i-DT33P / i-DT33XW Indicator
Ethernet Option
Instruction Manual
INTRODUCTION
This Ethernet interface kit is for use with the Defender® 3000 i-DT33P and iDT33XW indicators.
CAUTION: THE INSTALLATION AND CONNECTIONS SHOULD ONLY BE PERFORMED BY AUTHORIZED AND QUALIFIED PERSONNEL, SUCH AS AN ELECTRICIAN.
KIT CONTENTS
- Cable
- Screws(3)
- Ethernet Cable
- Ethernet PC Board
- Watertight Seal (with One or Two Holes)
INTERFACE INSTALLATION
Install the Ethernet Interface kit to the indicator as the following.
WARNING: BEFORE INSTALLATION, POWER OFF THE INDICATOR AND REMOVE ALL EXTERNAL AC POWER CONNECTIONS.
i-DT33P
1. Remove the battery cover in the direction of the arrow, and take out dry batteries if used.
2. Slightly push the end of the charging cover (circled) in the direction of the arrow and take it out.
3. Remove the Ethernet sealing plug upwards.
4. Take out the cable supplied. Connect one side to the main board connector (circled), and pass the other side through the side hole.
5. Take out the Ethernet PC board supplied. Connect the cable’s another side to the connector on the Ethernet board (circled). Then insert the Ethernet board into the two card slots (circled) on both sides.
- Ethernet PC board
6. Take out the Ethernet cable supplied, and pass it through the hole where the sealing plug is removed, and then connect the Ethernet interface to the connector on the Ethernet board (circled).
7. Re-install the dry batteries if used and close the battery cover to finish installation.
i-DT33XW
1. Remove the six hex head bolts from the rear housing of the indicator.
2. Release the strain relief for option connector (circled) including the watertight seal in it on the bottom of the indicator and remove the plug at the end of it.
- Watertight Seal with One Hole
- Plug
3. Take out the cable and the Ethernet PC board supplied. Connect one side of the cable to the connector on the Ethernet PC board (circled) and the other to the indicator’s main board connector (circled).
4. Pin the upper right corner of the Ethernet PC board on the positioning column (circled). Make sure the back of the Ethernet PC board is facing up.
5. Screw the two screws (circled) on the Ethernet PC board to the indicator. Pass the Ethernet cable through the strain relief for option connector and attach the watertight seal to it, and then connect the Ethernet interface to the connector on the Ethernet board (circled).
Note: If you install RS232 and the Ethernet kit at the same time, please use the supplied watertight seal with two holes. Make sure to install the RS232 cable first before the connection of Ethernet kit. Pass the RS232 cable through one hole of the watertight seal and the Ethernet cable through another.
- Watertight Seal with Two Holes
6. Tighten the strain relief for option connector with the watertight seal in it (circled).
7. Close the housing and re-install the bolts to finish installation.
SETUP
Upon installation, the indicator will recognize the Ethernet interface kit and add related items to the menu.
Ethernet Connection
If DHCP is set to be ON , the IP share or router will automatically assign
IP Address. If the DHCP is set to be OFF , users need to setup the IP
Address, Subnet Mask, Gateway, Primary DNS and Secondary DNS. The Interface
has a preset IP Address. The Port Number is always 9761 and cannot be
changed.
Note: After the setup of the IP Address, Gateway, Preferred DNS and Alternate DNS, please reboot the indicator to enable the setting.
OPERATION
Output
With the interface installed, the indicator will operate in one of the
following two ways:
-
Auto Print = OFF, Stable = ON or OFF
Press the Print button will send the display data.
If Stable is set to be ON , the indicator will wait for a stable reading before sending the stable data. -
Auto Print = ON, Stable = ON or OFF
The indicator will automatically send data based on the settings in the menu.
If Stable is set to be ON , only stable values will be sent.
Data sent from the Interface is in standard ASCII format terminated with a CRLF (carriage return-line feed). For the output format, please refer to the indicator’s instruction manual.
Input
The indicator will respond to various commands sent via the interface adapter.
Terminate commands with [CRLF].
Commands
For the full list of commands, please refer to the indicator’s instruction
manual.
CONFIGURATION
The Ethernet option board is connected to the i-DT33P and i-DT33XW indicator through the RS232 interface. The baud rate is 19200.
If the option board is not connected, the submenu will not display.
Ethernet | Reset | no , yes |
---|---|---|
IP Address | 169.254.1.1-9761 | |
Version | 1.03 | |
Alt Print CMD | A~Z, P | |
Alt Tare CMD | A~Z, T | |
Alt Zero CMD | A~Z, Z | |
End | \ | |
Print Setup | Reset | no , yes |
--- | --- | --- |
Assignment | Demand , Auto On Stable, Auto On Accept, Interval (seconds), |
MTContinuous, OH-Continuous, SICS
Stable Weight Only [Demand]
| Off , On (LFT is forced to be On)
Mode [Auto On Stable]
| Load , Load and Zero
Time [Interval(seconds)]
| 1 ~5000
Link to [OH-Continuous]
| Off , Custom 1, Custom2
Checksum [MT-Continuous]
| On, Off
Template| Custom 1 , Custom 2
End| \
Note: refers to not applicable in the table above.
MAC Address
The Medium Access Control (MAC) Address of the Ethernet option board is
not editable.
Port
The Ethernet option board has a TCP server, with the port of 9761
(unmodifiable).
DHCP
If DHCP is set to be ON , the IP share or router will automatically assign
IP Address. If the DHCP is set to be OFF , users need to setup the IP
Address, Subnet Mask, Gateway, Primary DNS and Secondary DNS. The Interface
has a preset IP Address. The Port Number is always 9761 and cannot be
changed.
Note: After the setup of the IP Address, Gateway, Preferred DNS and Alternate DNS, please reboot the indicator to enable the setting.
IP Address
IP is 169.254.1.1 by default.
Subnet Mask
Subnet mask is 255.255.255.0 by default.
Gateway
Gateway is 169.254.1.1 by default.
Primary DNS & Secondary DNS
DNS value is not needed when the system is connected to a LAN, while the DNS
value is assigned by the ISP when it is connected to the internet.
Alternative Command
The alternative command of Print, Tare, Zero can be set to ‘a’ ~ ‘z’ or ‘A’ ~
‘Z’.
Note:
- The submenu will be displayed after the installation of the Ethernet option board.
- To set the IP address and other parameters, you need to go to the ScaleMate software. To obtain it, contact an authorized OHAUS dealer.