ATOMTEX AT6130A Pocket Radiation Monitor Instruction Manual

June 13, 2024
ATOMTEX

ATOMTEX AT6130A Pocket Radiation Monitor

Compact devices intended for gamma and X-radiation ambient dose equivalent and ambient dose equivalent rate measurement.

Operating principle

The device operating principle is based on the process of count rate measurement of impulses, generated in a Geiger-Muller counter tube under the influence of X and gamma radiation. The count rate is converted automatically into measurable physical values throughout the range. An energy-compensating filter allows for correcting energy dependence of sensitivity efficiently in the entire energy range of photon radiation. Microprocessor-based unit is responsible for controlling the radiation monitor operating modes, calculations, storing and displaying measurement results, and for self- checking function.

Applications

  • Radiation protective measures in case of nuclear disasters
  • Civil protection
  • Radioecology
  • Fire-fighting service
  • Customs service
  • Dosimetric monitoring in manufacturing facilities, health care, and other institutions

Features

  • Low weight and small size
  • Automatic compensation of intrinsic detector background
  • Sound and visual alarm in case threshold level is exceeded for dose and dose rate
  • Rapid reaction to statistically significant change of dose rate (measurement process restart)
  • Field operation capability over a wide temperature range
  • In search mode, each registered gamma quantum is indicated by a sound signal
  • Up to 2000 measurement results can be stored in non-volatile memory with information about measurement date and time
  • Measurement results, current time, date, and battery life indicator are displayed on the matrix LCD screen
  • Headphones can be attached when working in a noisy environment
  • Bright white backlit LCD screen
  • Display in either Sv/h or in rem/h (configurable per request)

Specification

Ambient gamma and X radiation dose rate equivalent indication range

  • AT6130A 0.01 μSv/h – 10 mSv/h (or 1 μrem/h – 1 rem/h)*
  • AT6130D 0.01 μSv/h – 100 mSv/h (or 1 μrem/h – 10 rem/h)

Ambient gamma and X radiation dose rate equivalent measurement range

  • AT6130A 0.1 μSv/h – 10 mSv/h (or 10 μrem/h – 1 rem/h)*
  • AT6130D 0.1 μSv/h – 100 mSv/h (or 10 μrem/h – 10 rem/h)*

Ambient gamma and X radiation dose equivalent indication range

ATOMTEX-AT6130A-Pocket-Radiation-Monitor-fig-2

  • AT6130A 1 nSv – 100 mSv (or 0.1 μrem – 10 rem)*
  • AT6130D 1 nSv – 1 Sv (or 0.1 μrem – 100 rem)*

Ambient gamma and X radiation dose equivalent measurement range
AT6130A 0.1 μSv – 100 mSv (or 10 μrem – 10 rem)
AT6130D 0.1 μSv – 1 Sv (or 10 μrem – 100 rem)

Limits of tolerable intrinsic relative error ±20%

  • of dose rate measurement in the range from
  • 0.1 μSv/h to 10 mSv/h (from 10 μrem/h to 1 rem/h)

X and gamma radiation energy range 50 keV – 3 MeV

  • 137 -1 Typical sensitivity to Cs gamma radiation 2.8 cps/(μSv·h )
  • Response time for dose rate ≤ 7s
  • change from 1 to 10 μSv/h (accuracy error ≤ ±10%)
  • Energy dependence ±30% relative to 662 keV (137Cs)

Radiation overloading Radiation monitors can withstand a 100-fold rise in dose rate measurement upper range limit for 5 minutes with readings not lower than the maximum

  • Burn-up life ≥ 100 Sv
  • Continuous run time ≥ 500 h
  • Working temperature range -20°C to +55°C
  • Relative humidity with air temperature ≤ 95%
  • ≤ 35°C without condensation
  • Drop protection From ≤ 1.5 m to hard surface
  • Protection class IР57

Power supply 2 x AAA-size batteries (LR 03)

  • or 2 x AAA-size rechargeable
  • cells with a nominal voltage of 1.2 V
  • Overall dimensions, weight 110x60x38 mm, 0.25 kg

Units of measure are chosen during the ordering procedure and cannot be altered later Design and specifications are subject to change without notice

The radiation monitors comply with

  • GOST 27451-87, GOST 28271-89, GOST 17225-85,
  • Safety requirements of IEC 61010-1:2010,
  • EMC requirements of EN 55011:2009,
  • IEC 61000-4-2:2008, IEC 61000-4-3:2008