OHAUS Defender 6000 Series Washdown Base
Defender TM 6000 Series Base
This manual covers installation and maintenance instructions for the OHAUS
Defender TM Series Base. Please read this manual completely before
installation and operation.
Safety Precautions
Before installation and operation, please ensure that you have read and
understood the safety precautions in the manual.
Installation
Unpacking
Unpack and inspect the product to make sure that all components have been
included. The package includes:
- The Defender Series Base
- Instruction manual
When purchased as a complete Defender Series Scale, the package will also include:
- The Defender Series Scale
- Power cord
- Indicator
Assembly
Weighing Pan
Remove the shipping spacers, and place the weighing pan securely over the
rubber load pads on the top frame of the base.
Note: Not included with some models.
Wiring Connections
When connecting the base to an OHAUS indicator equipped with the matching
circular connector, push the base connector onto the indicator connector and
turn the locking ring clockwise to lock it in place.
When connecting the base to an indicator that does not have the matching
connector, install the optional load cell cable adapter (P/N 30429931 sold
separately) to the indicator using the wiring
codes in Table 1. Then attach the load cell cable to the cable adapter.
Note: As an alternate wiring method, cut off the connector, strip the wires and attach them to the indicator using the wiring codes in Table 1. The extension cable accessory (P/N: 30429931) cannot be used if the scale will be used in a legal for trade or legally controlled application, because there is no provision to seal the connection between the indicator and platform.
Selecting the Location
To ensure accuracy, proper performance, and safety, locate and operate the
base on a stable, level surface. Avoid locations with rapid temperature
changes or excessive dust, air currents, vibrations, electromagnetic fields,
or heat.
Level the base by adjusting the four leveling feet until the bubble in the level indicator is centered. A wrench may be needed to loosen the locking nut above each leveling foot. When the base is level, retighten the locking nuts up against the base to lock each foot into place.
Note: Ensure that the base is level each time its location is changed.
Maintenance
Cleaning
To maintain surface integrity, do not allow debris to remain on the surface of
the painted or stainless steel base frame.
Remove the pan and use fresh water to clean the base frame
surface of any debris. Dry the base with a clean cloth or allow it
to air dry before replacing the pan. Repeat daily for best
care.
Troubleshooting
Aside from installing components and leveling adjustments, the Defender Series Base does not require any other adjustments as shipped from the factory.
Load Cell Connection Table
Type 1 | Type 2 |
---|---|
+ Excitation | + Excitation |
– Excitation | – Excitation |
+ Signal | + Signal |
– Signal | – Signal |
+ Sense | + Sense |
– Sense | – Sense |
Sense Shield | Deep Yellow |
Wire Color: Blue Black White Red Green Grey Yellow | Wire Color: Green Black |
White Red Light Yellow Blue Deep
Yellow
INTRODUCTION
This manual covers installation, and maintenance instructions for the OHAUS
Defender TM Series Base. Please read this manual completely before
installation and operation.
SAFETY PRECAUTIONS
For safe and dependable operation of this product, please comply with the
following precautions:
- Operate the base only under ambient conditions specified in these instructions Ensure that the load cell cable does not pose an obstruction or tripping hazard
- Do not operate the base in hazardous environments or unstable locations
- Do not drop loads on the base
- Do not lift the base by the top frame; always lift from the bottom frame when moving the base
- Service should only be performed by authorized personnel
PRODUCT INSTALLATION
Unpacking
Unpack and inspect the product to make sure that all components have been
included. The package includes the following:
- Defender TM Series Base
- Weighing Pan
- Instruction Manual
When purchased as a complete Defender Series Scale, the package will also include:
- Indicator
- Column Assembly
ASSEMBLY
Weighing Pan
Remove the shipping spacers, and place the weighing pan securely over the
rubber load pads on the top frame of the base.
Note: Not included with some models
Wiring Connections
When connecting the base to an OHAUS indicator equipped with the matching
circular connector, push the base connector onto the indicator connector and
turn the locking ring clockwise to lock it in place.
When connecting the base to an indicator that does not have the matching
connector, install the optional load cell cable adapter (P/N 30429931 sold
separately) to the indicator using the wiring codes in Table 1. Then attach
the load cell cable to the cable adapter.
Note: As an alternate wiring method, cut off the connector, strip the
wires and attach them to the indicator using the wiring codes in Table 1.
The extension cable accessory (P/N: 30429931) cannot be used if the scale will
be used in a legal for trade or legally controlled application, because there
is no provision to seal the connection between the indicator and platform.
TABLE 1. LOAD CELL CONNECTION
Type1. For model i-D3K1S, i-D6K1S, i-D15K1R, i-D30K1R, i-D60K1L bases
FUNCTION | WIRE COLOR |
---|---|
+ Excitation | Blue |
– Excitation | Black |
+ Signal | White |
– Signal | Red |
+ Sense | Green |
– Sense | Grey |
Shield | Yellow |
Type2. For model i-D150K1L base
FUNCTION | WIRE COLOR |
---|---|
+ Excitation | Green |
– Excitation | Black |
+ Signal | White |
– Signal | Red |
+ Sense | Light Yellow |
– Sense | Blue |
Shield | Deep Yellow |
Selecting the Location
To ensure accuracy, proper performance and safety, locate and operate the base
on a stable, level surface. Avoid locations with rapid temperature changes or
excessive dust, air currents, vibrations, electromagnetic fields or heat.
Level the base by adjusting the four leveling feet until the bubble in the
level indicator is centered. A wrench may be needed to loosen the locking nut
above each leveling foot. When the base is level, retighten the locking nuts
up against the base to lock each foot into place.
Note: Ensure that the base is level each time its location is changed.
PRODUCT MAINTENANCE
Cleaning
The base components should be kept clean and free of excessive material build
up.
- A damp cloth with water and a mild detergent may be used to wipe clean the external surfaces – do not use acids, alkalis, strong solvents or abrasive materials.
To maintain surface integrity, do not allow debris to remain on the surface of the painted or stainless steel base frame.
- Remove the pan and use fresh water to clean the base frame surface of any debris. Dry the base with a clean cloth or allow to air dry before replacing the pan. Repeat daily for best care.
Product Troubleshooting
Aside from installing components and leveling adjustments, the Defender Series
Base does not require any other adjustments as shipped from the factory.
Operational difficulties that may be encountered can often be traced to simple
causes such as:
- Loose or incorrect wiring connections Failure to remove the shipping spacers Obstructions to the base frame
- Unstable environments
- Incorrect calibration or setup of the indicator
If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS service agent. For service assistance or technical support in the United States, call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM EST. An OHAUS product service specialist will be available to provide assistance. Outside the USA, please visit our web site, www.ohaus.com to locate the OHAUS office nearest you.
TECHNICAL DATA
Technical Specifications
The technical data is valid under the following ambient conditions:
- Operating temperature: -10ºC to +40 ºC (14 °F to 104 °F)
- Relative humidity: 100% relative humidity, non-condensing
- Altitude: Up to 2,000m
Note: Not all models may be available in your area. Check with your local OHAUS representative.
TABLE 2.1 SPECIFICATIONS
MODEL| i-D3K1S| i-D6K1S| i-D15K1R| i-D30K1R|
i-D60K1L| i-D150K1L
---|---|---|---|---|---|---
Capacity| 3 kg| 6 kg| 15 kg| 30 kg| 60 kg| 150 kg
Approved Resolution| OIML 3000e
Safe Overload Capacity| 150% of capacity
Pan Dimensions| 254 x 254 mm
10 x 10 in
| 305 x 305 mm
12 x 12 in
| 400 x 500 mm
15.7 x 19.6 in
Base Construction
| ****
316 stainless steel top and bottom frame, hermetically sealed IP68/IP69K stainless steel loadcell.
| 316 stainless steel removable weighing platform with 316 stainless steel frame, hermetically sealed IP68/IP69K stainless steel
loadcell.
Repeatability (std.deviation)| 2d
Linearity| ±2d
Load Cell Capacity| 10 kg| 10 kg| 20 kg| 50 kg| 100 kg| 200 kg
Load Cell Cable| 50cm L x 6-wire| 1m L x 6-wire
Load Cell Type| 350 Ohm, SST, single point
Load Cell Excitation| 3.3-15V DC/AC
Load Cell Rated Output| 2mV/V
Load Cell Protection| IP69K
Net Weight| 6 kg / 13.2 lb| 7.5 kg / 16.5 lb| 16 kg / 35.3 lb
Shipping Weight| 7 kg / 15.4 lb| 8.5 kg / 18.7 lb| 18 kg / 39.7 lb
Drawings
Figure 1. Defender Base Dimension Drawing. TABLE 3. DIMENSIONS
Base
| A| B| C| D
---|---|---|---|---
Pan Depth| Pan Width| Height of Pan to Surface of Table|
Height of Pan
i-DxxxKxS| 254 mm| 254 mm| 106 mm| N/A
i-DxxxKxR| 305 mm| 305 mm| 106 mm| N/A
i-DxxxKxL| 400 mm| 500 mm| 136.5 mm| 80 mm
*For leveling purposes, the feet may be extended up to an additional 15 mm.
COMPLIANCE
Compliance to the following standards is indicated by the corresponding mark
on the product.
Mark | Standard |
---|---|
This product complies with the applicable harmonized standards of EU |
Directives 2011/65/EU (RoHS). The EU Declaration of Conformity is available
online at www.ohaus.com/ce.
| This product complies with the EU Directive 2012/19/EU (WEEE). Please
dispose of this product in accordance with local regulations at the collecting
point specified for electrical and electronic equipment.
For disposal instructions in Europe, refer to www.ohaus.com/weee.
ISO 9001 Registration
The management system governing the production of this product is ISO 9001
certified.
ACCESSORIES
TABLE 4. ACCESSORIES
Part Number | Description |
---|---|
30616409 | Wheel Kit, Defender Series |
30626654 | Mounting Set Front SST |
30572907 | External power bank sealing kit |
30424409 | Cable Extension 9m D52 |
30429931 | Cable Kit Extension for base |
30572908 | Column Kit 260mm 316 D61 |
30572909 | Column Kit 670mm 316 D61 |
LIMITED WARRANTY
OHAUS products are warranted against defects in materials and workmanship from
the date of delivery through the duration of the warranty period. During the
warranty period OHAUS will repair, or, at its option, replace any component(s)
that proves to be defective at no charge, provided that the product is
returned, freight prepaid, to OHAUS. This warranty does not apply if the
product has been damaged by accident or misuse, exposed to radioactive or
corrosive materials, has foreign material penetrating to the inside of the
product, or as a result of service or modification by other than OHAUS. In
lieu of a properly returned warranty registration card, the warranty period
shall begin on the date of shipment to the authorized dealer. No other express
or implied warranty is given by OHAUS Corporation. OHAUS Corporation shall not
be liable for any consequential damages.
As warranty legislation differs from state to state and country to country,
please contact OHAUS or your local OHAUS dealer for further details