OHAUS Defender Series Bench Scale Bases User Guide

June 12, 2024
Ohaus

Defender® Series Base

User Guide

INTRODUCTION

This manual covers installation, and maintenance instructions for the Ohaus Defender® Series Base. Please read this manual completely before installation and operation.

SAFETY PRECAUTIONS

For safe and dependable operation of this product, please comply with the following precautions:

  • Operate the base only under ambient conditions specified in these instructions.
  • Ensure that the load cell cable does not pose an obstruction or tripping hazard.
  • Do not operate the base in hazardous environments or unstable locations.
  • Do not drop loads on the base.
  • Do not lift the base by the top frame; always lift from the bottom frame when moving the base.
  • Service should only be performed by authorized personnel.
INSTALLATION
Unpacking

Unpack and inspect the product to make sure that all components have been included. The package includes the following:

  • Defender™ Series Base
  • Weighing Pan
  • User Guide

When purchased as a complete Defender Series Scale, the package will also include:

  • Indicator
  • Column Assembly
ASSEMBLY
Weighing Pan

Remove the shipping spacers, and place the weighing pan securely over the rubber load pads on the top frame of the base.

Note: Not included with some models.

Wiring Connections

When connecting the base to an Ohaus indicator equipped with the matching circular connector, push the base connector onto the indicator connector and turn the locking ring clockwise to lock it in place.

When connecting the base to an indicator that does not have the matching connector, install the optional load cell cable adapter (P/N 30429931sold separately) to the indicator using the wiring codes in Table 1. Then attach the load cell cable to the cable adapter.

Note: As an alternate wiring method, cut off the connector, strip the wires and attach them to the indicator using the wiring codes in Table 1.

  • Detailed instructions that describe how to prepare the load cell cable for connection to the terminal block inside an indicator.
  • The extension cable accessory (P/N: 30429931) cannot be used if the scale will be used in a legal for trade or legally controlled application, because there is no provision to seal the connection between the indicator and platform.

TABLE 1. LOAD CELL CONNECTION

Type1. DxxxRxx Bases
DxxxWxx Bases

FUNCTION WIRE COLOR
+ Excitation  Green
– Excitation Black
+ Signal White
– Signal  Red
+ Sense Blue
– Sense Brown
Shield Yellow
Selecting the Location

Level the base by adjusting the four leveling feet until the bubble in the level indicator is centered.

Note: Ensure that the base is level each time its location is changed.

CORRECT     INCORRECT

MAINTENANCE
Cleaning

The base components should be kept clean and free of excessive material build up.

  • A damp cloth with water and a mild detergent may be used to wipe clean the external surfaces ­- do not use acids, alkalis, strong solvents or abrasive materials.

To maintain surface integrity, do not allow debris to remain on the surface of the painted or stainless steel base frame.

  • Remove the pan and use fresh water to clean the base frame surface of any debris. Dry the base with a clean cloth or allow to air dry before replacing the pan. Repeat daily for best care.
Troubleshooting

An Ohaus product service specialist will be available to provide assistance. Outside the USA, please visit our web site, www.ohaus.com to locate the Ohaus office nearest you.

TECHNICAL DATA
Technical Specifications

The technical data is valid under the following ambient conditions:
Ambient temperature: -10ºC to +40 ºC
Relative humidity: 10% to 90% relative humidity, non-condensing Altitude: Up to 2,000 m
Operability is assured at ambient temperature -10ºC to +40 ºC

Note: Not all models may be available in your area. Check with your local Ohaus representative.

TABLE 1 SPECIFICATIONS

MODEL|

D6RQDR

| D15RQDR| D15RQDL| D30RQDL| D60RQDL| D60RTDL|

D150RTDL

---|---|---|---|---|---|---|---
Capacity|

3 / 6 kg

| 6 / 15 kg| 6 / 15 kg| 15 / 30 kg| 30 / 60 kg| 30 / 60 kg|

60 / 150 kg

Approved Resolution|

OIML 2 x 3000e

Safe Overload Capacity|

150% of capacity

Pan Dimensions|

305 x 305 mm
12 x 12 in

| 400 x 400 mm
15.7 x 15.7 in|

400 x 500 mm
15.7 x 19.6 in

Base Construction|

Stainless Steel platform with painted steel frame and rubber leveling feet

Repeatability (std.deviation)|

2d

Linearity|

± 2d

Load Cell Capacity|

11 kg

| 22 kg| 22 kg| 50 kg| 100 kg| 100 kg|

250 kg

Load Cell Cable|

2 m L x 6-wire

Load Cell Type|

350 Ohm, aluminum, single point

Load Cell Excitation|

5-15V DC/AC

Load Cell Rated Output|

2mV/V

Load Cell Protection|

IP67

Net Weight |

6 kg

| 11.5 kg|

17 kg

Shipping Weight|

7 kg

| 13.5 kg|

19 kg

TABLE 2 SPECIFICATIONS

MODEL| D60RQDX| D150RQDX| D60RQDV| D150RQDV| D300RQDV| D150RTDV| D300RTDV| D600RTDV
---|---|---|---|---|---|---|---|---
Capacity| 30 / 60 kg| 60 / 150 kg| 30 / 60 kg| 60 / 150 kg| 150 / 300 kg| 60 / 150 kg| 150 / 300 kg| 300 / 600 kg
Approved Resolution| OIML 2 x 3000e
Safe Overload Capacity| 150% of capacity
Pan Dimensions|

500 x 500 mm
19.6 x 19.6 in

| 610 x 610 mm
24 x 24 in|

600 x 800 mm
23.6 x 31.5 in

Base Construction|

Stainless Steel platform with painted steel frame and rubber leveling feet
2d

Repeatability (std.deviation)
Linearity| ± 2d
Load Cell Capacity| 100 kg| 250 kg| 100 kg| 250 kg| 500 kg| 250 kg| 500 kg | 750 kg
Load Cell Cable| 2 m L x 6-wire
Load Cell Type | 350 Ohm, aluminum, single point
Load Cell Excitation| 5-15V DC/AC
Load Cell Rated Output| 2mV/V
Load Cell Protection| IP67
Net Weight | 19 kg| 31 kg| 39.2 kg
Shipping Weight| 21 kg| 34 kg| 42.7 kg

TABLE 3 SPECIFICATIONS

MODEL| D6WQDR| D15WQDR| D15WQDL| D30WQDL| D60WQDL| D60WTDL| D150WTDL
---|---|---|---|---|---|---|---
Capacity| 3 / 6 kg| 6 / 15 kg| 6 / 15 kg| 15 / 30 kg| 30 / 60 kg| 30 / 60 kg| 60 / 150 kg
Approved Resolution| OIML 2 x 3000e
Safe Overload Capacity| 150% of capacity
Pan Dimensions|

305 x 305 mm
12 x 12 in

| 400 x 400 mm
15.7 x 15.7 in|

400 x 500 mm
 15.7 x 19.6 in

Base Construction| Stainless Steel platform with stainless steel frame and rubber leveling feet
Repeatability (std.deviation)| 2d
Linearity| ± 2d
Load Cell Capacity| 11 kg| 22 kg| 22 kg| 50 kg| 100 kg| 100 kg| 250 kg
Load Cell Cable| 2 m L x 6-wire
Load Cell Type| 350 Ohm, aluminum, single point
Load Cell Excitation| 5-15V DC/AC
Load Cell Rated Output| 2mV/V
Load Cell Protection| IP67
Net Weight| 5 kg| 13 kg| 15.5 kg
Shipping Weight| 6 kg| 14 kg| 18.5 kg

TABLE 4 SPECIFICATIONS

MODEL| D60WQDX| D150WQDX| D60WQDV| D150WQDV| D300WQDV| D150WTDV| D300WTDV| D600WTDV
---|---|---|---|---|---|---|---|---
Capacity| 30 / 60 kg| 60 / 150 kg| 30 / 60 kg| 60 / 150 kg| 150 / 300 kg| 60 / 150 kg| 150 / 300 kg| 300 / 600 kg
Approved Resolution | OIML 2 x 3000e
Safe Overload Capacity | 150% of capacity
Pan Dimensions|

500 x 500 mm
19.6 x 19.6 in

| 610 x 610 mm
24 x 24 in|

600 x 800 mm
23.6 x 31.5 in

Base Construction| Stainless Steel platform with stainless steel frame and rubber leveling feet
Repeatability (std.deviation)| 2d
Linearity| ± 2d
Load Cell Capacity| 100 kg| 250 kg| 100 kg| 250 kg| 500 kg| 250 kg| 500 kg| 750 kg
Load Cell Cable| 2 m L x 6-wire
Load Cell Type| 350 Ohm, aluminum, single point
Load Cell Excitation| 5-15V DC/AC
Load Cell Rated Output| 2mV/V
Load Cell Protection| IP67
Net Weight| 16.5 kg| 28 kg | 36.5 kg
Shipping Weight | 19.5 kg| 32.5 kg | 41 kg

COMPLIANCE

| This product complies with the applicable harmonized standards of EU Directive 2011/65/EU (RoHS).
The EU Declaration of Conformity is available online at www.ohaus.com/ce.
---|---
| Disposal

In conformance with the European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.

Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.

If you have any questions, please contact the responsible authority or the distributor from which you purchased this device.

Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.

For disposal instructions in Europe, refer to www.ohaus.com/eee.

Thank you for your contribution to environmental protection.

INSTRUCTION MANUAL

For more detailed information about the scale, please refer to the product instruction manual. The instruction manual can be obtained either by scanning the QR code or by going to the product page and from there download it.

Go to www.ohaus.com to register your product’s warranty.

QR Code:

Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054, USA
Tel: 973-377-9000
Fax: 973-944-7177

With offices worldwide
www.ohaus.com

P/N 30428871A © 2018 Ohaus Corporation, all rights reserved.

Printed in China

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals