120
and 240 Vac).
3312981-000_B
Table of Contents Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP- Protective Earth Resistance CodeManagement Module Ground Stud
To test the Protective Earth Resistance from the CodeManagement Module ground
stud: 1. Turn the defibrillator OFF.
2. Set up the test as shown below.
Previous Page
3. Set the safety analyzer controls to measure Ohms by selecting the Ohm button on the safety analyzer.
4. Verify that the measured protective earth resistance is less than 0.3 Ohms
(120 and 240 Vac).
3312981-000_B
Table of Contents Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP- Leakage Current Setup
To set up the remaining leakage current tests:
Note: Reference the figures below for test setup using QUIK-COMBO therapy
cable and test setup using STANDARD PADDLES.
1. Connect a banana cable between the CodeManagement Module ground stud and
the safety analyzer V/Ohm/A output.
2. Connect the customer ECG Lead (5-wire or 3-lead) cable between the
defibrillator and the 1210 box.
3. Connect the 1210 box to the safety analyzer at RA snap with an ECG snap to
banana plug cable.
4. Connect the Therapy cable (QUIK-COMBO or Standard Paddles) between the
defibrillator and the safety analyzer at LL and LA snaps.
5. Connect the AC power from the CodeManagement Module to the safety analyzer
AC output.
6. Connect the SpO2 Leakage cable (if equipped) between the defibrillator and
the safety analyzer at RL snap.
Previous Page
Warning
SHOCK HAZARD. Do not defibrillate when connected to the safety analyzer.
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
PIP- Leakage Setup
Performance Inspection Procedures
QUIK-COMBO Therapy Setup
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
PIP- Leakage Setup
Standard Paddles setup
Performance Inspection Procedures
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
Earth Leakage Test Setup
To perform the Earth Leakage test in accordance with IEC 60601, set up the
safety analyzer as follows:
Press the SETUP button on the safety analyzer Press the F4/MORE button
Press the F2/INSTRUMENT button Press the F1/STANDARD button Press the
UP/DOWN arrows to select the 60601 standard Press the DONE button
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP – Earth Leakage Test – Normal Condition (NC)
1. Turn the defibrillator ON. 2. Press the µA button on the safety analyzer.
3. Press the F1/EARTH button on the safety analyzer and set the safety
analyzer controls as follows:
Neutral Closed
Polarity Normal/Reverse
Current Mode AC only
Note: Pause briefly between switching polarity to prevent damage to the
defibrillator.
4. Verify the DUT AC Mains LED is ON.
5. Verify the defibrillator AC Mains LED is ON.
6. Verify the measured current is between 15 µA and 300 µA (120 Vac) or
between 15 µA and 500 µA (240 Vac).
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
PIP – Earth Leakage Test – Single Fault Condisiont (SFC)
1. Verify the defibrillator is ON. 2. Press the µA button on the safety
analyzer.
3. Press the F1/EARTH button on the safety analyzer and set the safety
analyzer controls as follows:
Neutral Open
Polarity Normal/Reverse
Current Mode AC only
Note: Pause briefly between switching polarity to prevent damage to the
defibrillator.
4. Verify the DUT AC Mains LED is OFF.
5. Verify the defibrillator AC Mains LED is OFF.
6. Verify the measured current is between 15 µA and 1000 µA (120 and 240
Vac).
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
Direct Equipment Leakage and Direct Applied Part Leakage Test Setup
PIP – Direct Equipment Leakage Test – Single Fault Condition (SFC)
To perform the Direct Equipment Leakage and Direct Applied Part Leakage tests
in accordance with IEC 62353, set up the safety analyzer as follows:
Press the SETUP button on the safety analyzer Press the F4/MORE button
Press the F2/INSTRUMENT button Press the F1/STANDARD button Press the
UP/DOWN arrows to select the 62353 standard Press the DONE button
1. Verify the defibrillator is ON.
2. Press the µA button on the safety analyzer.
3. Press the F1/DIRECT EQUIPMENT button on the safety analyzer and set the
safety analyzer controls as follows:
Previous Page
Earth Open
Polarity Normal/Reverse
Current Mode AC only
4. Verify the DUT and defibrillator AC Mains LEDs are ON.
5. Verify the measured current is between 15 µA and 300 µA (120 Vac) or between 15 µA and 500 µA (240 Vac).
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
Performance Inspection Procedures
PIP Instructions
Direct Applied Part Leakage Test Setup
PIP – Direct Applied Part Leakage Test ECG
To set up the safety analyzer to measure Direct Applied Part Leakage: Press
the µA button on the safety analyzer Press the F4/MORE button Press the
UP/DOWN arrows to select the appropriate A.P. groups Press F1/SELECT then
F1/Direct A.P.
1. Verify the defibrillator is ON.
2. Press the LEFT/RIGHT arrows to select the RA lead, and set the safety
analyzer controls as follows:
Polarity Normal/Reverse
Lead
RA (ECG group from Fluke 1210 adapter)
Previous Page
Warning
Shock Hazard. During Direct Applied Part Leakage tests, high voltage is present on the safety analyzer electrode snaps. Do not touch snaps or device connections during these tests.
3. Press the TEST button to measure the Direct Applied Part Leakage current.
4. Verify the DUT and defibrillator AC Mains LEDs are ON.
5. Verify the measured current is between 2 µA and 45 µA(120 and 240 Vac). 3312981-000_B
Table of Contents Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
Direct Applied Part Leakage Test – Therapy
1. Verify the defibrillator is ON.
2. Press the LEFT/RIGHT arrows to select the LL and LA leads, and set the
safety analyzer controls as follows:
Note: The A.P. group should be set up to measure the combined leakage of the
LL and LA leads.
Polarity Normal/Reverse
Lead
LL-LA (Therapy group)
3. Press the TEST button to measure the Direct Applied Part Leakage current.
Warning
Shock Hazard. During Direct Applied Part Leakage tests, high voltage is
present on the safety analyzer electrode snaps. Do not touch snaps or device
connections during these tests.
4. Verify the DUT and defibrillator AC mains LEDs are ON.
5. Verify the measured current is between 2 µA and 5000 µA (120 and 240 Vac).
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP Instructions
Performance Inspection Procedures
Direct Applied Part Leakage Test – SpO2
Note: Execute this test if the LIFEPAK 20e defibrillator/monitor is equipped
with SpO2.
1. Verify the defibrillator is ON.
2. Press the LEFT/RIGHT arrows to select the RL lead and set the safety
analyzer controls as follows:
Polarity Normal/Reverse
Lead
RL (SpO2 group)
3. Press the TEST button to measure the Direct Applied Part Leakage current.
Warning
Shock Hazard. During Direct Applied Part Leakage tests, high voltage is
present on the safety analyzer electrode snaps. Do not touch snaps or device
connections during these tests.
4. Verify the DUT and defibrillator AC Mains LEDs are ON. 5. Verify the measured current is less than 5000 µA (120 and 240 Vac).
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
PIP – Leakage Current Test Limits
Performance Inspection Procedures
Leakage Current Test Limit Summary TABLE – IEC 62353 Leakage Test Limits
Leakage Test to be Performed
Test Conditions
Range at 120V, 60HZ Range at 240V, 50HZ
ECG Direct Applied Part
Normal Reversed
2 – 45 µA 2 – 45µA
2 – 45 µA 2 – 45µA
Therapy (Apex, Sternum) Direct Applied Part Normal Reversed
2 – 5000 µA 2 – 5000 µA
2 – 5000 µA 2 – 5000 µA
SPO2 Direct Applied Part
Normal Reversed
0 – 5000 µA 0 – 5000 µA
0 – 5000 µA 0 – 5000 µA
Direct Equipment Leakage
Normal, Open Earth Reversed, Open Earth
15 – 300 µA 15 – 300 µA
15 – 500 µA 15 – 500 µA
Protective Earth Resistance
Measured from CodeManagement Module ground stud
0.3 Ohms
0.3 Ohms
Protective Earth Resistance
TABLE – IEC 60601 Leakage Test Limits
Measured from LIFEPAK20/20e defibrillator ground stud
0.3 Ohms
0.3 Ohms
Leakage Test to be Performed
Test Conditions
Range at 120V, 60HZ Range at 240V,50HZ
Earth Leakage Current
Normal Condition
15 – 300 µA
Single Fault Condition 15 – 1000 µA
3312981-000_B
Previous Page
Table of Contents Section Contents
15 – 500 µA 15 – 1000 µA
Back
Index
Next Page
CodeManagement Module
Test and Calibration Procedure
TCP Scope and Applicability
Test and Calibration Procedure
This section contains the Test and Calibration Procedures (TCP). Perform this
section as necessary after replacing components or to correct out-
ofspecification conditions detected during the PIP.
Refer to the PIP/TCP Resource Requirements for a listing of the necessary
qualifications for PIP equipment, test equipment verification, workstation
power, and personnel.
Refer to the PIP/TCP Test Equipment Requirements for a listing of test
equipment, including specifications, required to complete the PIP.
Throughout the body of this TCP, the CodeManagement Module is referred to as
the Device Under Test or DUT.
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
CodeManagement Module
TCP Instructions
Test and Calibration Procedure
TCP CO2 Calibration
Note: Perform this test only if PIP- CO2 Calibration Check fails or if
required by the LIFENET Device Agent (LDA).
To calibrate CO2: 1. Connect the RS-232 (DB9 M/F) cable from the DUT to the
test computer.
2. Launch the Lifenet Device Agent (LDA) from the test computer. 3. Follow
instructions as prompted by the LDA to calibrate CO2.
CO2 calibration kit REF 21330-000239
Cut CO2 filter line REF 11996-000068 to ~18 inches in length
remove luer fitting
to front-panel DUT CO2 connector
Calibration gas REF 21300-001572
4. Verify that the CO2 calibration was successful.
DUT
Previous Page
Table of Contents
3312981-000_B
Section Contents
Back
Index
Next Page
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>