DETOUR TORUS Stent Graft System Instruction Manual
- June 13, 2024
- DETOUR
Table of Contents
DETOUR TORUS Stent Graft System
Product Information
The DETOURTM System is a medical device used for percutaneous revascularization in patients with symptomatic femoropopliteal lesions. It consists of two main components:
- x TORUSTM Stent Graft System: This component includes the o TORUSTM Stent Graft and the o TORUSTM Stent Graft Delivery System.
- ENDOCROSSTM Device: This device is also part of the DETOURTM System.
Figure 1 shows the TORUS Stent Graft, while Figure 2 depicts the TORUS Stent Graft Delivery System. Figure 3 displays the ENDOCROSS Device with the RX GW Port.
Intended Use/Indications
The DETOUR System is intended for use in patients with symptomatic
femoropopliteal lesions ranging from 200mm to 460mm in length. It is
specifically designed for patients with chronic total occlusions (100mm to
425mm) or diffuse stenosis exceeding 70%. The DETOUR System is recommended for
individuals who may not be suitable candidates for surgical or alternative
endovascular treatments. However, it is important to note that the DETOUR
System and its components should not be used in the coronary and cerebral
vasculature.
Contraindications
The DETOUR System should not be used in patients who have a distal common
femoral artery (CFA).
Product Usage Instructions
- Carefully read all instructions provided in the user manual before using the DETOURTM System.
- Observe all warnings and precautions mentioned throughout the instructions to avoid complications.
- Ensure that the patient meets the indications for use, including having symptomatic femoropopliteal lesions within the specified length range and characteristics.
- Do not use the DETOUR System or any of its components in the coronary and cerebral vasculature.
- Avoid using the DETOUR System in patients with a distal common femoral artery (CFA).
Carefully read all instructions prior to use. Observe all warnings and precautions noted throughout these instructions. Failure to do so may result in complications.
Device Description
The DETOUR™ System is comprised of two main components:
- TORUS™ Stent Graft System; comprised of the
- TORUS™ Stent Graft
- TORUS™ Stent Graft Delivery System
- ENDOCROSS™ Device
The procedure using the DETOUR System is described in Section 1.
TORUS Stent Graft System
The implantable TORUS Stent Graft (Figure 1) is a flexible, self-expanding
composite structure made of a Nitinol (NiTi) wire frame encapsulated in an
Expanded Polytetrafluoroethylene (ePTFE) film and Fluorinated Ethylene
Propylene (FEP). The length of FEP is identical across all stent lengths,
while the length of ePTFE increases with stent length. The TORUS Stent Graft
is pre-loaded onto the TORUS Stent Graft Delivery System.
The TORUS Stent Graft Delivery System (Figure 2) is an 8-French (Fr) system. It is compatible with a 0.035” guidewire and has a 135cm working length. The handle of the delivery system consists of an internal pulley mechanism activated by turning an external knob. The handle also features fluid flush ports for the inner lumen and guidewire lumen. The TORUS Stent Graft Delivery System uses an outer sheath to maintain the TORUS Stent Graft implant in a compressed state. Once at the target site, the user can slide the outer sheath proximally (toward the handle of the device or toward the user) by turning the knob in the direction of the arrow to expose the self-expanding TORUS Stent Graft. The TORUS Stent Graft Delivery System has Platinum-Iridium (PtIr) radiopaque markers on both the proximal and distal ends of the TORUS Stent Graft landing zone (part of the inner shaft), and a marker band on the outer sheath to allow visualization of the position of the inner and outer sheath during deployment.
ENDOCROSS Device
The ENDOCROSS Device (Figure 3) is a dual guidewire delivery tool that uses a
0.025” Nitinol needle with a 15mm throw. The needle exits the delivery tool at
an angle of approximately 45° to the ENDOCROSS Device shaft. The ENDOCROSS
Device is an 8 Fr compatible device with a 133cm working length with dual
0.014” guidewire ports, a rapid exchange (RX) guidewire port, and a needle
guidewire port. The RX guidewire port is a back-loaded, rapid-exchange design
used for initial device placement. The needle guidewire port is the central
lumen, and the lumen exits through the needle and is used to deliver
guidewire(s) to the desired location. The ENDOCROSS Device also incorporates
an intra-luminal stabilizer and a PtIr alloy marker band used to support and
guide needle deployment, respectively. The ENDOCROSS Device features are
controlled using the outer handle and the button on the ENDOCROSS Device
handle.
The outer handle controls spring loading, stabilizer deployment, and needle activation. The user moves the outer handle proximal to distal, then rotates the handle counterclockwise to load the spring, deploy the stabilizer, and activate the needle for deployment in a single motion. The ENDOCROSS Device shaft is keyed to ensure that the needle deploys in the same orientation as the marker band. Subsequent depression of the handle button deploys the needle in the direction indicated by the marker band.
Intended Use/Indications
The DETOUR System is indicated for use for percutaneous revascularization in patients with symptomatic femoropopliteal lesions from 200mm to 460mm in length with chronic total occlusions (100mm to 425mm) or diffuse stenosis >70% who may be considered suboptimal candidates for surgical or alternative endovascular treatments. The DETOUR System, or any of its components, is not for use in the coronary and cerebral vasculature.
Contraindications
Do not use the DETOUR System in patients with:
- A distal common femoral artery (CFA) <7mm in diameter.
- An increased risk of deep vein thrombosis (DVT), such as patients with a recent history of DVT, thrombophilia, and disseminated malignancy.
- Untreated flow-limiting aortoiliac occlusive disease.
- Lack of patent single vessel tibial runoff to ankle.
- Known coagulopathy, bleeding diathesis, or thrombocytopenia that cannot be medically managed.
- Known hypersensitivities, allergies, or contraindications to any of the following:
- Nitinol
- Antiplatelet, anticoagulant, or thrombolytic therapy
- PTFE
- Contrast media that cannot otherwise be medically managed
- Aspirin
- Heparin
Sizing Tables
Measure the diameter of the reference vessel (i.e., the diameter of the normal
vessel immediately proximal and distal to the lesion) and the length of the
target lesion. Refer to the tables below for recommended sizing.
Table 1: Vessel Diameter
Labeled Device Diameter (mm)| Reference Vessel Diameter
(mm)
| Available Device Nominal Lengths (mm)| Recommended Balloon Diameter for
Post-Dilation (mm)
---|---|---|---
5.5| 4.5-5.5| 200| 5.5
6.0| 5.6-6.0| 100, 150, 200| 6.0
6.7| 6.1-6.7| 100, 150, 200| 7.0
Table 2: Recommended Sizing Chart
Lesion Length (mm)
| Number of Stent Grafts**| Stent Graft 1 Nominal Length (mm)|
Stent Graft 2 Nominal Length (mm)| Stent Graft 3 Nominal Length (mm)
---|---|---|---|---
200-220| 2| 150| 150| N/A
230-270| 2| 200| 150| N/A
280-320| 2| 200| 200| N/A
330-410| 3| 200| 200| 150
420-460| 3| 200| 200| 200
60mm of stent graft overlap is required
Note:** In the procedure using the DETOUR System, the femoral vein is used
as a conduit for a series of overlapping TORUS Stent Grafts that provides a
femoral-popliteal bypass. As accounted for in the sizing tables, the stent
grafts are overlapped by at least 60mm, and each end of the overall bypass is
fixated within the arterial wall by at least 30mm. Therefore, foreshortening
is not impactful when the TORUS Stent Graft is implanted as part of the
procedure using the DETOUR System.
Warnings
- In the DETOUR2 Clinical Study, venous event rates were higher in females than males. This should be considered in patient selection and post-procedure management.
- For single use only. Do not reuse, reprocess, or re-sterilize. Reuse, reprocessing, or re-sterilization may compromise the structural integrity of the device and/or lead to device failure, which in turn, may result in patient injury, illness, or death. Reuse, reprocessing, or re-sterilization may also create a risk of contamination of the device and/or cause patient infection or cross-infection, including, but not limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may lead to injury, illness, or death of the patient.
- Do not use the DETOUR System if the package is opened or damaged. Prior to use, perform a thorough examination of the pouch materials and seals to ensure there is no visually detectable damage.
- Do not use the DETOUR System beyond the stated use-by date on the package label.
- Inspect the DETOUR System prior to use. Do not use if the device appears to be bent, kinked, or damaged in any way. The ENDOCROSS Device should be test fired prior to use in the patient to ensure proper functionality. Do not use if the device does not function properly when tested, or if the device appears to be bent, kinked, or damaged in any way because vessel damage and/or inability to advance or withdraw the device may occur.
- Do not place finger or other body parts near distal tip of the ENDOCROSS Device prior to deploying the needle.
- Do not rotate the ENDOCROSS Device while stabilizer is deployed.
- Place hand and fingers on the inner handle (not on the guidewire port area) of ENDOCROSS Device during needle deployment to avoid pinching gloves during needle deployment.
- Do not advance the guidewire significantly through the needle tip prior to needle deployment. Doing so may damage the device and/or guidewire tip.
- Ensure the stabilizer and needle are retracted prior to removing the ENDOCROSS Device.
- Do not deploy the stabilizer prior to advancing through the proximal anastomosis.
- Do not exert excessive force while advancing the ENDOCROSS Device through the anastomosis. Ensure tension is applied to the guidewire at both access sites and gently advance the ENDOCROSS Device until femoral vein (FV) access is confirmed fluoroscopically. Re-dilate the anastomosis if significant resistance is encountered.
- If the ENDOCROSS Device needle extends through both walls of the target artery, partially retract the needle using the outer handle and advance guidewire until the guidewire can be advanced to gain arterial access.
- Do not exert excessive force while advancing the TORUS Stent Graft System through the anastomosis. Gently advance the TORUS Stent Graft System until distal arterial access is confirmed fluoroscopically. Re-dilate the anastomosis if significant resistance is encountered.
- When deploying the proximal stent graft please ensure that the proximal edge (exposed end) of the stent graft is 1-2mm above the ostium of the superficial femoral artery.
- Do not dilate outside the margins of the deployed stent graft(s).
- Do not cannulate or puncture the TORUS Stent Graft. Cannulating or puncturing the stent graft may result in damage to the ePTFE lining and/or the NiTi frame, resulting in compromised performance or failure of the stent graft.
- Do not cut the stent graft or attempt to surgically implant the stent graft. The stent graft should only be placed and deployed using the TORUS Stent Graft System on which it is pre-loaded.
- Inadvertent, partial, or failed deployment or migration of the stent graft may require surgical intervention.
- If strong resistance is felt during guidewire tracking, determine the cause of the resistance before proceeding further. If the cause cannot be determined, withdraw the ENDOCROSS Device.
Precautions
- The DETOUR System is intended for use by physicians who have received appropriate training.
- The DETOUR System must be used with visual guidance (e.g., fluoroscopy) by physicians skilled in percutaneous, endovascular techniques in a fully equipped catheterization or angiographic laboratory or a surgical suite.
- Store the DETOUR System in a cool, dry environment and away from direct sunlight.
- Do not expose the DETOUR System to organic solvents (e.g., alcohol).
- Consider recent DVT history in the target limb in case planning and post-procedure anticoagulation as use of the DETOUR System uses a bypass conduit within the femoral vein.
- Antiplatelet medication should be initiated prior to placement of the stent graft. Effective dual antiplatelet therapy should be maintained following the procedure at a dosage deemed appropriate by the physician.
- A knee prosthesis may prevent successful and accurate ENDOCROSS Device needle deployment.
- To maximize wire differentiation, it is recommended to use a 180cm long 0.014” guidewire through the ENDOCROSS Device RX guidewire lumen.
- When conducting the procedure using the DETOUR System, the TORUS Stent Graft should only be used in patients with a patent femoral vein > 10mm in diameter, or with a duplicate or accessory femoral vein.
- Do not deploy the stent graft prior to insertion. The stent graft is self-expanding and cannot be re-loaded on the delivery system.
- Do not use excessive force when manipulating the device through an introducer. Excessive force may damage the device.
- When removing the delivery system avoid displacing the stent graft by removing the delivery system under fluoroscopy.
MRI Safety Information
MRI Safety Information
A person with a TORUS Stent Graft in an overlapped configuration may be safely scanned anywhere in the body at 1.5T or 3.0T under the following conditions. Failure to follow these conditions may result in injury
Parameter| Condition
Device Name| Overlapping TORUS Stent Graft single lengths up to 200mm and
overlapping lengths up to 480mm
Static Magnetic Field Strength (B0)| 1.5T and 3.0T
MR Scanner| Cylindrical
B0 Field Orientation| Horizontal
Maximum Spatial Field Gradient| 5000 Gauss/cm or less
RF Excitation| Circularly Polarized (CP)
RF Transmit Coil Type| Integrated Whole Body Transmit Coil
Operating Mode| Normal Operating Mode
RF Conditions| Maximum Whole-body SAR: 2W/kg
Scan Duration| 15 mins of scanning
Scan Regions| Any landmark is acceptable
Temperature Rise| 1.5T: 2.9ºC after 15 minutes of continuous scanning.
3.0T: 2.3ºC after 15 minutes of continuous scanning Cooling due to blood flow inside the covered stent and perfusion in the vascular bed surrounding the covered
stent was included in the assessment of in-vivo temperature rise.
Image Artifact| The presence of overlapping TORUS Stent Grafts may produce an image artifact of 0.6cm. Some manipulation of scan parameters may be needed to compensate for the artifact.
Potential Adverse Events and Complications
Below is a list of the potential adverse effects (e.g., complications)
associated with the use of the DETOUR System.
- Access vessel (arterial/venous) occlusion
- Myocardial infarction
- Amputation
- Pain (insertion site, leg and/or foot)
- Aneurysm or pseudoaneurysm
- Peripheral ischemia
- Arteriovenous (AV) fistula
- Renal insufficiency or failure secondary
- Bleeding complications to contrast medium
- Death
- Shock
- Device or deployment
- Side branch vessel occlusion malfunction/failure
- Stenosis or occlusion
- Drug reactions to antiplatelet agents or
- Stroke or transient ischemic attack contrast medium
- Thrombosis
- Edema
- Vessel wall trauma (dissection,
- Embolism (peripheral or pulmonary) perforation, or rupture)
- Fever in absence of infection
- Vessel spasm
- Hemorrhage or hematoma
- Venous flow disruption (deep vein
- Hypotension/hypertension
- Infection local or systemic including bacteremia or septicemia
- Worsening claudication
- Malposition
- Migration
Special Populations
- The DETOUR System has not been studied in women who are pregnant.
- The DETOUR System has not been studied in pediatric patients (< 18 years of age).
Clinical Studies
DETOUR® Endovascular Technique for long OcclUsive fem-pop Revascularization –
2 (NCT03119233)
The purpose of the DETOUR2 study was to establish a reasonable assurance of
safety and effectiveness of the procedure with the DETOUR System for
percutaneous revascularization in patients with symptomatic femoropopliteal
lesions from 200mm to 460mm in length in the US, Latvia, and Germany under IDE
G170083. Overall, 220 subjects (including roll-in patients) were treated with
the investigational device.
- Study Endpoints
The primary safety endpoint was freedom from major adverse events (MAEs) through 30 days post-procedure. Major Adverse Events include death, clinically driven target lesion revascularization (CD-TLR), amputation of the treated limb, occlusive-symptomatic deep vein thrombosis (DVT), pulmonary embolism (PE), or procedure-related bleeding requiring any transfusion of packed red blood cells or surgery. A performance goal for freedom from 30-day MAE of 84.0% was established for this endpoint. The safety performance goal was based on an aggregate of published trial data as described by VIVA Physicians Inc. (VPI) and adjusted to reflect the greater risk associated with the DETOUR2 study population. The study device was considered to have achieved the safety objective if the lower limit of the one-sided lower 97.5% confidence limit based on the exact method is greater than 84%.
The primary effectiveness endpoint was patency at 12 months post procedure.
Patency is defined as the absence of CD-TLR and absence of recurrent target
lesion diameter stenosis >50% by imaging (e.g., duplex ultrasound peak
systolic velocity ratio [PSVR] of >2.5 within the stent or immediately 1cm
above or below the treated segment). If both duplex ultrasound and angiography
were available, angiography took precedence. The primary effectiveness
endpoint of 12-month patency was evaluated by comparing the proportion of
successful subjects to a literature-derived performance goal of 60.4%.
The analyzed patient cohorts are defined as:
- Roll-ins: The first two subjects at each U.S. site were considered roll-in subjects. Roll-in subjects were pre-identified. Not all sites enrolled roll-in subjects.
- Intention to Treat (ITT): All subjects who received the intervention and were enrolled in the IDE Study (excluding roll-ins). The ITT Cohort was the primary analysis set to determine if the primary safety endpoint was met in the study.
- Modified ITT (MITT): Only those subjects/lesions where a DETOUR System: TORUS Stent Graft was implanted. The MITT Cohort is a subset of the ITT Cohort. The MITT Cohort is the primary analysis set to determine if the primary effectiveness endpoint was met in the study.
- Per Protocol (PP): Subset of the MITT group which excludes all subjects with major protocol deviations (e.g., violations of eligibility criteria) and missing the data required to evaluate the primary effectiveness endpoint.
Of the 220 subjects who received the ENDOCROSS Device, 18 were considered roll-in subjects and the remaining 202 (91.8%) are considered part of the ITT Cohort. Two subjects from the ITT group did not receive the TORUS Stent Graft, thus 200 patients were part of the MITT Cohort. Of the MITT Cohort, 197 (89.5%) were considered part of the Per Protocol Cohort.
Accountability of PMA Cohorts
At the time of database lock, 220 patients enrolled in the DETOUR2 study at 36
sites. 85.5% (188) patients are available for analysis for the 12-month post-
operative visit. Overall, 220 subjects were treated with the ENDOCROSS Device.
Figure 4 depicts the accountability of subjects at the study follow-up time
points. Of the 220 subjects enrolled, 83.6% (184) patients are available for
analysis for the 12-month post-operative visit. Eighteen (18) subjects in the
ITT group exited before the 12-month window: 2 enrolled but did not receive
the ENDOCROSS device, 7 subject withdrawals, 2 physician withdrawal, 5 deaths,
and 2 lost-to-follow-up.
Methods
In eligible patients, the ENDOCROSS device was used to deliver a guidewire
from the arterial segment proximal to the beginning of the target lesion,
through the femoral vein and back into the artery distal to the chronic total
occlusion or diseased segment. TORUS Stent Grafts of appropriate dimension
were selected. Stent grafts were then deployed in series, starting distally,
until the bypass was complete. Follow-up visits occur at 1month, 6-months,
12-months, 24-months, and 36-months post-procedure with arterial and venous
duplex-ultrasound assessment at each follow-up visit. Stent-graft x-ray also
occurred at the 12-month follow-up visit. An independent Clinical Events
Committee (CEC) was used to review and adjudicate primary and secondary safety
endpoints, including major adverse events. A Data Safety Monitoring Board
(DSMB) reviewed safety data. All angiograms, duplex ultrasound studies and
x-rays were submitted to the central Imaging Core Lab for analysis.
Peri, Intra and Post-Procedural Medication Regimens for the DETOUR2 Study
For study subjects, Dual Antiplatelet Therapy (DAPT) was required for the
duration of the DETOUR2 study (3 years), starting at least 24 hours prior to
the index procedure, or a loading dose during the index procedure. The minimum
loading doses required were 75mg of ASA, and 300-600 mg of Clopidogrel if not
on long-term therapy. It was recommended that the patients remain on ASA
indefinitely following completion of the study.
Table 3. Summary of Recommended Peri, Intra and Post-Procedural
Medication Regimens for the DETOUR2 Study
Medication| Peri-Procedure (<24 hours of index procedure)| Intra-
Procedure| Post-Procedure
---|---|---|---
Antiplatelet| Aspirin (ASA)| Loading dose of 300-325 mg within 24 hours
prior to procedure, if not on long-term aspirin therapy| N/A| 73-325 mg per
day, indefinitely
Clopidogrel (or similar antiplatelet agent or alternative agent, per operator
discretion)
| Loading dose of 75-300 mg within 24 hours prior to procedure, if not on
long-term clopidogrel (or similar) therapy| N/A| Clopidogrel 75 mg per day for
3 years (or per prescribing dose if a similar or alternative antiplatelet
agent is used)
Anticoagulation| IV Heparin / Bivalirudin (or other thrombin inhibitor)|
N/A| Maintain anticoagulation per hospital standard of care (minimum ACT >250
sec recommended)
| Anticoagulation administered at dose prescribed per institutional standard, at
operator discretion
Baseline Characteristics
Specific demographics and baseline characteristics for the DETOUR2 Study ITT
population are presented in Table 4. Medical history and health status are
provided in Table 5. Sub-group analysis of the ITT population is provided in
the Safety and Effectiveness Results sections.
Table 4: Baseline Demographics – ITT Population
Variable | All Subjects |
---|---|
Age at Consent (years) | 68.9 ± 9.38 (202) |
69.0 (47, 88)
Body Mass Index (BMI)| 28.74 ± 5.028 (202)
28.50 (18.3, 45.0)
Sex| —
Female| 26.2% (53/202)
Male| 73.8% (149/202)
Race1| —
American Indian or Alaska Native| 0.5% (1/202)
Asian| 0.0% (0/202)
Black or African American| 8.9% (18/202)
Native Hawaiian or Other Pacific Islander| 0.0% (0/202)
White| 86.6% (175/202)
Other| 4.5% (9/202)
Ethnicity| —
Hispanic or Latino| 4.6% (9/196)
Not Hispanic or Latino| 79.6% (156/196)
Not Reported| 15.8% (31/196)
Patient demographics and medical history are summarized in Table 4 and Table
5. The mean age was 68.9 ±9.4 years and 26.2% were female. 86.6% of the
patients were white. Comorbidities included coronary artery disease (87.6%),
hypertension (87.6%), diabetes (34.7%), prior history of smoking (91.1%), and
renal insufficiency (10.9%). The mean ABI was 0.61 ±0.22 at baseline. Most
patients (77.7%) were Rutherford Clinical Category (RCC) 3. The remainder of
the patients (22.3%) were RCC 4 and RCC 5.
Table 5: Medical History and Health Status – ITT Population
Variable | All Subjects |
---|---|
Renal Insufficiency (with or without intervention) | 10.9% (22/202) |
History of Smoking | — |
Current/Previous | 91.1% (184/202) |
Never | 8.9% (18/202) |
History of Peripheral Venous Disease (DVT, Thrombophlebitis, etc.) | 0.0% |
(0/202)
Peripheral Arterial Disease| 98.0% (198/202)
Previous Peripheral Intervention| 60.5% (121/200)
Previous Peripheral Vascular Surgery| 16.8% (34/202)
Hypertension| 87.6% (177/202)
Diabetes Mellitus| 34.7% (70/202)
Type 1| 1.4% (1/70)
Type 2| 98.6% (69/70)
Hypercholesterolemia| 73.2% (145/198)
Hyperlipidemia| 73.3% (140/191)
Coronary Artery Disease| 46.0% (92/200)
Congestive Heart Failure| 12.4% (25/201)
NYHA I| 16.0% (4/25)
NYHA II| 48.0% (12/25)
NYHA N.A.| 36.0% (9/25)
History of Myocardial Infarction| 21.9% (44/201)
History of Cerebrovascular Disease| 13.9% (28/201)
Stroke| 60.7% (17/28)
TIA| 21.4% (6/28)
Other| 17.9% (5/28)
Target Limb Ankle-Brachial Index (ABI)| 0.61 ± 0.22 (193)
0.60 (0.00, 1.55)
Contralateral Limb Ankle-Brachial Index (ABI)| 0.82 ± 0.26 (189)
0.82 (0.00, 1.63)
Target Limb Toe-Brachial Index (TBI)| 0.45 ± 0.19 (58)
0.43 (0.00, 0.83)
Contralateral Limb Toe-Brachial Index (TBI)| 0.59 ± 0.23 (57)
0.60 (0.00, 1.10)
Rutherford Clinical Category| —
Grade 3| 77.7% (157/202)
Grade 4| 17.8% (36/202)
Grade 5| 4.5% (9/202)
Table 6 summarizes the baseline lesion characteristics determined by the Core
Lab. Target lesions were relatively equally distributed between the right and
left limb. Pre-procedure chronic occlusion (CTO; 100% stenosis) was observed
in 96.0% (194/202) of subjects and diffuse stenosis (>70% stenosis) was
observed in 97.0% (196/202) of subjects. The mean lesion length was 327.14 ±
61.38mm. The mean CTO length was 217.31 ± 85.98mm. The calcification grade for
lesions was predominantly severe (70.4%, 126/179) followed by none/mild
(29.1%, 52/179) and then moderate (0.6%, 1/179).
Two vessel run-off to the foot was observed in 69.0% (129/187) of subjects.
There were no subjects who had zero run-off vessels to the foot. The popliteal
artery was involved in approximately 10% of lesions, and approximately 44% of
subjects had below-the-knee arterial disease as well.
Table 6: Core Lab Baseline Lesion Characteristics – ITT Population
Variable | All Subjects |
---|---|
Total Occlusion (100% stenosis) | 96.0% (194/202) |
Diffuse Stenosis (>70% stenosis) | 97.0% (196/202) |
In-Stent Restenosis | 17.3% (35/202) |
Lesion Length (Normal to Normal, mm) | 327.14 ± 61.38 (196) |
328.15 (194.6, 520.3)
Calcified Length (mm)| 64.12 ± 77.50 (178)
41.10 (0.0, 415.7)
CTO Length (mm)| 217.31 ± 85.98 (191)
232.50 (0.0, 436.1)
Calcification| —
None/Mild| 29.1% (52/179)
Moderate| 0.6% (1/179)
Severe| 70.4% (126/179)
Of the 202 ITT subjects, the device was implanted in 200. Two (2) subjects did
not receive the device: one (1) case was aborted because there was a
complication with a percutaneous transluminal angioplasty (PTA) balloon which
extended the procedure time, and one (1) case was aborted because venous
access was difficult.
Table 7 summarizes site reported procedure characteristics. The majority of
subjects were treated under conscious sedation (40.6%; 82/202) or local
anesthesia (28.2%; 57/202). Mean procedure time was 181.4 ± 90.55 minutes,
with a mean fluoroscopy time of 46.4 ± 19.51 minutes. An average of 208.1 ml ±
111.74 of contrast was used per subject. Estimated blood loss was 50 ± 57.50
ml per subject.
Table 7: Site Reported Procedure Characteristics – ITT
Variable | All Subjects |
---|---|
Type of Anesthesia Used | — |
General | 12.4% (25/202) |
Local | 28.2% (57/202) |
Epidural/Spinal | 5.9% (12/202) |
Conscious Sedation | 40.6% (82/202) |
Other | 12.9% (26/202) |
Estimated Blood Loss (ml) | 50.0 ± 57.50 (201) |
30.0 (0, 400)
Contrast Volume Used (ml)| 208.1 ± 111.74 (199)
180.0 (50, 900)
Fluoroscopy Time (min)| 46.4 ± 19.51 (199)
42.0 (12, 122)
Total Procedure Time (min)| 181.4 ± 90.55 (202)
163.0 (55, 495)
Post-Procedure Core Lab Angiography Data
Core laboratory-reported post-procedure angiographic data are presented in
Table 8.
Table 8: Post-Procedure Core Lab Angiography Data
Post-Procedure Characteristics | All Subjects |
---|---|
Distance of Proximal TORUS Edge from Superficial Femoral Artery Ostium (mm) |
5.91 ± 5.672 (183)
4.40 (0.0, 50.8)
Distance from Distal TORUS Edge to Tibial Plateau (mm)| 42.55 ± 36.22 (183)
28.00 (2.4, 150.5)
Proximal Intra-Arterial TORUS Length| 48.10 ± 17.68 (187)
45.90 (13.2, 125.6)
Distal Intra-Arterial TORUS Length| 59.44 ± 20.82 (187)
54.40 (19.2, 129.4)
Overlap Lengths (mm)| —
Proximal (or only)| 72.00 ± 20.54 (169)
69.00 (28.5, 139.1)
Middle| 43.60 ± 23.11 (3)
35.80 (25.4, 69.6)
Distal| 66.14 ± 16.54 (134)
63.15 (28.1, 179.9)
Run-Off Vessels to the Foot| —
0| 0.0% (0/183)
1| 15.3% (28/183)
2| 67.2% (123/183)
3| 17.5% (32/183)
Technical success in the MITT population, defined as successful delivery of the investigational devices to the identified area and removal of delivery system was 100%, in the 200 subjects. Procedural success in the MITT population, defined as successful delivery of the investigational device to the identified area and removal of the delivery system in the absence of in- hospital MAEs, was 98.5% (197/200). Three (3) patients had MAEs prior to discharge: Two (2) subjects were reported to have major bleeding, and one (1) subject was reported to have had a symptomatic DVT.
Safety Results
The primary safety endpoint was freedom from a composite of Major Adverse
Events (MAE) at 30 days (Table 9) as adjudicated by the CEC. MAEs were defined
as any of the following: all-cause mortality, CD-TLR, amputation of the
treated limb, occlusive-symptomatic deep vein thrombosis, pulmonary embolism,
or procedure-related bleeding. There were 15 total 30-day MAEs in 14 subjects:
3 CD-TLR, 5 DVT and 7 major bleeding events. Freedom from MAE at 30 days was
93.0% (185/199), with a lower exact one-sided 97.5% confidence interval of
88.5% compared to the PG 84%. Thus, the primary safety endpoint was met. The
composite primary safety endpoint is summarized in Table 9. There were 199
subjects of the 202 subjects in the ITT group evaluated for the primary safety
endpoint at 30 days. The 30-day MAE rate for the roll-in cohort (n=18) was
comparable to the pivotal [ITT] cohort (5.6% in roll-in versus 7.0% in pivotal
ITT). Major adverse events are reported in Table 10. Follow-up is on-going
through 36 months.
Table 9: Freedom from Major Adverse Event (MAE) at 30 Days – ITT
Population
All Subjects | Lower 97.5% Exact Confidence Limit | |
---|---|---|
Freedom from MAE1 | 93.0% (183/199)2 |
15 total 30-day MAE events in 14 subjects:
· 3 CD-TLR
· 5 DVT
· 7 Major Bleeding Events
| ****
88.5% > 84% P
Table 10: Major Adverse Events Over Time: Kaplan Meier Estimates – ITT Population
Event | 30 Days | 6 Months | 12 Months |
---|---|---|---|
MAE | 7.0% [3.5%, 10.5%] | 13.6% [8.8%, 18.3%] | 19.8% [14.2%, 25.4%] |
Death Events1 | 0.0% [0.0%, 0.0%] | 0.5% [0.0%, 1.5%] | 2.6% [0.4%, 4.9%] |
CD-TLR | 1.5% [0.0%, 3.2%] | 6.6% [3.1%, 10.0%] | 12.3% [7.7%, 17.0%] |
Amputation of Treated Limb | 0.0% [0.0%, 0.0%] | 1.0% [0.0%, 2.4%] | 2.1% [0.1%, |
4.1%]
Deep Vein Thrombosis| 2.5% [0.3%, 4.7%]| 3.5% [1.0%, 6.1%]| 4/1% [1.3%, 6.8%]
Pulmonary Embolism| 0.0% [0.0%, 0.0%]| 0.0% [0.0%, 0.0%]| 0.0% [0.0%, 0.0%]
Procedure Related Bleeding| 3.5% [1.0%, 6.0%]| 3.5% [1.0%, 6.0%]| 3.5% [1.0%,
6.0%]
Serious Adverse Events that Occurred in the PMA Clinical Study:
Overall, 66.3% (134/202) of subjects had a serious adverse event (SAE) within
12 months; almost half were vascular disorders based on the System Organ
Category (SOC) classification. Of all SAEs, 24.3% (49/202) were device related
SAEs, as adjudicated by the CEC. The most common SOC for device-related SAEs
were Vascular Disorders, with 48/202 subjects (23.8%) experiencing at least
one SAE in the SOC.
Overall, 17.3% (35/202) procedure related SAEs were adjudicated by the CEC,
were observed through 12months. The most common SOC for procedure-related SAEs
was Vascular Disorders, with 21/202 subjects (10.4%) experiencing at least one
adverse event in this SOC.
Table 11: Serious Adverse Events – ITT Population
Adjudicated Event
| ****
Body System or Organ Class / Preferred Level Term
| Through 12 Months
%
(n/N)
[m]
---|---|---
All Serious Adverse Events
| ****
Any Event
| 66.3% (134/202) [355]
Vascular Disorders
| 48.5% (98/202) [167]
Serious Device Related Adverse
| ****
Any Event
| 24.3% (49/202) [67]
Event| ****
Vascular Disorders
| 23.8% (48/202) [59]
Serious Procedure Related
| ****
Any Event
| 17.3% (35/202) [57]
Adverse Event| ****
Vascular Disorders
| 10.4% (21/202) [26]
Secondary Safety Endpoints
Secondary safety endpoint Kaplan Meier Estimates were performed on the ITT
analysis set (Table 12). Major Adverse Limb Events (MALE) were reported at 30
days, 6 months, and 12 months, with a rate of 4.0%, 10.5%, and 15.7%,
respectively.
Major Bleeding was reported in eight (8) subjects, with nine (9) events
occurring within 30 days of the index procedure. Eight (8) of these events
were related to the procedure. Five (5) of these eight (8) procedural related
events were due to blood loss anemia, there were two (2) gastrointestinal
bleeds and one retroperitoneal hematoma.
Symptomatic DVT occurred in eight (8) subjects through 12 months with no new
symptomatic DVT in the subjects followed through 24 months and 36 months. Most
DVT (7/8) occurred within 6 months of the index procedure. There has been no
Pulmonary Embolism (PE) observed in the ITT Cohort at the time of this report.
There was one (1) perioperative Myocardial Infarction (MI) in the ITT
population, through 30 days (0.5%; 1/199). There were three (3) MIs through 12
months in the ITT cohort and seven (7) MIs total in patients followed through
36 months.
Table 12: Secondary Safety Endpoints Over Time: Kaplan Meier Estimates –
ITT Population
Event | 30 Days | 6 Months | 12 Months |
---|---|---|---|
Major Adverse Limb Events (MALE) | 4.0% [1.3%, 6.7%] | 10.5% [6.3%, 14.8%] |
15.7% [10.7%, 20.9%]
Major Bleeding| 4.0% [1.3%, 6.7%]| 6.0% [2.7%, 9.3%]| 6.6% [3.1%, 10.0%]
Event| 30 Days| 6 Months| 12 Months
---|---|---|---
Symptomatic Deep Vein Thrombosis (DVT)| 2.5% [0.3%, 4.7%]| 3.5% [1.0%, 6. %]|
4.1% [1.3%, 6.8%]
Pulmonary Embolism (PE)| 0.0% [0.0%, 0.0%]| 0.0% [0.0%, 0.0%]| 0.0% [0.0%,
0.0%]
Myocardial Infarction (MI)| 0.5% [0.0%, 1.5%]| 0.5% [0.0%, 1.5%]| 1.6% [0.0%,
3.3%]
Hematoma| 0.5% [0.0%, 2.8%]| N/A| N/A
Stent Thrombosis| 3.5% [1.0%, 6.0%]| 9.0% [5.1%, 13.0%]| 15.8% [10.7%, 20.9%]
Venous Outcomes
The definition of a DVT is any organized clot that is occluded within the deep
venous system and results in a lack of flow and a lack of compressibility of
the vein. This definition is distinct from catheter-related thrombosis or
protein/fibrin deposition that have typically been reported with the use of
indwelling central venous catheters and cardiac implantable electronic
devices, which are largely asymptomatic, do not occlude the vein, and can be
found in the vicinity of the device. Venous events are described below based
on CEC adverse event definitions and clinical relevance. The methodologies of
categorization of these events are not mutually exclusive.
Table 13 categorizes venous events based upon the CEC definitions of serious
and non-serious.
Serious DVTs are those that resulted in either death, life-threatening
illness/injury, permanent impairment of a body structure of a body function,
inpatient or prolonged hospitalization, or medical or surgical intervention to
prevent life-threatening illness or injury or permanent impairment.
Non-serious DVTs are those that did not result in death, life-threatening
illness/injury, permanent impairment of a body structure or a body function,
inpatient or prolonged hospitalization, or medical or surgical intervention to
prevent life-threatening illness or injury or permanent impairment. These may
include fibrin sheaths which did not occlude the vein, are largely
asymptomatic, and can be found in the vicinity of the device.
Table 13: Deep Vein Thrombosis (DVT) – ITT Population
Event
| 30 Days
% (95% CI)
{n/N}
[m]
| 6 Months
% (95% CI)
{n/N}
[m]
| 12 Months
% (95% CI)
{n/N}
[m]
---|---|---|---
Non-serious DVT
| 10.9% (7.0, 16.0)
{22/202) [22]
| 13.9% (9.4, 19.4)
{28/202} [30]
| 13.9% (9.4, 19.4)
{28/202} [30]
Non-serious Device Related DVT| 6.9% (3.8, 11.4)
{14/202} [14]
| 8.9% (5.4%, 13.7%)
{18/202} [18]
| 8.9% (5.4%, 13.7%)
{18/202} [18]
Non-serious Procedure Related DVT| 10.9% (7.0, 16.0)
{22/202} [22]
| 10.9% (7.0, 16.0)
{22/202} [22]
| 10.9% (7.0, 16.0)
{22/202} [22]
Serious DVT (or Endpoint DVT)| 2.5% (0.8, 5.7)
{5/202} [5]
| 4.0% (1.7, 7.7)
{8/202} [8]
| 4.0% (1.7, 7.7)
{8/202} [8]
Event
| 30 Days
% (95% CI)
{n/N}
[m]
| 6 Months
% (95% CI)
{n/N}
[m]
| 12 Months
% (95% CI)
{n/N}
[m]
---|---|---|---
Serious Device Related DVT| 1.5% (0.3, 4.3)
{3/202} [3]
| 2.5% (0.8, 5.7)
{5/202} [5]
| 2.5% (0.8, 5.7)
{5/202} [5]
Serious Procedure Related DVT| 2.5% (0.8, 5.7)
{5/202} [5]
| 3.0% (1.1, 6.4)
{6/202} [6]
| 3.0% (1.1, 6.4)
{6/202} [6]
All DVT Events (serious and non-serious, regardless
of relation to device or procedure)
| ****
27 events
| ****
38 events
| ****
38 events
Figure 5 below organizes the venous events by clinical relevance. The events were categorized as symptomatic DVT, which are venous events that are occlusive within a deep vein and associated with clinical symptoms, and those that are not. Symptomatic DVTs (or endpoint DVTs) were considered as part of the safety endpoint. Venous events that are not symptomatic are likely the result of fibrin deposition/sheath. Details on gender and treatment are also reported. There were a total of 38 venous events reported in 36 patients from patients reporting symptoms and venous duplex surveillance at follow-up visits. Most were found incidentally on follow-up imaging. All of these events were reviewed by an independent Clinical Events Committee and Medical Monitor. There were eight (8) symptomatic DVTs in eight (8) patients that were considered as part of the safety endpoint. However, one (1) was later assessed to be an arterial occlusion in the bypass graft. Symptomatic DVT occurred in 7.5% (4/53) of females and 2.0% (3/147) of males. All 7 symptomatic DVTs resolved. There were a total of 30 venous events in 28 patients that were not considered symptomatic or occlusive, and most were found incidentally on follow up imaging. One (1) was later assessed to be an arterial occlusion of the SFA and popliteal artery, Venous events occurred in 15.1% (8/53) of females and 12.9% (19/147) of males; 14 patients started on anticoagulation and 13 patients had no additional treatment. All but one (1) of the venous events resolved within 6 months. The one venous event was resolved by year 2, without any further intervention. Overall, there were no significant interventions (e.g., lysis, thrombectomy) used in the treatment of any of the venous events. There was no progression in size, or progression to pulmonary embolism.
Duplex Venous Observation Scale (DVOS)
Duplex Venous Observation Scale was performed at 30 days, 6 months and 12
months, as shown in Table 14 in the ITT Cohort. The analysis demonstrates a
DVOS of 0 for the majority of subjects indicating the vein was patent with no
presence of fibrin or thrombus.
Table 14: Duplex Venous Observation Scale – ITT
DVOS| 30 Days (N=173 1 )| 6 Months (N=175)| 12
Months (N=171)
---|---|---|---
0| 83.2% (144)| 94.3% (165)| 96.5% (165)
1| 0.6% (1)| 1.1% (2)| 0.0% (0)
2| 1.7% (3)| 0.0% (0)| 0.6% (1)
3| 0.6% (1)| 0.0% (0)| 0.0% (0)
4| 11.0% (19)| 4.0% (7)| 2.3% (4)
5| 2.9% (5)| 0.6% (1)| 0.6% (1)
Venous Clinical Severity Scale (VCSS) and Villalta Scale Results
Results for the VCSS scales at baseline and for each visit interval (30 days,
6 months, 12 months, 24 months and 36 months) are summarized in Table 15. The
VCSS instrument provides a general assessment for patients with chronic venous
disease. A negative change from baseline indicates improvement in the
parameter assessed. At 30 days, 6 months and 12 months, patients reported less
pain. In the patients that have reached the 24- and 36-months follow-up, the
pain and varicose veins were improved. However, overall VCSS score at 1 year
was 0.3 ± 1.98, indicating on average the study subjects had a decline in
their status with regard to their preexisting chronic venous disease.
Table 15: Venous Clinical Severity Score – ITT
Visit| All Available| Change from Baseline (Follow-up –
Baseline)
---|---|---
Baseline| 0.9 ± 0.89 (202)
1.0 (0,2)
| —
30 Day| 1.7 ± 2.23 (194)
1.0 (0. 14)
| 0.7 ± 2.11 (194)
0.0 (-2, 12)
6 Months| 1.2 ± 2.08 (189)
1.0 (0, 19)
| 0.3 ± 1.99 (189)
0.0 (-2, 17)
12 Months| 1.3 ± 1.97 (173)
1.0 (0, 12)
| 0.3 ± 1.98 (173)
0.0 (-2, 11)
Results for the Villalta scales scores for each visit interval (30 days, 6
months, 12 months, 24 months and 36 months) are summarized in Table 16. The
Villalta score is used for post-thrombotic syndrome. It combines a patient
self-assessment of symptoms and a healthcare assessment of clinical signs of
post-thrombotic syndrome. A negative change from baseline indicates
improvement in the parameter assessed. At 30 days, 6 months, 12 months, the
overall Villalta score showed an improvement. The overall change in score at 1
year was -0.6 ± 3.23. In the patients who had reached 24- and 36-months
follow-up, an improvement remains.
Table 16: Villalta Scale – ITT
Visit
| ****
All Available
| ****
Change from Baseline (Follow-up – Baseline)
---|---|---
Baseline| 2.0 ± 3.08 (202)
1.0 (0.17)
| —
30 Day| 1.7 ± 2.86 (194)
0.5 (0, 19)
| -0.4 ± 3.29 (194)
0.0 (-12, 17)
6 Month| 1.3 ± 2.37 (189)
0.0 (0, 16)
| -0.8 ± 3.16 (189)
0.0 (-14, 13)
12 Month| 1.5 ± 2.76 (173)
0.0 (0.17)
| -0.6 ± 3.23 (173)
0.0 (-13, 11)
Stent Graft Separation and Migration
Assessment of Stent Graft Separation and Migration was performed at 30 days, 6
months, 12 months, 24 months and 36 months, as shown in Table 17, through
differing imaging modalities. One ultrasound detected stent graft separation
and migration at the 12-month follow-up visit. One 12-month X-ray detected
three (3) migrations (1.8%; 3/169) and one stent separation (0.6%; 1/169).
There were no stent graft fractures at 12-months, and no additional
separations, migrations, or fractures have been seen in patients who have
reached follow-up through 36 months.
Table 17: Stent Graft Separation and Migration via Ultrasound and X-ray
-ITT
30 Days | 6 Months | 12 Months | ||
---|---|---|---|---|
Ultrasound | Stent Graft Separation | 0.0% (0/192) [0.0%, 1.9%] | 0.0% | |
(0/188) [0.0%, 1.9%] | 0.6% (1/180) [0.0%, 3.1%] | |||
0.0% | 0.0% | 0.6% | ||
Stent Graft Migration | (0/192) | (0/188) | (1/180) | |
[0.0%, 1.9%] | [0.0%, 1.9%] | [0.0%, 3.1%] | ||
Stent Graft | 0.0% | 0.0% | 0.6% | |
Separation or | (0/192) | (0/188) | (1/180) | |
Migration | [0.0%, 1.9%] | [0.0%, 1.9%] | [0.0%, 3.1%] | |
30 Days | 6 Months | 12 Months | ||
--- | --- | --- | --- | --- |
X-Ray | Stent Graft Separation | **** |
N/A
| ****
N/A
| 1.8% (3/169) [0.4%, 5.1%]
| | | 0.6%
Stent Graft Migration| N/A| N/A| (1/169)
| | | [0.0%, 3.3%]
Stent Graft| | | 1.8%
Separation or| N/A| N/A| (3/169)
Migration| | | [0.4%, 5.1%]
Effectiveness Endpoints
The analysis of effectiveness was based on the 188 evaluable patients at the
12-month time point. The primary effectiveness endpoint was primary patency,
defined as freedom from clinically driven target lesion revascularization (CD-
TLR) and absence of recurrent target lesion restenosis of >50% diameter
reduction by imaging (duplex ultrasound peak systolic velocity ratio of >2.5
or angiography) within the stent or 1cm immediately above or below the treated
segment at 12 months. When both imaging modalities are available, angiography
takes precedence. Primary patency at 12 months was 68.1% (128/188). The lower
bound of an exact one-sided 97.5% confidence interval is 60.9%. As this is
above the prespecified 60.4% threshold of the null hypothesis, the null
hypothesis is rejected, and the evidence supports that the primary
effectiveness is met. The primary effectiveness outcomes are summarized in
Table 18. Information on the 12-month primary patency rate with the roll-in
cohort is limited by a low number of patients available for analysis (n=18).
The primary patency Kaplan-Meier estimate is 57.1% and is within the 95%
confidence band of the roll-in cohort
Table 18: Primary Effectiveness Endpoint, Primary Patency at 12 Months –
MITT
All Subjects | Lower 97.5% Exact Confidence Limit |
---|
Absence of CD-TLR1 and absence of recurrent target lesion diameter stenosis
50%| 68.1% (128/188)
Patency failures 30% (60/200) CD-TLR: 14% (28/200) [35]
PSVR >2.5: 24% (48/200)
| ****
60.9% > 60.4%
In order to meet the criteria for Clinically Driven (CD), the CEC members reviewed the reported TLR source documentation against the CD-TLR definition. To meet the definition of CD-TLR, at least two (2) reviewers must have determined that the event met at least two of the first three criteria. It is noted that the criteria did not need to be the same between the reviewers, only that at least two clinical criteria are met. All TLR are described in Figure 6 below.
Secondary Effectiveness Endpoints
All secondary effectiveness analyses were performed on the primary analysis
cohort, the MITT analysis set.
Table 19: Secondary Effectiveness Endpoints – MITT Population
30 Days | 6 Months | 12 Months | |
---|---|---|---|
92.9% | 94.8% | 97.2% | |
Clinical Success | (182/196) | (182/192) | (173/178) |
[88.3%, 96.0%] | [90.6%, 97.5%] | [93.6%, 99.1%] | |
Primary Patency* | 96.5% | 87.9% | 72.3% |
[93.9%, 99.0%] | [83.4%, 92.4%] | [66.0%, 78.6%] | |
Secondary Patency* | 100.0% | 95.4% | 89.1% |
[100.0%, 100.0%] | [92.5%, 98.4%] | [84.7%, 93.5%] | |
Major Index Limb Amputations Over Time* | 0.0% [0.0%, 0.0%] | 0.5% [0.0%, 1.5%] |
1.6% [0.0%, 3.3%]
Table 20: Secondary Effectiveness Baseline Comparison – MITT Population
Baseline | 30 Days | 6 Months | 12 Months |
---|
Limb Ischemia: Rutherford Clinical Category (continuous) – MITT
| | | 3.3 ± 0.53| 0.6 ± 1.04| 0.5 ± 1.05| 0.6 ± 0.96
All| (200)| (196)| (192)| (178)
Available| 3.0| 0.0| 0.0| 0.0
| (3,5)| (0, 5)| (0, 5)| (0, 5)
| | | 3.2 ± 0.50| 3.3 ± 0.53| 3.3 ± 0.54
Paired Data| Baseline| —| (196)
3.0
| (192)
3.0
| (178)
3.0
| | | (3, 5)| (3, 5)| (3, 5)
| | | 0.6 ± 1.04| 0.5 ± 1.05| 0.6 ± 0.96
Paired Data| Follow-up| —| (196)
0.0
| (192)
0.0
| (178)
0.0
| | | (0, 5)| (0, 5)| (0, 5)
Paired Data| Change| —| -2.7 ± 1.12| -2.7 ± 1.07| -2.7 ± 1.05
from| (196)| (192)| (178)
| Baseline| 30 Days| 6 Months| 12 Months
---|---|---|---|---
| | Baseline (Follow-up
– Baseline)
| | -3.0
(-5, 1)
| -3.0
(-5, -2)
| 0.0
(0, 5)
Ankle-Brachial Index (ABI) – MITT
Target Limb
| ****
All Available
| | 0.61 ± 0.22
(191)
0.60
(0.00, 1.55)
| 0.99 ± 0.18
(183)
1.00
(0.00, 1.71)
| 0.99 ± 0.21
(173)
1.00
(0.37, 1.98)
| 0.95 ± 0.21
(170)
0.97
(0.00, 1.97)
Paired Data
| ****
Baseline
| ****
—
| 0.61 ± 0.22
(178)
0.61
(0.00, 1.55)
| 0.61 ± 0.22
(169)
0.59
(0.00, 1.55)
| 0.60 ± 0.22
(166)
0.59
(0.00, 1.55)
Paired Data
| ****
Follow-up
| ****
—
| 0.99 ± 0.18
(178)
1.00
(0.00, 1.71)
| 0.98 ± 0.21
(169)
1.00
(0.37, 1.98)
| 0.96 ± 0.21
(166)
0.97
(0.00, 1.97)
Paired Data
| Change from Baseline (Follow-up
– Baseline)
| ****
—
| 0.38 ± 0.25
(178)
0.38
(-0.66, 1.20)
| 0.38 ± 0.24
(169)
0.38
(-0.49, 1.12)
| 0.36 ± 0.27
(166)
0.37
(-0.74, 1.04)
Contralateral Limb
| ****
All Available
| | 0.81 ± 0.26
(187)
0.81
(0.00, 1.63)
| 0.87 ± 0.24
(181)
0.90
(0.21, 1.88)
| 0.89 ± 0.24
(170)
0.93
(0.28, 1.97)
| 0.88 ± 0.25
(163)
0.89
(0.02, 1.97)
Paired Data
| ****
Baseline
| ****
—
| 0.82 ± 0.26
(172)
0.82
(0.00, 1.63)
| 0.83 ± 0.26
(163)
0.82
(0.00, 1.63)
| 0.82 ± 0.26
(157)
0.82
(0.00, 1.63)
Paired Data| Follow-up| ****
—
| 0.87 ± 0.24
(172)
0.90
(0.21, 1.88)
| 0.89 ± 0.24
(163)
0.92
(0.28, 1.97)
| 0.88 ± 0.25
(157)
0.89
(0.02, 1.97)
Paired Data| Change from Baseline (Follow-up
– Baseline)
| ****
—
| 0.05 ± 0.21
(172)
0.04
(-0.44, 1.16)
| 0.06 ± 0.23
(163)
0.04
(-0.53, 1.25)
| 0.06 ± 0.28
(157)
0.04
(-1.51, 1.25)
6 Minute Walk Test (6MWT) – MITT
| All Available| | 208.8 ± 133.18 (142) 194.5 (18, 800)| 259.7 ± 147.84 (131)
229.8 (0, 864)| —| 288.4 ± 141.51 (118) 291.5 (0, 1004)
Paired Data| Baseline| —| 213.4 ± 132.63 (130) 199.6 (23, 800)| —| 209.8 ±
118.73 (117) 198.0 (23, 650)
Paired Data| Follow-up| —| 260.3 ± 148.26 (130) 230.7 (0, 864)| —| 289.4 ±
141.76 (117) 292.0 (0, 1004)
Paired Data| Change from Baseline| ****
—
| 46.8 ± 105.29 (130)| ****
—
| 79.6 ± 144.71 (117)
| Baseline| 30 Days| 6 Months| 12 Months
---|---|---|---|---
| | (Follow-up –Baseline)| | 30.2 (-270, 427)| | 61.0 (-366, 913)
Secondary Effectiveness assessments of the MITT group were performed at 30 days, 6 months, 12 months, 24 months and 36 months as shown in Table 19 and Table 20. At each time point in the study, the change from baseline showed RCC improved. At 30 days, 6 months, and 12 months, RCC improved by an average score of 2.7. At 24 months and 36 months, RCC improved by a score of 2.8 and 2.9. Using Kaplan Meier methodology, the primary patency at 1 year was 72.3%, and the secondary patency at 1 year was 81.8%. Ankle-Brachial Index (ABI) change from baseline on the target limb averaged a 0.4 (±0.25) change at 30 days and a 0.4 (±0.27) change as 12 months. 6 Minute Walk Test (6MWT) change from baseline averaged a 46.8 (±105.29) change at 30 days and a 79.6 (±144.71) change at 12 months.
Reinterventions
Interventions were classified as TVR, TLR, or CD-TLR, as presented in Table
21. TVR and TLR were further classified as lysis or non-lysis procedures.
Lysis was defined as the use of thrombolytic drug therapy during the
intervention or reporting of thrombolytic techniques by the site. It is noted
that some patients may have had multiple procedures, therefore counts may not
sum across rows or columns.
Table 21: Reinterventions
| # Events in Interval Up to 30 Days (0-30 days) % (n/N)[m] 1| # Events in Interval 6 Months (31-180 days) % (n/N)[m] 1| # Events in Interval 12 Months (181-390 days) % (n/N)[m] 1| Total Events to 12 Months
(390 days) (n/N)[m] 1
---|---|---|---|---
All TVR| 2%| 6.5%| 13%| 21.5%
| (4/200)| (13/200)| (26/200)| (43/200)
| [4]| [15]| [33]| [52]
TVR (excluding| 2%| 5%| 13%| 19%
lysis only) 2,3| (4/200)| (10/200)| (26/200)| (38/200)
| [4]| [12]| [30]| [46]
TVR (including| 0%| 1.5%| 1.5%| 2.5%
lysis only)| (0/200)| (3/200)| (3/200)| (5/200)
| [0]| [3]| [3]| [6]
All TLR| 2%| 6%| 11.5%| 19.5%
| (4/200)| (12/200)| (23/200)| (39/200)
| [4]| [14]| [29]| [47]
TLR (excluding| 2%| 4.5%| 11%| 16.5%
lysis only) 2,3| (4/200)| (9/200)| (22/200)| (33/200)
| [4]| [11]| [26]| [41]
TLR (including| 0%| 1.5%| 1.5%| 2.5%
lysis only)| (0/200)| (3/200)| (3/200)| (5/200)
| [0]| [3]| [3]| [6]
| # Events in Interval Up to 30 Days (0-30 days) % (n/N)[m]
1| # Events in Interval 6 Months (31-180 days) % (n/N)[m]
1| # Events in Interval 12 Months (181-390 days) % (n/N)[m]
1| Total Events to 12 Months (390 days) (n/N)[m] 1
---|---|---|---|---
CD-TLR| 1.5%| 5%| 9%| 14%
| (3/200)| (10/200)| (18/200)| (28/200)
| [3]| [12]| [20]| [35]
Subgroup Analyses
The following subgroups were assessed for impact on Primary Safety and Primary
Effectiveness: gender and region. Please note that the stratification of the
study into these smaller cohorts reduces the sample size available to
accurately assess estimates and may impact interpretability.
Gender
The subgroup analysis by gender showed that females had a greater risk of
primary patency failure at one year compared to males.
Table 22: Primary Patency at 12 Months by Gender – MITT
Subgroup| All Subjects| Lower 97.5% Exact Confidence|
P-value
---|---|---|---
Female| 49.0% (24/49)| 34.4%| 0.001
Male| 74.8% (104/139)| 66.8%
There are multifactorial issues that impact the assessment of study outcomes for the female gender. Further information is available below on the post hoc, multivariate regressions analysis, completed to provide additional information. There are large studies that show that female gender does not contribute to graft failure and reduced patency. It has been hypothesized that reasons for poorer outcomes in women are multifactorial, including anatomical variables (e.g., smaller vessels), presentation with more advanced disease, and biological characteristics that influence the long-term outcomes of revascularization.
Region
Additionally, the analysis showed that primary patency was significantly
better in the OUS patients compared with the US patients.
Table 23: Primary Patency at 12 Months by Region – MITT
Subgroup| All Subjects| Lower 97.5% Exact Confidence Limit|
P-value
---|---|---|---
OUS sites| 89.4% (42/47)| 76.9%| <0.001
US sites| 60.1% (86/143)| 51.6%
The US population had higher rates of ISR patients [21.6% (33/153) vs. 4.1% (2/49); p=0.004], a shorter patent length of the above knee popliteal section [79.53 ± 44.2 (146) vs 99.15 ± 53.2 (44); p=0.015], more CFA disease present (9.2% vs 0%; p=0.024), and smaller distal CFA diameters [7.65 ± 1.7 (150) vs. 8.73 ± 1.4 (48); p<0.001]. These lesion characteristics may have contributed to a lower primary patency for the US cohort. Post-hoc multivariate logistic regression analyses, (Table 24) were conducted to select the covariates that best explain the variability in primary endpoints.
In summary, it was found that smaller distal CFA diameters, shorter
height/heavier weight, longer CTO length and history of smoking were
predictors of poor patency. When accounting for these variables, gender became
a statistically insignificant predictor.
When evaluating the differences between the US and OUS cohorts, each of these
predictor variables provided the OUS with an advantage. That is, the European
participants were taller and had larger vessels. Therefore, it is uncertain
whether outcomes for these cohorts are more directly correlated to anatomic
and lesion characteristics. However, with small numbers in these groups, it is
difficult to draw definitive conclusions.
Table 24: Odds Ratio Estimates for Patency Model
Odds Ratio Estimates
Effect| Point Estimate| 95% Wald Confidence Limits
Region: OUS vs US| 0.403| 0.154| 1.057
Distal CFA Diameter| 0.754| 0.595| 0.954
Height| 0.924| 0.881| 0.968
Weight| 1.031| 1.005| 1.058
CTO Length| 1.046| 0.999| 1.094
Smoking Status (current vs former)| 2.707| 1.323| 5.538
Directions for Use
The DETOUR System is supplied sterile and is intended for Single Use Only.
OVERVIEW OF THE PROCEDURE USING THE DETOUR SYSTEM
The DETOUR System is utilized to perform the procedure using the DETOUR
System. The procedure using the DETOUR System is an endovascular femoral-
popliteal bypass using the femoral vein as a conduit for TORUS Stent Grafts
which provides a bypass of the diseased arterial segment. Utilizing standard
endovascular techniques, the ENDOCROSS Device is used to create an arterio-
venous connection above the diseased arterial segment, and then a veno-
arterial connection below the diseased arterial segment. TORUS Stent Grafts
are then placed from distal to proximal sequentially to provide the bypass.
Multiple TORUS Stent Grafts may be utilized during a procedure using the
DETOUR System. A graphical representation of the procedure using the DETOUR
System is provided in Figure 7.
MATERIALS REQUIRED FOR THE PROCEDURE USING THE DETOUR SYSTEM
- Commercially available vascular snare (e.g., EN Snare® Endovascular Snare (Merit Medical), Amplatz Goose Neck™ snare (Medtronic)
- 0.014” guidewires, 300cm length (2x) and a 180cm length (1x)
- 0.035” guidewire, at least 300cm in length
- 8F introducer sheath (at least 45cm in length) for contralateral arterial access
- 4F or 6F introducer sheath for venous access
- 4F or 5F Sizing/Marker and/or Exchange Catheter
- Angioplasty balloons of various diameters and lengths
- Inflation device for use with angioplasty balloons
- Ancillary tools to gain arterial and venous access using standard technique
PRECAUTION: To minimize wire differentiation it is recommended to use a 180cm long 0.014” guidewire through the ENDOCROSS Device RX guidewire lumen.
ENDOCROSS DEVICE USE AND ANASTOMOSIS CREATION
ENDOCROSS Device Set Up
-
Inspect the ENDOCROSS Device pouch for damage. Do not use device if the package has been opened or damaged.
WARNING: Do not use the ENDOCROSS Device if the package is opened or damaged. Prior to use, perform a thorough examination of the pouch materials and seals to ensure there is no visually detectable damage.
WARNING: Do not use the DETOUR System beyond the stated use-by date on the package label. -
Using sterile technique, remove the ENDOCROSS Device from the package and transfer to the sterile field. Inspect the catheters for any visible signs of damage.
WARNING: Inspect the DETOUR System prior to use. Do not use if the device appears to be bent, kinked, or damaged in any way. The ENDOCROSS Device should be test fired prior to use in the patient to ensure proper functionality. Do not use if the device does not function properly when tested, or if the device appears to be bent, kinked, or damaged in any way because vessel damage and/or inability to advance or withdraw the device may occur. -
Flush the ENDOCROSS Device with heparinized saline through the ports and advance the outer handle to ensure the Stabilizer deploys and Needle can be deployed by depressing the Button.
-
After deploying the needle, flush with heparinized saline through the rear needle guidewire port, then retract the needle and the outer handle until collapsing the stabilizer and returning needle to the “locked” position. The “locked” position is indicated when a “click” is heard upon retraction of the outer handle.
WARNING: Do not place finger or other body parts near distal tip of the ENDOCROSS Device prior to deploying the Needle.
Isolate Vascular Targets
WARNING: The distal CFA must be measured; the DETOUR System is
contraindicated for patients with a distal common femoral artery (CFA) <7mm in
diameter.
- Gain tibial venous access and place an appropriately sized introducer sheath.
- Advance a snare per the product’s instructions for use into the femoral vein (FV) at a level just above the arterial lesion.
- Gain contra-lateral common femoral arterial access using standard technique and place an 8Fr sheath.
- Load a 300cm 0.014” guidewire through the rear Needle guidewire port of the ENDOCROSS Device and advance until the distal tip is just visible then retract approximately 5cm.
- Backload the 0.014” guidewire in RX Port and advance ENDOCROSS Device 3-4cm distal from the SFA ostium.
Create Proximal Anastomosis
-
Rotate imaging to superimpose the ENDOCROSS Marker Band positioned in the superficial femoral artery (SFA) over the FV, adjusting both lateral and cranial/caudal planes to optimize overlap of the FV and SFA.
-
If desired, rotate 90 degrees in either direction to obtain orthogonal views.
-
Once the ENDOCROSS Device is superimposed over the snare in the FV, orient the ENDOCROSS Marker Band to a posterior image. (Figure 5).
-
On the ENDOCROSS Device, pull back the outer handle until it clicks into the starting position. Then advance the outer handle forward until the stabilizer is deployed and the needle is spring-loaded, then rotate the outer handle to load the needle in place. Re-confirm the orientation of the ENDOCROSS Device marker band. If orientation is not adequate retract the Outer Handle and repeat until appropriate orientation is achieved with Stabilizer deployed.
WARNING: Do not rotate the ENDOCROSS Device while Stabilizer is deployed. -
Depress the Deployment Button to deploy the Needle into the adjacent FV with the ENDOCROSS marker band in the posterior crossing position.
WARNING: Place hand and fingers on the inner handle (not on the guidewire port area) of the ENDOCROSS Device during needle deployment to avoid pinching gloves during needle deployment. -
Rotate to an orthogonal view, and advance the 300cm 0.014” guidewire through the Needle to confirm FV access. Repeat Steps 2-5 until venous access is confirmed.
WARNING: Do not advance the guidewire significantly through the Needle tip prior to Needle deployment. Doing so may damage the device and/or guidewire tip. -
NOTE: If the ENDOCROSS Device Needle extends through both walls of the target vein, partially retract the Needle using the Outer Handle and advance guidewire into snare to gain venous access.
-
Once the 300cm 0.014” guidewire position within the venous lumen is confirmed, snare the floppy to stiff transition portion of the 0.014” guidewire, and retract the needle using the outer handle on the ENDOCROSS Device until it “clicks” into the locked position.
-
Withdraw the snare and the 300cm 0.014” guidewire through the venous access site. Secure guidewire with a hemostat or torque device.
-
Withdraw the ENDOCROSS Device while ensuring the distal portion of the guidewire remains externalized at the ipsilateral venous access site.
-
Re-prep the ENDOCROSS Device by flushing with heparinized saline through the ports and advance the outer handle to ensure the Stabilizer deploys and Needle can be deployed by depressing the Button.
-
After deploying the needle, flush with heparinized saline through rear needle guidewire port then retract the needle and the outer handle, collapsing the stabilizer and returning needle to the “locked” position.
WARNING: Ensure the Stabilizer and Needle are retracted prior to removing the ENDOCROSS Device. -
Use an appropriately sized PTA balloon to dilate the anastomosis between the SFA and FV.
Create Distal Anastomosis
-
Backload the proximal end of the externalized 300cm 0.014” guidewire into the tip of the ENDOCROSS Device until it exits the rapid-exchange (RX) port and re-advance the ENDOCROSS Device to the newly created proximal anastomosis.
-
Load a 300cm 0.014” guidewire through the needle guidewire port of the ENDOCROSS Device and advance until the distal tip is just visible then retract approximately 5cm.
-
While maintaining tension on the 0.014” guidewire at both access sites, gently advance the ENDOCROSS Device through the proximal anastomosis into the FV.
WARNING: Do not deploy the Stabilizer prior to advancing through the proximal anastomosis.
WARNING: Do not exert excessive force while advancing the ENDOCROSS Device through the anastomosis. Ensure tension is applied to the guidewire at both access sites and gently advance the ENDOCROSS Device until FV access is confirmed fluoroscopically. Re-dilate anastomosis if significant resistance is encountered.
PRECAUTION: A knee prosthesis may prevent successful and accurate ENDOCROSS Needle deployment. -
Continue advancing the ENDOCROSS Device until the Crown Marker Band is at the desired crossing location in the Popliteal vein that corresponds to the area distal to the lesion in the Popliteal artery.
-
Visualize the target artery (either distal SFA or proximal Popliteal Artery) and rotate the C-arm with fluoroscopy to superimpose the Femoral/Popliteal vein at the target artery re-entry location. To optimize the overlap of the popliteal artery and popliteal vein both lateral and cranial/caudal planes adjustment may be needed. Verify overlap with angiography.
-
If desired, rotate 90 degrees in either direction to obtain the orthogonality of the vessels.
-
In the superimposed position, while injecting contrast, using the ENDOCROSS Marker Band, orient the marker band to the anterior position towards the artery.
-
Advance and rotate the Outer Handle until the stabilizer is deployed. Re-confirm the orientation of the ENDOCROSS Marker Band and adjust as necessary.
-
Maintain tension on the externalized 0.014” guidewire prior to Needle deployment.
-
Depress the Button to deploy the needle through the venous and arterial walls and into the adjacent SFA/Popliteal artery.
WARNING: If the ENDOCROSS Device Needle extends through both walls of the target artery, partially retract Needle using the Outer Handle and advance guidewire until you are able to advance the guidewire to gain arterial access. -
Advance the 300cm 0.014” guidewire through the Needle to confirm arterial access. Repeat Steps 6-9 as necessary until 0.014” guidewire placement is confirmed in target artery.
-
Withdraw the ENDOCROSS Device while ensuring distal arterial 0.014” guidewire position is maintained.
WARNING: Ensure the Stabilizer and Needle are retracted prior to removing the ENDOCROSS Device. -
Use an appropriately sized PTA balloon to dilate the distal and proximal anastomosis sites.
-
Using an exchange catheter replace the 0.014” guidewire with a 0.035” 300cm guidewire.
-
Remove the externalized 300cm 0.014” guidewire.
TORUS STENT GRAFT SYSTEM USE
- Sizing and Selection of the TORUS Stent Graft
- Prior to opening the sterile package, check to ensure that the diameter and length of the TORUS Stent Graft are correct and confirm that the product has not reached the expiration date.
- Select the appropriate stent graft diameters for the distal and proximal vessels using Table 1 and Table 2 above.
- Select the appropriate stent graft lengths based on the length of the proposed bypass.
* **A.** The lengths provided in “Sizing” are the nominal stent graft lengths.
* **B.** It is important that the stent graft occupy at least 3cm of the arterial lumen on both the distal and proximal sides of the lesion.
4. When overlapping (telescoping) grafts, at least 60mm of overlap is required.
* **C.** A 5.5mm stent graft may only be overlapped within a 5.5mm stent graft.
* **D.** A 6.0mm stent graft may be proximally overlapped within a 5.5mm or 6.0mm stent graft.
* **E.** A 6.7mm stent graft may be proximally overlapped within a 6.0mm or 6.7mm stent graft.
-
Set-Up
-
Inspect the TORUS Stent Graft System pouch for damage. Do not use the device if the package has been opened or damaged.
-
Using sterile technique, remove the device from the package and transfer to the sterile field. Inspect the device for any visible signs of damage.
PRECAUTION: Do not deploy the stent graft prior to insertion. The stent graft is self-expanding and cannot be re-loaded on the delivery system. -
Flush Guidewire Lumen using the luer-lock connector on the proximal end of the device using heparinized saline until fluid exits the catheter tip.
-
Flush the stent graft Lumen using the flush port on the device handle until fluid exits the sides of the TORUS Stent Graft System, proximal to the stent graft location.
-
Stent Graft Placement
-
Insert the TORUS Stent Graft System over the 300cm 0.035” guidewire and advance to the treatment site.
PRECAUTION: Do not use excessive force when manipulating the device through an introducer. Excessive force may damage the device. -
Advance the TORUS Stent Graft System through the proximal and distal anastomoses to the intended target landing zone in the artery
WARNING: Do not exert excessive force while advancing the TORUS Stent Graft System through the anastomosis. Gently advance the TORUS Stent Graft System until distal arterial access is confirmed fluoroscopically. Re-dilate anastomosis if significant resistance is encountered. -
Advance the distal marker band of the TORUS Stent Graft System a minimum of 3cm distal to the distal anastomosis, leaving the proximal marker band within the Femoral Vein.
-
Hold the TORUS Stent Graft Delivery System deployment handle flat on the table with the knob facing up and turn the knob counterclockwise (i.e., in the direction of the arrow) to deploy the stent graft.
-
Once the stent graft is completely deployed, remove the TORUS Stent Graft Delivery System.
PRECAUTION: When removing the delivery system avoid displacing the Stent Graft by removing the delivery system under fluoroscopy.
PRECAUTION: When deploying multiple TORUS Stent Grafts, it is recommended to have at least 60mm of overlap of each stent graft. -
If needed, insert a subsequent TORUS Stent Graft System over the same 0.035” guidewire and advance through the proximal anastomosis to the level of the first stent graft placed.
-
Position the distal marker band on the TORUS Stent Graft Delivery System inside the deployed TORUS Stent Graft allowing for approximately 60mm of overlap.
-
For the final TORUS Stent Graft placement, ensure the proximal marker band is at least 3cm above the proximal anastomosis in the SFA and at the bifurcation of the SFA and profunda. Ensure that there is at least 60mm of overlap when deploying a stent graft inside of a previously placed stent graft.
-
After confirming position at both locations, deploy the stent graft completing the bypass, remove the TORUS Stent Graft Delivery System.
WARNING: When deploying the proximal stent graft please ensure that the proximal edge (exposed end) of the stent graft is 1-2mm above the ostium of the superficial femoral artery (Figure 4). -
Dilate each stent graft along its entire length using the balloons designated in Table 1 of the Sizing section above, ensuring that full dilatation of both anastomosis sites is performed so that optimal laminar flow is achieved.
WARNING: Do not dilate outside the margins of the deployed stent graft(s). -
Perform final angiogram and venogram.
-
Remove the 0.035” guidewire and proceed with hemostasis techniques (i.e., vessel closure devices, manual compression).
-
Patient Surveillance
-
Refer to AHA and SVS guidelines and recommendations for antiplatelet and/or anticoagulation medication in the post-operative period.
-
Refer to AHA and SVS guidelines on patient surveillance.
Disposal
After use, dispose of product and packaging in accordance with hospital, administrative and/or local government policy.
Storage and Handling
Store devices in a cool, dry environment away from direct sunlight.
Definitions of Symbols
Manufacturer
Endologix LLC 1971 Milmont Drive Milpitas, CA 95035 U.S.A.
www.endologix.com
800-983-2284 (949) 595 7200
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>