ACCT MG2-50 Professional Digital Scale User Manual

June 17, 2024
ACCT

ACCT logo MG2-50
USER MANUAL
PROFESSIONAL DIGITAL SCALE

MG2-50 Professional Digital Scale

ACCT MG2 50 Professional Digital Scale - Symbols

Thanks for purchasing the ACCT Digital Milligram Scale!
This User Guide is designed to get you up and running quickly with your scale. Scales areprecisioninstruments and require proper care and skill to get the best results.
If you havequestions orneedsupport, please feel free to reach out our customer support: acct_support@163.com

What’s In The Box

ACCT MG2 50 Professional Digital Scale - The Box

Key Descriptions

| – Turns the scale ON or OFF.
---|---
|  – Changes the weighing unit.
|  – Zero’s the scale. Also subtracts the weight of an empty container (see Tare Function).
|  – Customize the time of auto-off.

Weighing Your First Item

  1. Press the key to turn the scale on.
  2. Place the item on the center of the weighing platform.
  3. Take your reading from the display. You can press the key to convert between weighing units.
  4. The scale’s auto-off function will trun the scale off after 3 minutes of inactivity or you can press the key to turn the scale off.

Things To Avoid

ACCT MG2 50 Professional Digital Scale - Avoid

Tare Function

The Tare function allows you to subtract the weight of a container so that the display only shows the weight of the contents.ACCT MG2 50 Professional
Digital Scale - FunctionPlace an empty tray or container on the platform.
Press the key to subtract the container’s weight.
Add items to the container to see their weight on the display.
NOTE: The weight of the container will subtract from the available capacity and will leave a negative value on the display once the weight is removed. Press again to return the display to zero.

Auto-Off

  1. Turn off the scale.
  2. Press and hold button, “0”/ “60”/ “120”/ “180” will be displayed repeatedly.
  3. Release button when the display shows the time you want. (if “0”is selected, the auto-off function will be disable.)
  4. The display will show “PASS”, then turn off automatically.

Accuracy

Your scale comes calibrated from the factory and most users wont need to calibrate their scale for a long time.
You should check your scale’s accuracy periodically using calibration weights and perform calibration if necessary.

Calibration

**** Calibration adjustment is permanent and should only be performed using accurate calibration weights that match the requirements for your scale (see the chart on back). Do not use coins or other household items to calibrate.

  1. Prepare the required weight for your scale. (see chart on back)
  2. Find a flat and level surface to perform calibration and let the scale adjust to room temperature.
  3. Press to turn the scale on and wait for the scale to stabilize.
  4. Make sure nothing is on the platform then press and hold , the display will show “CAL” first, then flash “20.00”.
  5. Place a 20g calibration weight on the center of the platform.
  6. The display will show “40.00”, then place another 20g weight calibration to the center of the platform.
  7. The display will show “pass” then return to normal weighing mode.
    You may now turn the scale off or continue weighing.

Specifications

Max Capacity| 50g / 1.7637 oz / 50000 mg/ 1.6075 ozt / 32.150 dwt / 250.000 ct / 771.62 gn
---|---
Readability| 0.001 g / 0.0001 oz / 1mg 0.0001 ozt / 0.001 dwt / 0.005 ct / 0.02 gn
Weighing Units| g, oz, ozt, dwt, ct, gn, mg
Dimensions| Overall: 5.3×3.7×1.4in (LxWxH)
Platform: 1.8×1.8in (LxW)
Display: 1.26×0.52in (LxW)
Calibration Weight| 2x20g
Power| 2 x AAA Batteries
Warranty Warranty| 2  Year Limited Warranty