RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor Instruction Manual

June 13, 2024
RadonTec

RadonTec-logo

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor

Product Information

The product is a Modbus device that allows for communication with other devices using the Modbus RTU protocol. It supports serial communication via a COM port and operates at a baud rate of 9600. The device has various settings and parameters that can be configured to suit the user’s requirements.

Product Usage Instructions

  1. Select “Modbus RTU” in the Options menu.

  2. Set the Serial device to “Com”, the Serial port to “4”, and the Baud to “9600”.

  3. Go to Settings and switch the Base Addr to “0”.

  4. Set the Modbus Mode to “RTU” and the Slave Addr to “1”.

  5. Important: Set the Scan Rate to “10000 ms” (every 10 seconds).
    Note that the recommended scan rate for the sensors’ radon value is 10 minutes, as it is updated every 10 minutes. A faster scan rate may disrupt the measurement process.

  6. Read the desired registers according to the provided examples or specific requirements.

  7. If you require further assistance or have any questions, please get in contact with us.

Note: The provided text contains some unrelated characters and information, which have been omitted from the instructions.

Real-time radon measurements
The AlphaTracer is a professional, highly innovative real-time radon sensor for fixed installation in smart buildings for connection to ventilation controls or other devices suitable for digital input signals. It measures Rn-222 activity concentration in indoor air, which is provided as a proportional digital output signal.

Sensor Specifications

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(1\)

General Device Specifications

Dimensions, Weight and Placement

Weight: 90 g
The sensor can be mounted in any conceivable position, as it is not sensitive to light or vibration.

Calibration
Recalibration is recommended every 3 years. If the sensor is exposed to very high concentrations (on average more than 1,000 Bq/m3), calibration is recommended every 2 years. A factory calibration is offered by the company RadonTec GmbH. The sensor itself does not require any maintenance.

Safety and warning notices

Please read the data sheet carefully before commissioning the AlphaTracer. Damage caused by non-observance of the instructions in the data sheet is excluded from any liability.

  • The radon sensor may only be installed by qualified personnel and must be installed in accordance with the data sheet.
  • To avoid damage to the sensor, check that it has been connected correctly before applying the supply voltage.
  • When unpacking, it must be checked whether the correct device version has been delivered.
  • The AlphaTracer may only be used in the manner described in this operating manual. If the product is used in any other way, this may result in danger to the operator as well as destruction of the sensor.
  • No technical modifications may be made to the sensor. The sensor housing should only be cleaned with a damp cloth. Do not use scouring agents or cleaning agents containing solvents. Furthermore, the sensor must not be used if the ambient conditions are not within the limits specified in the data sheet.
  • The sensor must not be used in an explosive atmosphere.
  • Failure to observe the safety instructions may result in damage to the sensor and injury to the user.

Power and Connectivity Options

  • Power 5 V (7 – 24 V DC, 16 mA) via IST 1.25 mm 5 Pin
  • Communication 2-wire modbus interface 7 – 24 V, GND

There are several ways to connect the sensor. As soon as the sensor is supplied with voltage, the measurement starts.

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(3\)

  1. DC jack specifications (5.5 mm x 2.5 mm), please use a DC plug 5.5 mm x 2.5 mm. Vcc = 7 – 24 V
  2. Terminal block is for serial communication, B-
  3. Terminal block is for serial communication, A+
  4. Terminal block is used for the shield
  5. Terminal block is used for the power supply Vcc, please provide +7 – 24 V
  6. Terminal block is used for the GND
  7. Stereo jack socket, is not supplied in the Modus version

The terminal block has following specifications: Pitch: 3,5/3,81 mm, Poles: 3P and 2P, Screw: M2

  • Please supply the AlphaTracer only one time with voltage!

Modbus

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(4\)

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(5\)

As default the sensor is using slave address 1. The radon value can be found in register 20.

Example 1 – QModMaster
You can use the open source program QModMaster, which you can download here: https://sourceforge.net/projects/qmodmaster/
Just check which Com Port your USB to R$485 adapter is using. In our case it’s COM11.

Select Modbus RTU… in the Options menu.

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(7\)

Set Serial device to Com, Serial port to 4 and Baud to 9600 and go to Settings and switch the Base Addr to 0.

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(8\)

Set the Modus Mode to RTU and Slave Addr to 1. Important, put the Scan Rate to 10000 ms (which means every 10 seconds). Normally we recommend to set the scan rate to 10 minutes, because the sensors radon value is updated every 10 minutes. A faster scan rate just disturbs the sensors radon measurement process.

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(9\)

In this example, we read 7 registers.

  1. Radon value live
  2.  Radon value 24h
  3. Radon value long since start
  4. Radon entries since start

Define Function Code Read Holding Registers (0x03) and Start Address 0, Number of Registers 7 and the Data Format Dec. Connect and activate Scan (so that the tool automatically updates the information every 10 seconds). Now you can see the sensor address in register 0.

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(10\)

How to change the slave id address?
If you want to use more than one sensor on the cable, then you need to give each sensor its own address. By default each sensor is on the Modbus slave id

You can change the slave id as follows:

  1. Select Function Code Write Single Register (0x06)
  2. Now you can write for example the slave address to 7
  3. Press the button Read/Write

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(11\)

If you now try to read out the input register again, you will get an error message because the tool cannot reach the sensor on Slave Addr 1, that means that the change of the slave address has worked.

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(12\)

Let’s change the Slave Addr to 7 and the Start Address to 20 and test again.

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(13\)

As a result the connection works and we get the radon live value on address 20.

Example 2 – Rilheva Tool
There is a free version called Rilheva Modus Poll that you can download here: https://www.rilheva.com/rilheva-modbus-poll-desktop-edition

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(14\)

Comparative measurement

RadonTec Modbus 485 AlphaTracer Office Patented Radon Sensor FIG
\(15\)

In this example, we have induced (under laboratory conditions) a very rapid increase in the Radon-222 activity concentration from 30 to over 2,500 Ba/m?. Both AlphaTracer variants show, only delayed in time, the same total exposure as the reference instrument AlphaGuard (approx. 15,000 Euro acquisition cost). Thus, the AlphaTracer Office reaches the measurement curve of the reference instrument with a time delay of approx. 30 min, the AlphaTracer Industry with a delay of approx. 100 min due to its 1P54 class. A very fast increase or decrease of the radon concentration leads to a delayed reproduction of the concentration change for radon measuring instruments with a denser housing and consequently slower air entry, but the correct measurement of the total exposure is not falsified or impaired by this.

Get in contact with us!

Website: radontec.de
Email: info@radontec.de
RadonTec GmbH • Hauptstraße 5 • 89426 Wittislingen

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals