ACCT MG2-50 Professional Digital Scale User Manual
- June 17, 2024
- ACCT
Table of Contents
MG2-50
USER MANUAL
PROFESSIONAL DIGITAL SCALE
MG2-50 Professional Digital Scale
Thanks for purchasing the ACCT Digital Milligram Scale!
This User Guide is designed to get you up and running quickly with your scale.
Scales areprecisioninstruments and require proper care and skill to get the
best results.
If you havequestions orneedsupport, please feel free to reach out our customer
support: acct_support@163.com
What’s In The Box
Key Descriptions
| – Turns the scale ON or OFF.
---|---
| – Changes the weighing unit.
| – Zero’s the scale. Also subtracts the weight of an empty container (see
Tare Function).
| – Customize the time of auto-off.
Weighing Your First Item
- Press the key to turn the scale on.
- Place the item on the center of the weighing platform.
- Take your reading from the display. You can press the key to convert between weighing units.
- The scale’s auto-off function will trun the scale off after 3 minutes of inactivity or you can press the key to turn the scale off.
Things To Avoid
Tare Function
The Tare function allows you to subtract the weight of a container so that the
display only shows the weight of the contents.Place an empty tray or
container on the platform.
Press the key to subtract the container’s weight.
Add items to the container to see their weight on the display.
NOTE: The weight of the container will subtract from the available
capacity and will leave a negative value on the display once the weight is
removed. Press again to return the display to zero.
Auto-Off
- Turn off the scale.
- Press and hold button, “0”/ “60”/ “120”/ “180” will be displayed repeatedly.
- Release button when the display shows the time you want. (if “0”is selected, the auto-off function will be disable.)
- The display will show “PASS”, then turn off automatically.
Accuracy
Your scale comes calibrated from the factory and most users wont need to
calibrate their scale for a long time.
You should check your scale’s accuracy periodically using calibration weights
and perform calibration if necessary.
Calibration
**** Calibration adjustment is permanent and should only be performed using accurate calibration weights that match the requirements for your scale (see the chart on back). Do not use coins or other household items to calibrate.
- Prepare the required weight for your scale. (see chart on back)
- Find a flat and level surface to perform calibration and let the scale adjust to room temperature.
- Press to turn the scale on and wait for the scale to stabilize.
- Make sure nothing is on the platform then press and hold , the display will show “CAL” first, then flash “20.00”.
- Place a 20g calibration weight on the center of the platform.
- The display will show “40.00”, then place another 20g weight calibration to the center of the platform.
- The display will show “pass” then return to normal weighing mode.
You may now turn the scale off or continue weighing.
Specifications
Max Capacity| 50g / 1.7637 oz / 50000 mg/ 1.6075 ozt / 32.150 dwt / 250.000 ct
/ 771.62 gn
---|---
Readability| 0.001 g / 0.0001 oz / 1mg 0.0001 ozt / 0.001 dwt / 0.005 ct /
0.02 gn
Weighing Units| g, oz, ozt, dwt, ct, gn, mg
Dimensions| Overall: 5.3×3.7×1.4in (LxWxH)
Platform: 1.8×1.8in (LxW)
Display: 1.26×0.52in (LxW)
Calibration Weight| 2x20g
Power| 2 x AAA Batteries
Warranty Warranty| 2 Year Limited Warranty