The Tenney Junior temperature test chambers have been in the market for over 50 years. They are well suited for use in electronic, military, and pharmaceutical quality assurance and reliability testing, as well as research testing and production processes. In keeping with the needs of today’s lab, we specifically designed these benchtop and floor chambers to have a compact exterior yet an ample interior workspace to maximize valuable floor space.

Features

  • All models feature vapor-tight, continuously welded stainless steel interiors.
  • Structural reinforcement is used at all critical points.
  • Temperature conditions are controlled with the Watlow F4T profiling type dual channel 1/4 DIN controller
  • Data logging and Graphical trend chart for Watlow F4T controller
  • Fiberglass insulation surrounds the temperature chamber to maximize insulating characteristics, thus ensuring minimal thermal transfer.
  • 120V power supply cord with male NEMA 5-15P plug
  • 3″ port is standard on left side

Options

  • GN2 Purge System
  • Dry Air Purge System
  • Tempguard IV Over-temperature Protection
  • Viewing windows up to 6″ x 8″ and thermally insulated
  • Interior lighting
  • Shelving, adjustable and removable
  • Automatic CO2 or LN2 cooling boost system
  • Recording instruments
  • Alternative power supply options for many global current schemes
  • Cart system to allow chamber to be moved within a facility
  • Stacking option for bench model
  • 50 Hz -CE mark available
  • Intrinsically safe interior
  • NIST Calibration
  • Tenney Environmental offers calibration services accredited to the ISO 17025 standard
  • Additional engineered to order custom configurations available

Specifications

Model

|

TJR

|

TUJR

---|---|---

Cubic Feet

|

1.12

Liters

|

31.71

Work Space (WxDxH)

|

16″ X 11″ X 11.75″
406mm x 279mm x 298mm

|

16″ X 11″ X 11.75″
406mm x 279mm x 298mm

Exterior (WxDxH)

|

37″ X 22″ X 31 .75″
940mm x 559mm x 806mm

|

25.75″ X 22″ X 60″
655mm x 559mm x 1524mm

Approx. Unit Weight

|

260Ibs / 118kg

|

292Ibs / 133kg

Approx. Crated Weight

|

32Slbs / 148kg

|

36Slbs / 165kg

Temperature Range


Standard Temperature Range

|

-68°C to  180°C

Optional Temperature Range

|

-75°C to  200°C

Cooling Rates (Empty Chamber) Min (°C/min)


180°C to -65°C

| 81 (3)| Ambient to -35°C|

14 (4.2)

125°C to -35°C

| 43 (3.7)| Ambient to -40°C|

17 (3.8)

85°C to -40°C

| 31 (4)| Ambient to -40°C|

17(3.8)

Heating Rates (Empty Chamber) Min (°C/min)


-65°C to 180°C

| 64 (3.8)| -68°C Ambient| 14 (6.6)
-20°C to 85°C| 23 (4.5)| Ambient to 85°C|

14(7.8)

-40°C to 85 °C

| 25 (5)| Ambient to 125°C| 27 (5.5)
-35°C to 180°C| 61 (3.5)| Ambient to 180°C|

50 (4.1)

Live Load Capacity (Empty Chamber)

Temperature Range

|

-40°C

| 170 Watts
-54°C|

145 Watts

-65°C

| 105 Watts
-73°C|

60 Watts

 |

TJR

|

TUJR

---|---|---

Temperature Control

|

± 1.0°C after stabilization

Air Flow Rate

| 50CFM
Heat Capacity|

500W

Refrigeration

| Two (2) 1/2 HP
Condenser|

Air cooled

Condenser Discharge Air

| Exits on right side of the chamber
Heat of Rejection|

Maximum 6300 BTU’s per hour

Power

| 120V-1PH-60Hz
Amp- Fuse|

20

Amp – Under Load

| 16
NEMA Plug|

Male NEMA 5-l SP

Standard Ports Included (loc)

| 3″ Standard Port w/Plug centered left sidewall
Decibel Level|

Less than 72 dBa as measured 3 feet from chamber, S feet from the floor at right angles

Load Per Shelf

| 25 lbs/11kg per shelf (maximum chamber load including shelf and floor load not to exceed 30 lbs / 14 kg)
Number of Shelves Standard|

None

Max Number of Shelves

| 3 shelves
Shelf Spacing|

2″/50mm on Pilasters

Chamber Spacing Requirements

| 18″/458mm on all sides
Floor Load Rating|

30lbs/14kg

Performance is based upon an empty chamber with air temperature measured at the supply air stream, operating at 24°C (75°F) ambient air.
Operating at 50Hz or higher than 24°C ambient air, performance may be reduced. Consult factory regarding any special cooling requirements.

Tenney Environmental
2821 Old Route 15 | New Columbia, PA 17858 USA | TEL 570-538-7200 | FAX 570-538-7380
www.tenney.com

TPS-Logo.png

Tenney, Lunaire, Blue M, Gruenberg Lindberg/MPH, Wisconsin Oven

Tenney-Logo.png

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals