OHAUS ION-100A Static Ionizer Instruction Manual
- June 12, 2024
- Ohaus
Table of Contents
Authorised OHAUS Australia Dealer
www.techadv.com.au
sales@techadv.com.au Ionizer 100A
Instruction Manual
INTRODUCTION
This manual contains installation, operation and maintenance instructions for
the Ionizer 100A.
When using the ionizer, basic safety procedures must be followed. Read with
attention all instructions before using the ionizer.
1.1 Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show
safety issues and warnings. Ignoring the safety notes may lead to personal
injury, damage to the equipment, malfunctions and false results.
Signal Words
CAUTION Indicates a hazardous situation with low risk, resulting in
damage to the device or the property or in loss of data, or injuries if not
avoided.
Attention: Indicates important information about the product
Warning Symbols
General hazard
** Direct current
1.2 Safety Precautions
CAUTION:** READ ALL SAFETY WARNINGS BEFORE INSTALLING, MAKING CONNECTIONS,
OR SERVICING THIS EQUIPMENT. FAILURE TO COMPLY WITH THESE WARNINGS COULD
RESULT IN PERSONAL INJURY AND/OR PROPERTY DAMAGE. RETAIN ALL INSTRUCTIONS FOR
FUTURE REFERENCE.
- The equipment is supplied with a grounded power cable, which must be connected to a compatible grounded power receptacle.
- Do not position the equipment such that it is difficult to disconnect the power cord from the power receptacle.
- Make sure that the power cord does not pose a potential obstacle or tripping hazard.
- Operate the equipment only under ambient conditions specified in these instructions.
- The equipment is for indoor use only. Use the equipment only in dry locations.
- The equipment should only be operated by properly trained personnel.
- Turn the equipment off when not in use.
- Disconnect the equipment from the power supply during cleaning or maintenance.
- Do not touch the electrode pins while the equipment is turned on.
1.3 Features
The Ionizer 100A generates bipolar ions continuously by DC corona discharge
separately from positive and negative discharge electrodes and directs the
ionized air onto the charged body to eliminate static electricity. The
generated ions are well-balanced in polarity and can eliminate static
electricity regardless of the polarity of the charged body.
Static electricity
Generally, insulators such as plastics are prone to be charged with static
electricity when the ambient relative humidity is below 45%RH. This may affect
weighing and cause a measurement error of several milligrams. The Ionizer 100A
can eliminate static electricity very effectively.
INSTALLATION
2.1 Package Contents
- Ionizer
- Instruction Manual
- AC Adapter
2.2 Installing Components
Follow the steps below to install the ionizer.
Step 1. Place the ionizer on the table next to the balance as shown below. The
distance between the ionizer and the balance should be more than 20 mm. Step
2. Adjust the height and angle of the ionizer by loosening the two lock knobs
on each side of the ionizer and moving it to the optimal position.
Step 3. Tighten the two lock knobs.
Step 4. Connect the AC adapter to a power outlet.
OPERATION
3.1 Overview of Controls
Item | Description |
---|---|
1 | Power connection |
2 | Power switch |
3 | Power indicator lamp |
4 | Electrode pins |
3.2 Operating the Ionizer
Step 1. Turn on the ionizer by pressing the power switch.
Step 2. Place the sample in front of the ionizer in the working area as shown
below. The distance between the sample and the ionizer should be around
10~40cm.
MAINTENANCE
CAUTION: Turn Off the ionizer with the switch and disconnect the AC
adapter before starting any maintenance work.
Do not allow liquids to enter the housing.
4.1 Cleaning
Clean the electrode pins using alcohol and cotton swabs as shown below.
Clean the housing using a soft cloth, dampened with water and a mild
detergent.
Attention: Do not use harsh chemicals to clean the balance, as the finish
may be damaged.
4.2 Replacing the Electrode Pins
Using needle nose pliers, grip the electrode pin and pull it out of the
socket. Gently push the new electrode pin into the socket until it is fully
inserted in the socket. Replacement electrode pins can be ordered from OHAUS,
please see below for information. Access ories:
SP Electrode pins, ION-100A | PN: 30133391 |
---|
TECHNICAL DATA
5.1 Specifications
Model| ION-100A (Europe: P/N 30095929, Americas: P/N 30130302, Asia-Pacific:
P/N 30130303)
---|---
Ion Greneration Method| DC Corona Discharge
Ionic concentration| >1*106 EA/cm3 (10cm distance)
Effective Static Removal Range| 100 to 400mm from the outlet
Static Elimination Time (approx.)| 1s/10cm, 5s/20cm (from ±1000V to ±100V)
Ozone Concentration| <0.05 ppm
Electrode Probes| Tungsten(Φ)2.5mm/Length: 17.8mm) (Durability: 15,000 hours)
Fan Speed| Fan Free
Power Input|
12V ****0.5A
Ambient conditions| Indoor use only
Altitude| Up to 2000 m
Specified Temperature Range| 0°C to 40°C
Operating Temperature Range| 5°C to 40°C / 41°F to 104°F
Humidity| maximum relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50% relative humidity at 40°C
Mains supply voltage fluctuations| up to ±10% of the nominal voltage
Installation category| II
Pollution degree| 2
Net Weight| 0.5kg/1.1Ib
Shipping Dimension (cm)| (W x H x L) 282115 / 11 x 8.27 x 5.9 in
Shipping Weight| 1.2kg/2.6Ib
5.2 Dimensions
COMPLIANCE
Compliance to the following standards is indicated by the corresponding mark on the product.
Mark | Standard |
---|---|
**** | This product complies with the applicable harmonized standards of EU |
Directives 2011/65/EU (RoHS), 2014/30/EU (EMC) and 2014/35/EU (LVD). The EU
Declaration of Conformity is available online
atwww.ohaus.com/ce.
| EN 61326-1
Disposal
** In conformance with the European Directive 2002/96/EC on Waste
Electrical and Electronic Equipment (WEEE) this device may not be disposed of
in domestic waste.
This also applies to countries outside the EU, per their specific
requirements. Please dispose of this product in accordance with local
regulations at the collecting point specified for electrical and electronic
equipment. If you have any questions, please contact the responsible authority
or the distributor from which you purchased this device. Should this device be
passed on to other parties (for private or professional use), the content of
this regulation must also be related.
For disposal instructions in Europe, refer to
www.ohaus.com/weee Thank you for your
contribution to environmental protection.
FCC Note
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures: Reorient or relocate the receiving antenna,
increase the separation between the equipment and receiver or consult the
dealer or an experienced radio/TV technician for help.
Industry Canada Note
This Class B digital apparatus complies with Canadian ICES-001.
ISO 9001 Registration**
In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to
ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the
OHAUS quality management system is compliant with the ISO 9001 standard’s
requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to
the ISO 9001:2008 standard.
OHAUS Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054 USA
Tel: +1 973 377 9000
Fax: +1 973 944 7177
With offices worldwide
www.ohaus.com P/N 30093331I © 2016 OHAUS
Corporation, all rights reserved
Printed in China
Documents / Resources
|
OHAUS ION-100A Static
Ionizer
[pdf] Instruction Manual
ION-100A Static Ionizer, ION-100A, Static Ionizer, Ionizer
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>