OHAUS Defender 3000 Series Base Scale Instruction Manual
- June 12, 2024
- Ohaus
Table of Contents
OHAUS Defender 3000 Series Base Scale Instruction Manual
Compliance
This product conforms to the EMC directive 8 9 /3 3 6 /EEC and the Low Voltage Directive 7 3 /2 3 /EEC. The complete declaration of Conformity is available from Ohaus Corporation
Disposal
In conformance with the European Directive 2002/96 EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to
ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the
Ohaus quality management system is compliant with the ISO 9001 standard’s
requirements. On May 15, 2003, Ohaus Corporation, USA, was re-registered to
the ISO 9001:2000 standard.
INTRODUCTION
This manual covers installation, and maintenance instructions for the Ohaus Defender 3000 Series Base. Please read this manual completely before installation and operation.
SAFETY PRECAUTIONS
For safe and dependable operation of this product, please comply with the following precautions:
- Operate the base only under ambient conditions specified in these instructions
- Ensure that the load cell cable does not pose an obstruction or tripping hazard
- Do not operate the base in hazardous environments or unstable locations
- Do not drop loads on the base
- Do not lift the base by the top frame; always lift from the bottom frame when moving the base
- Service should only be performed by authorized personnel
INSTALLATION
Unpacking
Unpack and inspect the product to make sure that all components have been included. The package includes the following:
- Base
- Weighing Pan
- Warranty Card
- Instruction Manual
When purchased as a complete Defender 3000 Series Scale, will also include:
- 3000 Series Indicator
- Column Assembly
Assembly
Weighing Pan
Place the weighing pan securely over the rubber load pads on the top frame of
the base.
Wiring Connections
Connect the load cell cable to an indicator using the wiring codes in Table 1
below. Refer to the indicator manual for information about load cell
connections, setup and calibration
Note: When purchased as a Defender Series Scale, the base is already pre- wired to the 3000 Series Indicator.
TABLE 1. SIX WIRE LOAD CELL CONNECTION.
FUNCTION | WIRE COLOR |
---|---|
+ Excitation | Green |
– Excitation | Black |
+ Signal | Red |
– Signal | White |
+ Sense | Blue |
– Sense | Brown |
Shield | Yellow |
Selecting the Location
To ensure accuracy, proper performance and safety, locate and operate the base on a stable, level surface. Avoid locations with rapid temperature changes or excessive dust, air currents, vibrations, electromagnetic fields or heat
Level the base by adjusting the four leveling feet until the bubble in the level indicator (located at the rear of the base) is centered. A wrench may be needed to loosen the locking nut above each leveling foot. When the base is level, retighten the locking nuts up against the base to lock each foot into place.
Note: Ensure that the base is level each time its location is changed.
CORRECT
INCORRECT
MAINTENANCE
Cleaning
The base components should be kept clean and free of excessive material build up.
- Damp cloth with water and a mild detergent may be used to wipe clean the external surfaces—do not use acids, alkalis, strong solvents or abrasive materials and chemicals
Troubleshooting
Aside from installing components and leveling adjustments, the Defender Series Base does not require any other adjustments as shipped from the factory.
Operational difficulties that may be encountered can often be traced to simple causes such as:
- Loose or incorrect wiring connections
- Obstructions to the base frame
- Unstable environments
- Incorrect calibration or setup of the indicator
If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus service agent. For service assistance or technical support in the United States call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM EST. An Ohaus product service specialist will be available to provide assistance. Outside the USA, please visit our web site, www.ohaus.com to locate the Ohaus office nearest you.
TECHNICAL DATA
Technical Specifications
- The technical data is valid under the following ambient conditions:
- Ambient temperature: -10ºC to +40 ºC
- Relative humidity: 10% to 90% relative humidity, non-condensing
- Height above sea level: Up to 4,000m
- Operability is assured at ambient temperature: -10ºC to +40 ºC
SPECIFICATIONS
MODEL| D30BR| D60BR| D60BL| D150BL| D150BX|
D300BX
---|---|---|---|---|---|---
Capacity| 30 kg| 60 kg| 60 kg| 150 kg| 150 kg| 300 kg
Maximum Resolution| 6000d| 6000d| 6000d| 7500d| 7500d| 6000d
Pan Dimensions| 305 x 355 x 75 mm| 420 x 550 x 80 mm| 500 x 650 x 90 mm
Base Construction| Stainless Steel platform with painted steel frame and
rubber leveling feet
Safe Overload Capacity| 125% of capacity
Repeatability (std. deviation)| 1d
Linearity| ±1d
Load Cell Cable| 2 m L x 6-wire| 2.3 m L x 6-wire
Load Cell Type| 350 Ohm, aluminum, single point
Load Cell Excitation| 5-15V DC/AC
Load Cell Rated Output| 2mV/V
Load Cell Protection| IP67
Net Weight| 9 kg| 16 kg| 28 kg
Shipping Weight| 11 kg| 19 kg| 32 kg
MODEL| D30VR| D60VR| D60VL| D150VL| D150VX| D300VX
Capacity| 30 kg| 60 kg| 60 kg| 150 kg| 150 kg| 300 kg
Maximum Resolution| 6000d| 6000d| 6000d| 7500d| 7500d| 6000d
Pan Dimensions| 305 x 355 x 75 mm| 420 x 550 x 80 mm| 500 x 650 x 90 mm
Base Construction| Stainless Steel platform with stainless steel frame and
rubber leveling feet
Safe Overload Capacity| 125% of capacity
Repeatability (std. deviation)| 1d
Linearity| ±1d
Load Cell Cable| 2 m L x 6-wire| 2.3 m L x 6-wire
Load Cell Type| 350 Ohm, stainless steel, single point
Load Cell Excitation| 5-15V DC/AC
Load Cell Rated Output| 2mV/V
Load Cell Protection| IP67
Net Weight| 9 kg| 21 kg| 33 kg
Shipping Weight| 11 kg| 24 kg| 37 kg
Drawings
Figure 1. Defender Base Dimension Drawing.
TABLE 3. DIMENSIONS.
Base
| A| B| C| D| E
---|---|---|---|---|---
Pan Depth| Pan Width| Height of Pan to Surface of Table*| Height of Pan| Depth
Including Level Indicator
D30BR, D60BR| 355 mm| 305 mm| 123 mm| 75 mm| 380 mm
D60BL, D150BL| 550 mm| 420 mm| 136 mm| 80 mm| 575 mm
D150BX, D300BX| 650 mm| 500 mm| 143 mm| 90 mm| 675 mm
D30VR, D60VR| 355 mm| 305 mm| 119 mm| 75 mm| 380 mm
D60VL, D150VL| 550 mm| 420 mm| 145 mm| 80 mm| 575 mm
D150VX, D300VX| 650 mm| 500 mm| 154 mm| 90 mm| 675 mm
*For leveling purposes, the feet may be extended up to an additional 11mm.
TABLE 4. ACCESSORIES.
Description | Part Number |
---|---|
Column Kit, 35 cm painted steel | 80251743 |
Column Kit, 70 cm painted steel | 80251744 |
Column Kit, 35 cm stainless steel | 80251745 |
Column Kit, 70 cm stainless steel | 80251746 |
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus. This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages. As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for further details.
Costumer Support
Ohaus Corporation
19A Chapin Road
P.O. Box 2033
Pine Brook, NJ 07058-2033, USA
Tel: 973-377-9000
Fax: 973-944-7177
With offices worldwide
www.ohaus.com
References
- Lab Balance | Industrial Scales | Lab Equipment | Lab Instruments | OHAUS
- Lab Balance | Industrial Scales | Lab Equipment | Lab Instruments | OHAUS
- Manual-Hub.com – Free PDF manuals!
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>