SAFETY INSTRUCTION

This chapter contains important safety instructions that you must follow when operating the KA3000 & KA6000 series and when keeping it in storage. Read the following before any operation to insure your safety and to keep the best condition for the KA3000 & KA6000 series.

Safety Symbols These safety symbols may appear in this manual or on the series.

General Introduction

Safety Guidelines

  • Do not block or obstruct the cooling fan vent opening.
  • Avoid severe impacts or rough handling that leads to damage.
  • Do not disassem static els goriare qualified as service personnel.

AC INPUT

  • AC Inut Voltage: 110V/120V/220V/230V, 50/60 Hz.
  • Connect the protective grounding conductor of the AC power cord to an earth ground, to avoid electrical shock.

Operation Environment

  • Location: Indoor, no direct sunlight, dust-free, almost non-conductive pollution ( note below).
  • Relative Humidity: <80%
  • Altitude: < 2000m
  • Temperature: 0-40°C

Storage environment

  • Location: Indoor
  • Relative Humidity: <70%
  • Temperature:- 10-70°C

FUSE

Model 110/120V 220/230V
KA3003D/P T4A/250V T2A/250V
KA3005D/P T5A/250V T3A/250V
KA301OD/P T10A/250V T5A/250V
KA6002D/P T5A/250V T3A/250V
KA6003D/P T5A/250V T3A/250V
KA6005D/P T10A/250V T5A/250V
  • To ensure fire protection, replace the fuse only with the specified type and rating.
  • Disconnect the power cord before fuse replacement.
  • Make sure the cause of fuse blowout is fixed before fuse replacement.

OVERVIEW

Models Introduction

Models V Meter A Meter USB Resolution
KA3003O 4digit 4digit NO 10mV/1 mA
KA3003P 4digit 4digit Yes 10mV/1 mA
KA3005O 4digit 4digit NO 10mV/1 mA
KA3005P 4digit 4digit Yes 10mV/1 mA
KA3010O 4digit 4digit NO 10mV/1 mA
KA3010P 4digit 4digit Yes 10mV/1 mA
KA6002/3D 4digit 4digit NO 10mV/1 mA
KA6002/3P 4digit 4digit Yes 10mV/1 mA
KA6005O 4digit 4digit NO 10mV/1 mA
KA6005P 4digit 4digit Yes 10mV/1 mA
Main Characteristics

Performance

  • Low noise: cooling fan controlled by heatsink temperature
  • Compact size, light weight

Operation

  • Constant voltage / constant
  • Corrent operation
  • Output On / Off control
  • Digital panel control
  • 4 pairs of panel setup save/recall
  • Conarse and fine voltage / current control
  • Software calibration
  • Beep output
  • Button lock function

Protection

  • Overload protection
  • Reverse polarity protection

Interfaces

  • USB/RS232 for remote control (only for KA3003P,КА3005Р, КА6002Р, КА6003Р, КА6005Р & КА3010Р)

Front Panel Overview

KORAD-KA3000-6000-Series-Digital-Control-And-Programmable-DC-Power-Supply-
fig- \(2\)

DISPLAY

Voltage level   Voltmeter displays the setup value of output voltage.
Current level  Displays the setup value of output current.

Status Indication

OVP

OVP is the indicator of overvoltage protection. When overvoltage function is turned on, ovp indicator lights on; when output voltage is higher than protection setup value due to unexpected conditions,output cuts off and OVP indicator flickers; Press the button OVP again, and the power supply recovers.

OCP
OCP is OCP indicator. When overcurrent function is turned on,   OGP indicator lights on.

C. C
C.C is constant current indicator. When power supply is in the mode of constant current, this light is on.

G. V

C.V is constant voltage indicator. When power supply is in the mode of constant voltage, this light is on.

OUT
OUT is output indicator. If light on, there is voltage output in the output terminal.

Storage Indication

Indication of saving and recalling 4 setups stored internally;
When LOCK indication turns on, the front panel button operation is locked.

Brief Introduction of Panel Buttons

Saves or recalls panel settings. For settings, 1 ~ 4 are available.
For save/recall details, see Page 13. For M5, press M4 and adjust the knob, then you will enter M5 mode.

KORAD-KA3000-6000-Series-Digital-Control-And-Programmable-DC-Power-Supply-
fig- \(8\)

KORAD-KA3000-6000-Series-Digital-Control-And-Programmable-DC-Power-Supply-
fig- \(9\)

KORAD-KA3000-6000-Series-Digital-Control-And-Programmable-DC-Power-Supply-
fig- \(10\)

REAR PANEL INTRODUCTION

KORAD-KA3000-6000-Series-Digital-Control-And-Programmable-DC-Power-Supply-
fig- \(11\)

USB dependent interface based on remote control order (see Page 14); only for KAXXXXP series, such as KA3003P and so on.

RS232 dependent interface based on remote control order (see Page 14); only for KAXXXXP series, such as KA3003P and so on.

The power cord socket mainly accepts AC values: 115V /230V, 50 / 60 Hz. Please refer to the fuse parameters on the back fuse label to replace the specified fuse.

Make sure the correct type of fuse is installed before power up.

OPERATION

Connect AC power cord

Connecting AC power cord and selecting the corresponding AC voltage according to the back label on voltage; then connecting the AC power cord to the socket on the back panel.

power on  Press the power switch to turn power on. The display initializes, showing the model of the machine and then showing the setting last time.

power off Press the power switch again to turn power off.

Output On / Off

Panel Operation
Press the output button to turn on output; and the button LED also turns on. Pressing the Output button again to turn off the output and the LED.

Note: If there are any of the following conditions, the output will automatically turn off.

  1. OVP turns on and there are unusual OVP on the output terminal.
  2. The setting voltage is more than that of the OVP.
  3. Recalling other setups from the memory.

Beep On / Off

  • Panel Operation
    By default, the beep sound is enabled. To turn off the beep, press the OCP(BEEP) button for 2 seconds. A beep comes out and the beep setting will be turned off.
    To enable the beep, press the OCP(BEEP) button again for 2 seconds.

Front Panel Lock

  • Panel Operation
    Press the LOCK button to lock the front panel button operation. To unlock, press the LOCK button for 2 seconds.

Output Setup

Panel Operation

  1. Connecting the load to the front port, red (+), black(-).

  2. Setting output voltage and current.
    Press the button Voltage / Current selection to switch voltage adjustment and current adjustment Adjusting voltage and current with Voltage / Current Adjustment knob. By default, the Voltage and Current knob work in the coarse mode. To activate the fine mode, press the buttons to selsct the coarse mode or the fine mode.

  3. Turning on the output and pressing the output button.
    The button LED turns on and displays CV or CC mode.

SAVE / RECALL SETUP

Save Setup

  • Background
    The front panel settongs can be stored into one of the four internal memories.

  • Contents
    The following list shows the setup contenes.

    • Fine / coarse knob editing mode

    • Beep on / off

    • Output voltage / current level
      The following settings are always saved as “off”

    • Output on / off

    • Front panellock on / off

  • Panel operation
    Press one of the 4 buttons (M1, M2, M3, M4) and the LED light turns on accordingly. After you adjust the value,it is savde automatically once it stops blinking.

Recall Setup

The front panel settings can be recalled from one of the four internal memories.

Recalls panel settings. For settings, 1 ~ 4 are available.

  • Indication of saving and recalling 4 setups stored internally;
  • Press one of the 1 ~ 4 Memory buttons, for example number 1.
  • The panel settings saved in memory NO. 1 are recalled. The LED M1 turns on.

Note:
When a setting is recalled, the output automatically turns off.

REMOTE CONTROL

Remote Control Setup
All the models with the suffix “P”, such as KA3003P,KA3005P, KA6002P, etc. can be connected to the PC through interfaces USB/RS232 on the back of the machine and controlled by the remote control.

COM setting
Set up the COM port inside the PC according to the following list.

  • Baud rate: 9600
  • Parity bit: None
  • Data bit: 8
  • Stop bit: 1
  • Data flow control: None

Functionality check Run this query command via the terminal application such as MTTTY (Multi-threaded TTY).

*IDN?
This should return the identification information:
Manufacturer, model name, software version.
KORAD KA3003P Vx.xx

REMOTE CONTROL PROCEDURES

Entering the Remote Control Mode

  1. Connect the USB cable.
  2. The power supply will automatically connect. After normal connection,there will be a tweet from the power supply itself.
  3. The panel buttons are locked, so the power supply can only controlled by the computer.
    NOTE: KORAD software must be installed first.

Exiting from the Remote Control Mode

  1. Close the remote control software.
  2. Disconnect USB from the back.
  3. The power supply disconnects; a tweet from the beep with the hint that the remote control is over.
  4. The power supply automatically comes into the panel control mode.

Specifications

Note: The specifications below are tested under the conditions of temperation 25°C+-5°C and the warm-up for 20 minutes.

Models KA3003D / P KA3005D / P KA301OD / P KA6003D / P KA6005D / P
Voltage Range

0-30V

|

0-30V

|

0-30V

|

0-60V

|

0-60V

Current Range|

0-3A

|

0-5A

|

0-1 OA

|

0-3A

|

0-5A

Load Regulation
Voltage Current| O. 01% +2mv

O. 1%+5mA

| O. 01% +2mv

O. 1% +10mA

| O. 01% +3mv O. 1 %+ 20mA| O. 01%+ 2mv

O. 1 % +5mA

| O. 01% +2mv

O. 1%+10mA

Line Regulation
Voltage Current| O. 01%+3mv

O. 1%+3mA

| O. 01%+3mv

O. 1%+3mA

| O. 01% +3mv

O. 1%+3mA

| O. 01%+3mv

O. 1 % +3mA

| O. 01% +3mv

O. 1% +3mA

Setup Resolution
Voltage Current| 10mV

1mA

| 10mV

1mA

| 10mV

1mA

| 10mV

1mA

| 10mV

1mA

Setup Accuracy (25°C + – 5°C )
Voltage Current| 0.5 %+20mV

0.5%+5mA

| 0.5%+20mV

0.5% +1OmA

| 0.5% +20mV

0.5% +20mA

| 0.5%+30mV

0.5% +5mA

| 0.5%+30mV

0.5%+1OmA

Ripple(20-20M)
Voltage Current| 1mvrms 3mArms| 2mvrms 3mArms| 2mvrms 5mArms| 1mvrms 3mArms| 1mvrms 3mArms
Temp. Coefficient
Voltage Current| 150ppm 150ppm| 150ppm 150ppm| 150ppm 150ppm| 150ppm 150ppm| 150ppm

150ppm

Read Back Resolution
Voltage Current| 1omv 1mA| 10mv 1mA| 1omv 1mA| 10mv 1mA| 10mv 1mA
Read Back Temp. Coefficient
Voltage Current| 150ppm 150ppm| 150ppm 150ppm| 150ppm 150ppm| 150ppm 150ppm| 150ppm 150ppm
Reaction Time

Voltage Rise Voltage Drop

| 1 OOmS

100mS

(10% Rated load)

| 1 OOmS

1 OOmS

(10% Rated load)

| 100mS

100mS

(10%  Rated load)

| 1 OOmS

1 OOmS

(10% Rated load)

| 1 OOmS

1 OOmS

(10% Rated load )

Interface (for programmable models only)
RS232, USB
Accessories
User manual 1, Power cord1, USB cable (for programmable models KA3003P, KA3005P, KA6002P, KA6003P, KA6005P & KA3010P)
Weight and Dimension
11Omm(W)172mm(H)305mm(D): KA6005x8kg & KA301Ox8.3kg

11Omm(W)165mm(H)265mm(D): KA3003x3. 7kg, KA3005x4.3kg, KA6002x4.2kg & KA6003x4 .55kg.

The proprietary information in this manual is protected by copyrights.
Any photocopies, reproductions or translation to another language are not allowed unless it is permitted by KORAD Technology officially.

And all rights are reserved.
The information in this manual is correct when printing. However,KORAD will continuously improve products and reserve the rights to change specifications, equipment, and maintenance procedures at any time without notice.
DONGGUAN KORAD TECHNOLOGY CO., LTD.

DONGGUAN KORAD TECHNOLOGY CO., LTD.
Address: F9, Building D2, Kechuang Center, Songshan Lake Intelligent Valley,No. 7 Yanhe North Road, Liaobu Town, Dongguan City, Guangdong Province.
Tel: +86-769-81111584
Fax: +86-769-81111804
E-mail: krissy@koradtechnology.com
www.koradtechnology.com
www.korad.com.cn

FAQ

Q1: The panel buttons don’t work when power on.
A1: The panel is locked. Press the button for over 2 seconds, and then the panel will unlock.

Pressing ON/OFF, there is no output when power on.

Current setup is 0.

Output voltage rises slowly when output button is on.

Current setup is too small.

Making OCP on and pressing output switch; and then the output is automatically shut off

Current protection value setup is too small. You could press output switch and then make OCP on.