OWEN MUMFORD Rapport Classic ED Pump Instruction Manual
- June 4, 2024
- OWEN MUMFORD
Table of Contents
- OWEN MUMFORD Rapport Classic ED Pump
- Troubleshooting Guide
- How the Rapport™ Classic Vacuum Therapy Device (V.T.D.) Works
- Choosing the Correct Constriction Ring
- Choosing a Transfer Sleeve
- Mounting the Constriction Ring
- Fitting the Pump and the Penile Tube Together
- Using the Pump to Create a Vacuum
- Cleaning Instructions and Care of your Rapport™ Classic V.T.D.
- Warranty
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
OWEN MUMFORD Rapport Classic ED Pump
Troubleshooting Guide
Problem | Likely cause | Remedy | Comments |
---|---|---|---|
Handle sticking | The vacuum pump has been submerged in water or immersed in | ||
running water | Allow at least 24 hours to dry. Do not try to force the handle. |
Please do not immerse the vacuum pump in water.
Clean as instructed
|
No erection| No vacuum because of air leak caused by poor seal against the
penis or seal between tube and pump| Use a smaller constriction ring, more
lubricating jelly or trim pubic hair| The problem is often associated
with inexperience, it can resolve itself as you become more adept at using the device
Partial erection| Not enough pumping strokes to create a full erection| Use a
larger seal sleeve| With continued use the problem often resolves itself
Rapid loss of erection (in less than 5 minutes)| Vacuum effect lost because of
air leak, poor seal against the penis, excessive pubic hair or constriction
ring too large| Use a smaller constriction ring or more lubricating jelly|
Experiment with varying sizes of constriction ring and / or seal sleeves
Delayed loss of erection (after 5 to 10 minutes)| Insufficient tension from
constriction rings| Use smaller or a combination of constriction rings| The
problem is often associated
with inexperience, it can resolve itself as you become more adept at using the device
Pulling the skin of the scrotum| Lubricant on scrotum, improper pumping
technique, sleeve too large| Remove lubricant from scrotum, use a smaller
sleeve, use the modified pumping technique (ask your physician), or angle
cylinder downward during pumping|
Discomfort or pain… during pumping| Pumping too rapidly| Pump slower and
continue to use the device| With continued use the problem often resolves
itself
from constriction ring| Constriction ring too small| Use a larger constriction
ring and continue to use the device| With continued use the problem usually
resolves itself
during intercourse| Inadequate lubrication or pressure on a sensitive area|
Apply more lubrication or change your intercourse position|
during ejaculation| Long period of abstinence, infection or inflammation of
the prostate gland| Continue to use the device| With continued use the problem
often resolves itself – if pain or discomfort persists, consult your physician
Redness, irritation or bruising| Pumping too rapidly| Pump slower or use fewer
strokes| Consult your physician if this problem persists
Penis feels cold
(to you or your partner)
| Constriction of blood flow to the penis or lubrication| Wash off lubricant
or warm lubricant before use|
Penis pivots on its base| Constriction of blood flow to the penis| If
manageable, apply the device after you have a partial erection|
Warnings
Do not submerge the vacuum pump in water or immerse in running water. Doing so
may cause the handle to stick and prevent the vacuum pump from working
correctly. If the vacuum pump has been submerged refer to the troubleshooting
guide.
Cautionary points on the use of vacuum therapy devices
- Use only the minimum amount of vacuum necessary to achieve an erection and increase the amount of vacuum pressure slowly. This will prevent bruising.
- If you experience any pain when you use the vacuum pump to create an erection, lower the vacuum pressure immediately by pressing the release valve.
- Never wear the constriction ring for more than 30 minutes and allow a minimum of 60 minutes between uses.
- Do not risk falling asleep while wearing a constriction ring.
- An instruction DVD is available, in addition to a video guide found at www.owenmumford.com, which shows a demonstration of the Rapport™ Classic V.T.D. being used to create an erection.
- Caution is advised when using a vacuum pump following the removal of a penile implant because of possible surgical scarring and subsequent impairment.
- Misuse of a vacuum pump may aggravate already existing medical conditions such as Peyronie’s disease, priapism, and urethral stricture.
- Use of the vacuum pump may bruise or cause rupture of the blood vessels immediately below the surface of the skin, which may produce hemorrhage or hematoma.
- If any of the above problems occur, discontinue the use of the vacuum pump and consult your doctor.
- The vacuum pump does not guard against pregnancy.
- We advise you to continue to use your usual form of contraception.
- Be sure to start by using a larger size constriction ring to ensure that the constriction on the penis is no more than necessary.
- If your erection does not stay firm and hard the constriction ring may be too large and not enough blood is staying in the penis. Use the next size smaller.
- Always consult your healthcare professional before using the Rapport™ Classic V.T.D.
Owen Mumford will not be liable for any misuse of this product.
How the Rapport™ Classic Vacuum Therapy Device (V.T.D.) Works
An erection is created by means of a vacuum, which draws blood into the penis. The erection is maintained by placing a constriction ring at the base of the penis, which ensures that the blood remains in the penis until the ring is removed.
The constriction ring should be snug and comfortable fit without being too tight. The use of the Rapport™ Classic V.T.D. is not intended to replace your doctor’s diagnosis of the underlying cause of your impotence, but to help achieve and maintain an erection. It is always best to discuss your individual problem with a doctor when using this treatment. Initially, it may take some practice to familiarise yourself with the Rapport™ Classic V.T.D., so please take the time to read this instruction book, and to practice the technique of creating and maintaining your erection, before you attempt sexual intercourse. The instruction DVD and online instruction video guide show how to put the Rapport™ Classic V.T.D. together and demonstrate how to create an erection.
The Rapport™ Classic V.T.D. is a complete system with everything you need for immediate use. Inside the discreet storage bag you will find:
- Five sizes of Rapport™ constriction rings. Five separately packaged and numbered rings, each a different size, from No. 4 the smallest up to No. 8 the largest. The centre ring is designed to encircle the base of the penis. The handles are for easy application and removal.
- An easy-to-use Vacuum Pump that has a specially angled handle to make creating the erection simple. As a safety measure the vacuum pressure is controlled by a valve that will not allow the pressure to rise above 360mm of Hg. The vacuum pressure can be released at any time by pressing the pressure release valve on the top of the pump.
- A loading cone to assist the mounting of the constriction rings onto the transfer sleeve.
- A penile tube in durable clear material, which attaches to the pump unit at one end and the transfer sleeve with the constriction ring at the other.
- Two transfer sleeves give a choice of internal diameter. Both are specially shaped to hold the constriction ring and make it easy to transfer it onto the base of the penis.
- A tube of lubricating jelly. This helps to create the vacuum seal and prevents the penis from adhering to the penile tube.
- A card ring sizing guide for easy selection of the correct size constriction ring can be found at the back of this book.
- Depending on which pack you have received, a How to Use DVD may be included. An online instruction video guide is available at www.owenmumford.com.
Choosing the Correct Constriction Ring
One of the most important parts of the Rapport™ Classic V.T.D. is the constriction ring. The ring must encircle the penis and fit comfortably tight. The constriction ring works by retaining the blood in the penis to maintain a firm erection. Using the card ring sizing guide provided, place your flaccid (non-erect) penis into the slots until you find one which fits the circumference of your penis. Each slot has a ring size marked above it which corresponds to the ring you require.
If the Rapport™ Classic V.T.D. gives you a good erection, and with the constriction ring on your penis, your erection decreases prematurely, you will need to try a ring one size smaller. If in use the constriction ring seems too tight, next time use one size larger to see if it will maintain your erection. Try different sizes until you are satisfied with their fit. There are five sizes to choose from.
Choosing a Transfer Sleeve
There are two transfer sleeves provided with two internal diameters. Depending on the size of your penis you should choose which of the two is likely to be the most suitable. Remember that your erect penis will need to be withdrawn through the transfer sleeve. The transfer sleeves are specially shaped so that the constriction ring can be easily loaded and held in the flat groove. The unique shape also allows easy passage of the constriction ring off the transfer sleeve and onto the base of your penis. Because the transfer sleeves are made of a soft material they help to make a good seal against your body to achieve the vacuum inside the penile tube which creates the erection.
Mounting the Constriction Ring
When you have selected the most suitable transfer sleeve place it firmly onto the clear end of the penile tube. (Fig. 1). Fit the loading cone onto the transfer sleeve and apply a little water-soluble lubricating jelly to the top of the cone. Holding the constriction ring by the outer handles, simply stretch and pull the ring down the loading cone until it slips onto the transfer sleeve. Make sure that the constriction ring is located in the flat groove at the center of the transfer sleeve. (Fig. 2).
The constriction ring is now mounted and ready for use. Before using the Rapport™ Classic V.T.D. make sure you use the water-soluble lubricating jelly to lubricate the end of the transfer sleeve. Also lightly lubricate the penis to prevent it from sticking to the inside of the penile tube during insertion or while using the pump. (Fig. 3).
Fitting the Pump and the Penile Tube Together
The pump head has two small pegs on each side above the O-ring; these should be positioned into the two grooves on the top of the penile tube. The two parts are then locked together with a twist so that the two pegs slot down and locate at the lower end of the groove. (Fig. 4). If these two parts are then locked together correctly it will ensure the unit is sealed tight enough to create a vacuum inside the penile tube. (Fig. 5).
Using the Pump to Create a Vacuum
- Insert the penis into the penile tube and apply slight pressure to seal the transfer sleeve against your body at the base of the penis, to ensure a vacuum will be created. (Fig. 6) It may be necessary to groom the hair at the base of the penis to ensure a vacuum is created. Make sure the constriction ring on the transfer sleeve is in the correct position with the handles to the top and bottom. Steady the tube with one hand to prevent any jerky movements and squeeze and release the pump handle with the other hand. This works the pump to create the vacuum necessary. (Fig. 7) You may find that creating the erection will be easier and more comfortable if you stand up to use the pump. You can always steady the tube by placing it on a flat surface such as a table. If you are unsteady when standing up, create an erection sitting down on the edge of a chair.
- Continue pumping slowly to create the erection. If you feel any pain stop pumping and press the pressure release button on the pump to eliminate the vacuum. Contact your doctor. Please take care not to pump too fast as this may bruise or cause a rupture of the blood vessels immediately below the surface of the skin.
- NOTE: For your safety, the vacuum pump is automatically released through the release valve at a pressure higher than 360mm of Hg.
- When the desired erection has been achieved, continue to hold the tube tightly against your body. Hold the tube with one hand, still pushing it against your body and with the fingers of the other hand, ease the constriction ring off the transfer sleeve and onto the base of your penis. Always position the constriction ring with the handles to the top and bottom while on the base of your penis. (Fig. 8). Because of the special shape of the transfer sleeve, when the constriction ring is eased out of the flat groove it will then slip easily off the transfer sleeve and onto your penis. Some vacuum pressure may be lost during this transfer, but you will still need to press the release valve to lower the vacuum pressure to enable you to remove the penile tube. (Fig. 9). If you find the placing of the constriction ring onto your penis difficult try releasing a little of the vacuum by pushing the penile tube away from your body slightly then rolling the constriction ring off the transfer sleeve.
- With a constriction ring in place, a lasting erection should now be maintained and sexual intercourse can be accomplished.
- NOTE: If a condom is to be worn, it should be put on as a last step. Do not wear a condom during the vacuum process. (Fig. 10).
- To remove the constriction ring after intercourse, simply grasp the outer handles to stretch the central ring. As the blood is released from the penis it will decrease in size and the ring may be easily removed.
Cleaning Instructions and Care of your Rapport™ Classic V.T.D.
-
The Constriction Rings
After each use, the constriction ring should be washed with mild soap in warm water, rinsed well, and dried thoroughly. Return the ring to its bag. Make sure you only use water-soluble lubricating jelly with the constriction rings. Your rings should last for longer if cared for properly. -
The Penile Tube and Transfer Sleeve
After each use, the penile tube and transfer sleeve should be washed thoroughly with mild soap in warm water. Rinse and dry well before replacing them with the carrying bag. -
The Vacuum Pump
Very little maintenance is needed for the vacuum pump, but we recommend that you occasionally wipe the vacuum pump with a damp cloth to remove any dirt or lubrication. Do not submerge the vacuum pump in water or immerse the vacuum pump under running water, as this may cause the handle to stick and it will not be possible to create a vacuum.
Please take care when wiping the underside of your vacuum pump as the nonreturn valve (three-pronged rubber valve) can be pushed into the piston chamber with only a slight force. If this valve is not in position, no vacuum will be achieved.
It will be necessary from time to time to lubricate the O-ring on the pump head, particularly if you are finding it difficult to attach or remove the pump head to and from the penile tube. After cleaning the pump head apply a little petroleum jelly to the O-ring (the colored band) making sure you smear the lubricant all the way around the O-ring. This will ensure a total seal between the pump head and the penile tube and maintain the vacuum inside the penile tube.
NOTE:
This lubrication of the O-ring with petroleum jelly is the ONLY use of
petroleum jelly. Do not allow petroleum jelly to come in contact with the
constriction rings as it could damage them.
Warranty
This warranty is offered to you as a purchaser (“you”) by the Owen Mumford entity whose details are set out on the relevant device packaging (“we” or “us”). You cannot transfer this warranty to anyone else. This warranty is in addition to and does not affect, your rights as a consumer or purchaser under the laws of your country.
This warranty applies for twelve months from the date of purchase (the “warranty period”).
If the device is or becomes materially defective during the warranty period, please notify us by providing details of the defect and we will provide you with a replacement device. Subject to the laws of your country, we shall have no further liability to you after we have provided you with a replacement device.
This warranty will not apply to defects that are attributable to:
- fair wear and tear;
- handling, storage or use of the device other than in accordance with our instructions, or improper handling, storage or use of the device; or
- damage or breakage caused by accident or by external causes beyond our control.
Owen Mumford will not be liable for any misuse of this product.
When you contact us, we may ask you for details of the batch number or other
form of identification of the device in order for us to further investigate the
defect. We may also ask you to return the device and any provision of a
replacement device is subject to us receiving the defective device from you.
Owen Mumford Ltd, Brimsdown Industrial Estate, Worcester Road, Chipping Norton, Oxfordshire OX7 5XP, United Kingdom T: +44 (0)1608 645555 F: +44 (0)1608 646546 E: info@owenmumford.com.
Aquagel® Lubricating Jelly
Ingredients – Sodium polyacrylate (carbomer), Glycerol, Hiquest (disodium salt
of 1-hydroxy ethane-1, 1 phosphonic acid; sequestrant), Sodium Nipasept
(parabens mix – a preservative), Purified water. Made by ECOLAB. ECOLAB,
Lotherton Way, Garforth, Leeds LS25 2JY, England
www.ecolab.co.uk.
Aquagel® is not a product or trademark owned by Owen Mumford. For more information please contact the relevant manufacturer.
Made in England by / Fabriqué en Angleterre par / Hergestellt in England von
HEAD OFFICE
Owen Mumford Ltd
Brook Hill, Woodstock Oxfordshire OX20 1TU, United Kingdom
T : +44 (0)1993 812021
E: info@owenmumford.com
UK / IRELAND SALES & MARKETING
T: +44 (0)1608 645555
E: info@owenmumford.com
NORTH AMERICA
T: +1 770-977-2226
E: info@owenmumfordinc.com
FRANCE
T: +33 (0)2 32 51 88 70
E: fr.info@owenmumford.com
GERMANY
T: +49 (0)6026/97750
E: de.info@owenmumford.com
MALAYSIA
T: +60 (0)7 553 9617
E: my.info@owenmumford.com
MEXICO
T: +52 (55) 7036 1534
E: mx.info@owenmumford.com
W: owenmumford.com.
© Copyright Owen Mumford 2018. Latest revision date is June 2018 SM/2000/11/0000/08.