Ohaus 3000 Ranger Count 2nd Scale Interface Instruction Manual

June 12, 2024
Ohaus

Ohaus 3000 Ranger Count 2nd Scale Interface

Product Information

The Second Scale Interface Kit is designed for use with the Ohaus product Ranger Count 3000. It allows you to connect an additional platform to the Ranger Count 3000.

Kit Contents

  • 2nd Scale PC Board
  • Screws (4)
  • Pressure Button
  • Instruction Manual

Interface Installation

  1. Remove the red interface options cover and break off the plastic blocker from the options cover.
  2. Install the Interface PCB by pushing it through the pins and securing it with the 4 screws.
  3. Connect a cable to the interface.
  4. Place the cord in the groove and put the red interface option cover back on.

Product Usage Instructions

Setup

  1. Connect the main platform to the 2nd weighing platform.
  2. In the SETUP menu, switch on the Dual Scale (D.Scale) menu item.
  3. Set the CAP2 and Grad2 values in the SETUP menu.
  4. Perform a calibration by pressing and holding the key to switch between the main platform and the secondary platform.

Note:  Make sure to perform a calibration of the scale before using the weighing platform. Also, ensure that the weighing unit is set to kg or lb when using the high capacity platform. If the display shows error8.6, long press the [Units] key to switch the unit to kg or lb.

Accessories
For a complete listing of Ohaus printers and other accessories, please contact Ohaus Corporation or visit their website at www.Ohaus.com.

Compliance
This accessory has been tested and complies with the approvals listed in the applicable Instruction Manual.

Disposal
OHAUS Corporation 7 Campus Drive Parsippany, NJ 07054, USA  Tel: +1 973 377 9000Fax: +1 973 944 7177 With offices worldwide Website: www.ohaus.com

2nd Scale Interface

INTRODUCTION

This Second Scale Interface Kit is for use with the Ohaus product Ranger Count 3000. Through this kit you can connect additional platform with Ranger Count 3000.

KIT CONTENTS

  • 2nd Scale PC Board
  • Screws (4)
  • Pressure Button
  • Instruction Manual

INTERFACE INSTALLATION

Install the Interface module on the underside of the balance as shown.

  1. Remove the red interface options cover. Break off the plastic blocker from the option cover.Ohaus-3000-Ranger-Count-2nd-Scale-Interface-fig- \(1\)
  2. Install the Interface PCB. Push it through the pins and secure it with the 4 screws. Ohaus-3000-Ranger-Count-2nd-Scale-Interface-fig- \(2\)
  3. Connect a cable.
  4. Place the cord in the groove and put the red interface option cover back on.

SETUP

After the main platform is connected to the 2nd weighing platform, switch on the Dual Scale (D.Scale) menu item in the SETUP menu (see menu structure below).

Then set the CAP2 and Grad2 in the SETUP menu and do a calibration. Press and hold the key to switch between the main platform and the secondary platform.

CAL SETUP

  • Zero2: Reset
  • Span1: D.Scale
  • Span2: Cap2
  • Lin1 : Grad2
  • Lin2: Pwr. Un
  • GEO: Zero
  • End: Opt
    • A.Tare
    • Bp. Opt
    • Bp. Sig
    • Bp. Key
    • Accum
    • End

Note : Make sure to perform a calibration of the scale before using the 2nd weighing platform. Make sure the weighing unit is in kg or lb when using high capacity platform. (If display shows error8.6, long press [Units] key to switch unit to kg or lb.)

ACCESSORIES

For a complete listing of Ohaus printers and other accessories, contact Ohaus Corporation or visit www.Ohaus.com

COMPLIANCE
This accessory has been tested and complies with the approvals listed in the applicable Instruction Manual.

Disposal

In conformance with the European Directive 2002/96 EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. For disposal instructions in Europe, refer to www.ohaus.com/weee.

OHAUS Corporation 7 Campus Drive Parsippany, NJ 07054, USA Tel : +1 973 377 9000 Fax : +1 973 944 7177 With offices worldwidewww.ohaus.com

P/N 30064481B © 2013 Ohaus Corporation, all rights reserved Printed in China

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals