The Supine Stander is a medical device designed to support users in a standing position. It provides greater control over the head, trunk, pelvis, knees, and feet while allowing variable weight-bearing. The stander is suitable for users who lack sufficient anti-gravity control of the head and shoulder girdle. It is also equipped with a folding headboard to encourage independence for users with sufficient head and upper trunk control.
The Supine Stander comes in two models:
- E420 Small Supine Stander
- E430 Large Supine Stander
Key user dimension: height. Select the appropriate stander based on the user’s overall height. Choose the model that allows for growth. The user’s weight must not exceed the maximum working load.
Supine Stander Usage Instructions
Checking Your Order
Upon receiving your Supine Stander, check the contents of the package to ensure everything is included. The package should contain:
- Supine Stander (including padding, footboard, adjustable armrests, hip strap, chest strap, leg straps, one pair of laterals, one pair of head blocks, and double-locking casters).
- Components of the Supine Stander.
If your shipment is incomplete or damaged, please contact Customer Service.
Attaching Headboard
WARNING: To prevent spinal injury, make sure the headboard is in place before tilting the client to a horizontal position. Also, ensure that all body parts are clear of the joint between the headboard and mainboard during adjustment.
- Lift the headboard and slide the cleat (A) into the frame slot.
- Tighten the cleat knob (B) to secure the headboard.
Attaching Head Blocks
- Loosen the two knobs (C) on the head block adjusting clamp.
- Remove the lateral with a curved bracket (D), turn it upright, and insert the bracket into the clamp.
- Slide the curved bracket into the clamp to the desired spacing. If spacing of less than 8 is required, ensure that the brackets overlap.
- Tighten both knobs (C) to secure the head block.
For detailed instructions on operation, maintenance, cleaning, and warranty, please refer to the product manual available at http://www.rifton.com /customer-service/product-manuals.
- IMPORTANT
Please save this product manual for future reference. Additional copies are available at http://www.rifton.com/customer-service/product- manuals.
Key for users
- Use this key to determine which sections of this product manual apply to you.
- Technical Users For professionals who order and set up Rifton products.
- Home Users For caregivers who use Rifton products on a regular basis.
- Maintenance Personnel For anyone who is responsible for service or re-ordering of Rifton products and parts.
WARNING
- Thoroughly read and understand the information in this product manual before attempting to use this product. If the procedures and instructions in this product manual are not followed, serious injury or death could occur.
- A qualified professional must assess the appropriateness and safety of all equipment for each user.
- This product is intended for use by clients of unreliable judgment. Adult supervision is required at all times.
- To prevent falls and injuries:
- Do not use this product on rough and uneven terrain, around swimming pools, or near stairways.
- Ensure the appropriate use of straps and supports at all times. Straps and supports are provided for the safety of the user and must be carefully adjusted for comfort and security.
- Tighten all adjustment knobs before use and immediately after making any adjustments.
- Do not use this product for clients outside the height and weight limits specified in this manual.
- To prevent structural failure, which may result in serious injury or death:
- Inspect this product and components regularly for loose or missing screws, metal fatigue, cracks, broken welds, missing attachments, general instability, or other signs of excessive wear.
- Immediately remove this product from use when any condition develops that might make the operation unsafe.
- Do not use Rifton components or products for any purpose other than their intended use.
Recommended use
WARNING
To prevent falls and injury, do not use the Supine Stander as a transportation device.
- The Rifton Supine Stander is a Class 1 medical device. It supports the user in the standing position. The Supine Stander differs from a tilt table in that it allows much greater control of the head, trunk, pelvis, knees, and feet while allowing variable weight-bearing. It is especially suitable for users who lack sufficient anti-gravity control of the head and shoulder girdle.
- For users who have sufficient head and upper trunk control, the headboard folds down to encourage independence.
- The E430 large Supine Stander is recommended for children/adults between 46”– 72” (117 cm – 183 cm) tall and up to 250 lbs (114 kg).
- The E420 small Supine Stander is recommended for children/adults between 30”– 50” (75 cm – 25 cm) tall and up to 100 lbs (45 kg).
User and item dimensions
User dimensions – inches (cm) | small | large |
---|---|---|
Height | 30 –50 (76-127) | 46 –72 (117-183) |
Key user dimension: height
Select the appropriate Stander by the user’s overall height. Choose the model that allows for growth.
Important: The user’s weight must not exceed the maximum working load.
Item dimensions – inches (cm)| small| large
Board length and width| 53 x 16½ (135 x 42)| 73 x 19 (185 x 48)
Distance between laterals| 7–13 (18–33)| 9 –16 (23–41)
Height when horizontal| 28½ (72)| 21–30 (53–76)
Height when vertical| 54½ (138)| 74 (188)
Base length and width| 37½ x 27 (95 x 69)| 46½ x 29½ (118 x 75)
Max. working load in lbs. (kg)| 100 (45)| 250 (114)
Check your order
This product comes in a single carton, containing:
- Supine Stander (includes padding, footboard, adjustable armrests, hip strap, chest strap, leg straps, one pair of laterals, one pair of head blocks and double-locking casters).
- Components of the Supine Stander.
- You may not have ordered all of the available components, but use the diagrams that follow to confirm that your order is complete.
If your shipment is incomplete or in any way damaged on arrival, please call Customer Service, 800.571.8198.
Basic item
Headboard
WARNING
To prevent spinal injury, ensure the headboard is in place before tilting the
client to a horizontal position.
To prevent pinching or crushing, ensure that all body parts are clear of the
joint between headboard and mainboard during adjustment.
Attaching headboard:
Figure 6a: Lift the headboard, slide the cleat (A) into frame slot, and
tighten the cleat knob (B).
Attaching head blocks:
Figure 6a:
- Loosen two knobs (C) on the head block adjusting clamp.
- Remove the lateral with curved bracket (D), turn upright, and insert bracket into clamp.
- Slide curved bracket into clamp to desired spacing. Brackets must overlap if spacing of less than 8” is required.
- Tighten both knobs (C) to secure the head block.
Adjusting headboard:
- Figure 6a: Position headboard by raising it and sliding cleat (A) into the frame slot. Tighten cleat knob (B). If headboard is not needed, slide cleat up, tighten knob, and fold the headboard down.
Adjusting head blocks:
- Figure 6a: Loosen the two knobs (C) on the clamp, slide the block to the desired position, and tighten knobs securely.
Footboard
WARNING
To prevent falls, ensure the footboard is in place before tilting the client
to a vertical position.
To prevent pinching or crushing, correctly position the footboard at the
appropriate height before placing the client on the footboard.
To prevent injury, lock all four casters at all times, regardless of whether
or not the product is occupied, except when moving the stander from one
location to another.
Attaching sandals:
Figure 7a: Either large or medium sandals can be attached to the
footboard.
- Insert carriage bolts with bushings through the bottom of the footboard.
- Insert sandal base onto bolts and adjust sandal before tightening knobs (A) securely.
- Place sandal onto base and lock into place with latch (B).
Adjusting sandals:
Figures 8a and 8b: To adjust sandals, loosen knobs (A), move sandals to
the desired position, and tighten knobs. Sandal height can be built up or a
toe/heel angle achieved by using one or more wedges.
Attaching sandal wedges:
Figure 8a: Use latches (B) to add a wedge. Unlatch and remove sandal from
the base, place wedge over the base and latch it into place; any number of
wedges can be added. Place and latch the sandal on top of the wedges.
Adjusting footboard:
Figure 8b: The footboard can be adjusted to accommodate users of varying
heights. Loosen the clamp knobs (C), move board to the desired height, and
tighten knobs securely. For the user’s security, footboard is best left as
close as possible to the floor.
Adjusting the casters:
Figure 8c: Swivel casters allow the Supine Stander to be moved easily.
The foot-operated brake, locks both the wheel and the swivel
simultaneously.
Small Supine Stander mainboard (E420)
Adjusting the mainboard:
Figure 9a: The hand crank (A) controls the tilt angle of the mainboard.
By turning the crank, the board will begin to tilt. Continue to turn the crank
until the desired angle is reached.
The tilt angle is indicated on the board angle label. This can be used to
record the correct angle setting for each user.
Attaching laterals:
Figure 9b: Laterals (B) ship unassembled. Follow the instructions below
to attach them.
- Remove the knob and plastic washer, leaving the rubber washer on the lateral.
- Slide the bolt of the lateral through the slot in the board.
- Replace knob (C) and plastic washer on the bolt-on back of the board, and tighten knob (C) securely.
Adjusting laterals and straps:
Any lateral or strap can be adjusted in and out or up and down. To adjust in
or out, loosen the appropriate knobs, slide the lateral or strap to the
desired position, and tighten the knobs securely.
To adjust up and down or to completely remove a lateral or strap, loosen the
knob(s) completely and remove the lateral or strap. The lateral or strap can
then be moved to a different set of slots, and the knobs tightened securely.
Adjusting hip and leg straps:
WARNING
To prevent falls and injury, secure the strap safety closures (A) when in use.
Figure 10a: To adjust the height of hip and leg straps (A), pull the back of
the strap away from the hook-and-loop fabric at the back of the mainboard,
slide the strap to the desired height, and secure the hook-and-loop fabric
Adjusting armrests:
To prevent pinching or crushing, ensure body parts are clear of potential
crush points while adjusting the armrest.
Figure 10a: To adjust the height of the armrest, loosen both knobs (B),
slide the armrest to the desired position, and tighten the knobs securely. To
adjust the angle of the armrest, loosen the bottom knob only, lift or lower
the front of the armrest to the desired position, and tighten the knob
securely.
Large Supine Stander mainboard
(E430)
Adjusting the mainboard:
Figure 11a: The hand crank (A) controls transfer height adjustment and
the tilt angle of the mainboard. By turning the crank, the board in its
horizontal position will move vertically from 21″ to 30″(53cm to 76cm) and
then begin to tilt. Continue to turn the hand crank until the desired angle is
reached.
The tilt angle is indicated on the board angle label (B). This can be used to
record the correct angle setting for each user.
Attaching laterals:
Figure 11b: Laterals (D) ship in reverse position.
To reposition laterals:
- Loosen knobs (C) and slide laterals out from under the clamp.
- Rotate laterals into the correct position and replace them under the clamp.
Tighten knobs.
Use knobs (C) to adjust the laterals laterally and vertically.
Adjusting laterals and straps:
Figure 11c: Any lateral or strap can be adjusted in and out, or up and
down. Loosen the two knobs on the appropriate clamp, slide lateral or strap to
the desired position, and tighten the knobs securely.
To completely remove a lateral or strap, loosen clamp knobs enough to slide
the bracket out. The lateral or strap can then be moved to a different set of
clamps and the knobs tightened securely.
Adjusting laterals and straps cont.
If you are using the large Supine Stander for a client whose head does not
extend above the mainboard, remove the head blocks and the two clamps. Remount
the clamps in the top slots of the mainboard. Attach the head blocks as
described on
Adjusting hip strap:
WARNING
To prevent falls and injury, secure the strap safety closures (C) when in use
(see Figure 12b).
Figure 12a: It is possible to place the hip strap (A) in line with the
laterals. Loosen the clamp knobs enough to allow the hip strap’s metal bracket
to lie on top of the lateral. Tighten clamp knobs securely. Adjusting leg
straps:
To adjust height of leg strap, pull back of strap away from hook-and-loop
fabric at back of the mainboard, slide strap to the desired height, and secure
hook-and-loop fabric.
Adjusting armrests:
WARNING
To prevent pinching or crushing, ensure body parts are clear of potential
crush points while adjusting armrest.
Figure 12b: To adjust height of armrest, loosen both knobs (B), slide
armrest to desired position, and tighten knobs securely. To adjust angle of
armrest, loosen the bottom knob only, lift or lower front of armrest to
desired position, and tighten knob securely.
Components
Tray
Attaching the tray:
Figure 13a: A tray can be fitted to the armrests. Position both armrests
at the same height and angle. Attach tray by inserting the tray latches into
appropriate armrest holes. Adjusting the tray:
Figure 13b: The tray can be adjusted in and out by opening the tray
latches and moving them to a different set of holes in the armrests.
Hand Anchors
To prevent falls and resulting injury, do not use Rifton Anchors as a primary
support or as a standing support. Suction cups may release.
To prevent falls and cuts, do not use Rifton Anchors on thin glass.
Recommended use
Figure 14a: Hand Anchors can be used on any of our available trays. They
give the client an anchor to control excessive movement of the arms and
encourage stabilization of the shoulder girdle. The client may hold on with
both hands for stabilization training or with one hand, keeping the other hand
free. Operation
On a clean smooth table or tray, press down on suction cup and turn handhold
clockwise until secure. To remove, turn handhold counterclockwise.
Tip: Surface must be very smooth and completely clean to get a strong
seal. If suction cup does not hold well, wipe with a damp cloth or apply a
thin layer of petroleum jelly around the rim of the rubber pad to enhance the
vacuum.
Arm Anchors
To prevent falls and resulting injury, do not use Rifton Anchors as a primary
support or as a standing support. Suction cups may release.
To prevent falls and cuts, do not use Rifton Anchors on thin glass.
Recommended use
Figure 15a: The arm Anchor is intended to stabilize one arm on a table or
tray, or as a pair to stabilize both arms. Providing increased stability to
the forearm, arm Anchors may be beneficial for head and trunk control to allow
functional use of the free arm. Adjustment
Arm and wrist straps can be adjusted and secured with hook-and-loop fasteners.
Move handhold backward and forward or rotate by loosening the knob.
Operation
On a clean smooth table or tray, press down on suction cup, and lower lever.
Place arm on arm Anchor pad and secure it with arm and wrist straps. If
suction cup does not hold well, wipe with a damp cloth or apply a thin layer
of petroleum jelly around the rim of the rubber pad to enhance the vacuum.
Adjustable abduction wedge
Attaching
Figure 16a: Remove the knobs (A) and one washer off each threaded stud.
- Place abduction wedge in appropriate slot with one plastic washer on each threaded stud between wedge and stander.
- Put the second washer on each of the threaded studs between knobs and stander.
- Thread and tighten knobs securely.
Adjusting
Figure 16a: To adjust the amount of abduction, raise one wing of the
wedge to expose the adjustment screw (B). Turn screw out to increase
abduction. Raise the second wing, and adjust it in the same way. Round
abductor
Figure 16b: To attach the round abductor, remove the knob and place where
appropriate. Re-tighten knobs securely. Collar
Figure 16c: To attach the collar, wrap it around the round abductor, and
fasten the snaps.
Operation
- To prevent falls and injury, lock all four casters.
- Crank the Supine Stander to desired horizontal height for ease of transfer.
- Slide the armrest and the lateral away from the transfer area. Open the chest, hip and leg straps.
- Make sure that both the headboard and footboard are in place before transferring the user.
- Transfer the user onto the Supine Stander in a horizontal position, and make final adjustments to all straps, ensuring that the safety closures on the straps are secure.
- Adjust the head block and laterals.
- Secure the user’s feet in the sandals and adjust positioning.
- Adjust the armrests to the user’s elbow height, and set to the appropriate angle.
- Make any final adjustments, ensuring that everything is in place, and that all knobs are tight and secure.
- Crank the Supine Stander to the desired angle as indicated on the board angle label.
- Attach tray on armrests if desired.
Maintenance
This product is designed and tested for an expected life of 5 years when used
and maintained in accordance with this manual. At all times, users must ensure
that the product remains in a safe and useable condition, including regular
maintenance and inspections as specified in the manual.
To prevent structural failure, which may result in serious injury or
death:
- Inspect this product and components regularly for loose or missing screws, metal fatigue, cracks, broken welds, missing attachments, general instability or other signs of excessive wear.
- Immediately remove this product from use when any condition develops that might make operation unsafe.
- Do not use Rifton components or products for any purpose other than their intended use.
- Replace or repair components or products that are damaged or appear to be unstable.
- Use only Rifton authorized replacement parts. Order information for replacement parts is provided on the back of this product manual.
- Twice a year lubricate all four casters. Apply a drop of light oil to the ball bearings and to the axle on each side of the caster wheel.
- Once a year, check the lubrication of the threaded rod. If it seems dry, apply a general purpose grease to the length of the rod.
Cleaning
- Clean with disinfectant wipes or a solution of up to 10% bleach. Do not use excessive water on pads and wooden surfaces.
- Clean tray with mild soap or detergent. Avoid abrasive cleaners and aromatic spirits. Disinfectants can be used, but no solvents. Never scrape with razor blades or other sharp instruments. Do not clean in hot sun or at elevated temperatures. Do not send through a dishwasher.
- The components of the arm Anchor can be removed for cleaning. Wipe clean with disinfectant wipes or a solution of up to 10% bleach.
Warranty Statement
- If a Rifton product breaks or fails in service during the first year, we will replace it free of charge.
Materials
- Steel hardware items (nuts, bolts, screws, etc) are typically zinc or nickel plated, or stainless steel.
- Upholstery items (pads, support blocks, padded prompts, etc) are typically polyurethane foam with a fire-retardant cover made from expanded vinyl.
- Frames are typically steel or aluminum tubing, welded together, and coated with a baked-on paint finish. Some frame components may also be stainless steel.
- Straps are typically made of polypropylene or nylon webbing.
- Wooden components are typically birch plywood, solid maple, or laminated hardwood veneers, finished with a clear polyurethane lacquer.
- Plastic components are typically injection molded from a variety of industrial resins.
- All materials are latex, lead and phthalates free.
User modifications
WARNING
To prevent serious injury or death, do not modify or alter Rifton products or components, or use Rifton products or components in conjunction with products from other manufacturers. Rifton does not accept responsibility for any modifications or alterations made to our components or products after they leave our premises. Customers modifying or altering our components or products, or using them in conjunction with products from other manufacturers, do so at their own risk.
Rifton Contact Information
-
Emailsales@rifton.com
-
Phone 800.571.8198 9–5 EST
-
Mail Rifton Equipment
PO Box 260 Rifton NY 12471–0260 -
Fax 800.865.4674
-
Onlinewww.rifton.com
To order replacement parts
- Locate the ID number of the product on the small white label.
- Have this number available when you call 800.571.8198 for your customer service representative.
Use only replacement parts supplied by Rifton Equipment.
We are glad to supply replacement parts. Although Rifton makes every effort to
supply the correct parts and instructions for repairing or refurbishing your
equipment, you are responsible to make sure that the repairs or modifications
are correctly and safely completed.
Find letters of medical necessity and studies on the therapeutic benefits of
standing at:
www.rifton.com/supinestander
Documents / Resources
|
rifton E420 Supine
Stander
[pdf] Instruction Manual
E420, E430, E420 Supine Stander, Supine Stander, Stander
---|---
References
- Rifton | Adaptive Equipment
- Rifton | Product manual listing
- Rifton | Supine standing frames for adults and children are designed for partial weight-bearing in the supine position
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>