Wondfo WV01P0001 COVID-19 Influenza A and B Test User Guide
- August 7, 2024
- Wondfo
Table of Contents
Wondfo WV01P0001 COVID-19 Influenza A and B Test
Specifications
- Product Name: COVID-19 / Influenza A&B Home Test
- Model Number: WV01P0001
- Use: Emergency Use Authorization (EUA) for in Vitro Diagnostic Use with anterior nasal swab specimens
- Components: Test cassettes, buffer tubes, swabs, tube holder, quick reference instructions
Product Information
Intended Use
The COVID-19 / Influenza A&B Home Test is designed for the detection of
COVID-19 and influenza viruses using anterior nasal swab specimens.
Principle of Procedure
The test detects the presence of influenza A&B and COVID-19 viruses in the
specimen provided.
Reagents and Materials
The kit includes test cassettes, buffer tubes, swabs, tube holder, and
quick reference instructions for usage.
Storage and Stability
Store the kit components in a cool, dry place and use them before the
expiration date mentioned on the package.
Quality Control
Each test has a built-in internal procedural control to verify the proper
assembly and functioning of the test strip.
Product Usage Instructions
Preparing for the Test
- Check the expiration date on the outer box.
- Wash hands thoroughly or use hand sanitizer.
- Insert buffer tube into the tube holder.
- Remove test cassette from pouch and lay it flat.
Sample Collection
- Remove the swab from the pouch, avoiding contact with the tip.
FAQ
-
Q: What should I do if the control line does not appear after performing the test?
- A: If the control line does not appear, the test result is invalid. Perform a new test using a new swab and test kit.
-
Q: How should I store the kit components?
- A: Store the components in a cool, dry place and use them before the expiration date mentioned on the package.
COVID-19 / Influenza A &B Home Test
HEALTHCARE PROVIDER
INSTRUCTIONS FOR USE
- For use under Emergency Use Authorization (EUA) only
- For in Vitro Diagnostic Use
- For use with anterior nasal swab specimens
INTENDED USE
- The WELLlife™ COVID-19 / Influenza A&B Home Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens.
- This test is authorized for non-prescription home use with self-collected anterior nasal swab specimens from individuals aged 14 years or older, or with adult-collected anterior nasal swab specimens from individuals aged two (2) years or older. This test is only authorized for individuals with signs and symptoms of respiratory infection consistent with COVID-19 within the first five (5) days of symptom onset when tested at least twice over three days with at least 48 hours between tests.
- Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar.
- Results are for the simultaneous identification of SARS-CoV-2, influenza A virus, and influenza B virus proteins antigens, but do not differentiate between SARS-CoV and SARS-CoV-2 viruses and are not intended to detect influenza C antigens.
- The viral antigens targeted by this test are generally detectable from specimens collected
using nasal swabs during the acute phase of infection. Positive results indicate the presence of viral antigens, but clinical correlation with patient history and other diagnostic information is necessary to determine infection status.
-
Positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definitive cause of disease. Individuals who test positive with the WELLlife™ COVID-19 / Influenza A&B Home Test should self-isolate and seek follow-up care with their physician or healthcare provider as additional testing may be necessary.
-
All negative results are presumptive and confirmation with a molecular assay, if necessary for patient management, may be performed. Negative results do not rule out SARS-CoV-2, influenza A, and influenza B infection and should not be used as the sole basis for treatment or patient management decisions, including infection control decisions such as isolating from others and wearing masks. Negative results should be considered in the context of an individual’s recent exposures, history, and the presence of clinical signs and symptoms consistent with SARS-CoV-2, influenza A, and influenza B infection.
-
Individuals who test negative and continue to experience SARS-CoV-2 and/or influenza-like symptoms of fever, cough, and/or shortness of breath may still have SARS-CoV-2 and/or influenza infection and should seek follow up care with their physician or healthcare provider.
-
The WELLlife™ COVID-19 / Influenza A&B Home Test is only for use under the Food and Drug Administration’s Emergency Use Authorization.
SUMMARY AND EXPLANATION
Along with the common cold, influenza is one of the most common acute respiratory infections, producing symptoms such as headache, chills, dry cough, body aches, and fever. It affects 5%-20% of the United States population annually, resulting in more than 200,000 hospitalizations and 36,000 deaths.[1] The influenza A virus is typically more prevalent and is associated with the most serious influenza epidemics, while influenza B infections usually present with milder symptoms. Diagnosis is difficult because the initial symptoms can be similar to those caused by other infectious agents. Considering that the influenza virus is highly contagious, accurate diagnosis and prompt treatment of patients can have a positive effect on public health. Accurate diagnosis and the ability to distinguish between A or B antigens can also help reduce the inappropriate use of antibiotics and gives the physician the opportunity to prescribe an antiviral therapy. Initiation of antiviral therapy should begin as soon as possible after onset, ideally within 48 hours of the appearance of symptoms, as treatment may reduce the duration of symptoms.[2]
Coronaviruses are enveloped RNA viruses that are found broadly among humans, other mammals, and birds. The viruses are known to cause mild symptoms, but sometimes severe respiratory, enteric, hepatic, and neurological diseases can occur. Seven coronavirus species are known to cause human disease, four of which (229E, OC43, NL63 and HKU-1) are quite prevalent and can cause mild cold symptoms, especially in immunocompetent people.[3] There are three other strains that are known to cause severe acute respiratory disease. These strains include severe acute respiratory syndrome coronavirus (SARS-CoV), Middle East respiratory syndrome coronavirus (MERS-CoV), and the 2019 Novel Coronavirus (COVID-19). These strains are all zoonotic in origin and have been linked to sometimes fatal respiratory illness. The prevalence of SARS and MERS has been quite low in recent years; the Novel Coronavirus (COVID-19) was recently identified in December 2019. The main manifestations of illness include fever, fatigue, and dry cough. Nasal congestion, runny nose, sore throat, myalgia, and diarrhea are found in a few cases. Most epidemiological studies suggest a 1–14-day incubation period. The median incubation period is estimated at 5.1 days, with most developing symptoms before 11.5 days.[4] Infected but asymptomatic people can also be an infectious source. The WELLlife™ COVID-19 / Influenza A&B Home Test can provide rapid detection of influenza A, influenza B, and/or SARS-CoV-2 viral antigens from symptomatic patients.
PRINCIPLE OF PROCEDURE
The WELLlife™ COVID-19 / Influenza A&B Home Test consists of a test cassette that separately detects influenza A, influenza B, and SARS-CoV-2 antigens. The test procedure requires the anterior nasal swab specimen to be inserted into the prefilled extraction buffer tube to be solubilized, and then the specimen is eluted. The virus particles in the specimen are disrupted by the chemicals in the extraction buffer, exposing internal viral nucleoproteins. After the release of specimen, the swab is discarded. The extracted specimen is then dropped into the sample well of the test cassette.
If SARS-CoV-2, influenza A and/or influenza B antigens are present in the specimen, they will react with SARS-CoV-2 antibody coupled to dye particles and/or influenza antibody coupled to dye particles, migrate through the membrane as antigen-antibody-dye complexes, bind to the immobilized capture antibody line(s) on the membrane, and generate a colored pink to red line in the specific test line position. The rest of the sample and rabbit IgG dye particle complexes continue to migrate to the Control line position (C), where immobilized goat anti-rabbit IgG will capture the rabbit IgG dye particle complexes and form the Control line. Formation of the pink to red Control line serves as an internal control to demonstrate that test reagents are functional, antibody-dye conjugates in the dye pad have been hydrated and released and that sufficient sample has been applied to allow for migration through the Test and Control lines. If the Control line does not appear within the designated incubation time, the result is invalid, and the test should be repeated using a new test device and specimen.
If the antigen level is equal to or above the detection limit, a visible colored band appears at the test region. Absence of this pink to red colored band in the test region of test strip and only a visible control line will appear, suggests a negative result.
WELLlife™ COVID-19 / Influenza A&B Home Test has three Test lines, one for COVID-19, one for influenza A and one for influenza B. The three Test lines allow for the separate and differential identification of COVID-19, influenza A and/or B from a single specimen. If any
Test line appears in the test result window, together with the Control line, the test result is Page 2 of 15 positive for COVID-19 and/or influenza.
REAGENTS AND MATERIALS
The WELLlife™ COVID-19 / Influenza A&B Home Test kit configurations are
indicated below
Components| 1
Test/kit| 2
Tests/kit| 5
Tests/kit| 10
Tests/kit| 25
Tests/kit
---|---|---|---|---|---
Sealed Test Cassettes| 1| 2| 5| 10| 25
Buffer Tubes| 1| 2| 5| 10| 25
Swabs| 1| 2| 5| 10| 25
Tube holder(Top right corner on Box)| 1| 1| 1| 1| 1
Quick Reference Instructions(QRI)| 1| 1| 1| 1| 1
Materials Required but Not Provided
Timer or watch
WARNINGS AND PRECAUTIONS
-
Read the instructions fully and carefully before performing the procedure. Failure to follow the instructions may result in inaccurate test results.
-
In the USA, this product has not been FDA cleared or approved, but has been authorized by FDA under an Emergency Use Authorization. This product has been authorized only for the detection of proteins from SARS-CoV-2, influenza A, and influenza B, not for any other viruses or pathogens. The emergency use of this product is only authorized for the duration of the declaration that circumstances exist justifying the authorization of emergency use of in vitro diagnostics for detection and/ or diagnosis of COVID-19 under Section 564(b)(1) of the Federal Food, Drug, and Cosmetic Act, 21 U.S.C.
§ 360bbb-3(b)(1), unless the declaration is terminated or authorization is revoked sooner. -
Serial testing should be performed in individuals with SARS-CoV-2 negative results at least twice over three days (with 48 hours between tests) for symptomatic individuals. You may need to purchase additional tests to perform this serial (repeat) testing.
-
Consistent with serial testing recommendations for SARS-CoV-2, for multi-analyte tests, symptomatic individuals who test positive for influenza A or B on the initial test but test negative for SARS-CoV-2 should be tested again in 48 hours to evaluate for co-infection with SARS-CoV-2 infection.
-
An anterior nasal swab sample can be self-collected by individuals aged 14 years and older. Children aged 2 to 13 years should be tested by an adult.
-
Do not use on anyone under 2 years of age.
-
Wear a safety mask or other face-covering when collecting a specimen from a child or another individual.
-
This test may only be used in symptomatic individuals.
-
Do not use if any of the test kit contents or packaging is damaged or open.
-
Test components are single-use. Do not re-use the test strip, buffer liquid, or swab.
-
If any liquid spills from the buffer tube, discard test components and re-start test using new test components.
-
Do not read test results before 10 minutes or after 20 minutes. Results read before 10 minutes or after 20 minutes may lead to a false positive, false negative, or invalid result.
-
If uncertain how to proceed, contact Technical Assistance at 1-888-444-3657.
-
Do not touch swab tip when handling the swab.
-
To ensure accurate test results, avoid contamination with liquid gel hand soap, hand sanitizer cream lotion, and fast-drying 80% ethanol hand sanitizer.
-
Do not open the test contents until ready for use, if the test cassette is open for an hour or longer, false test results may occur.
-
Testing should be performed in an area with good lighting.
-
Do not use the test kit after its expiration date.
-
Keep testing kit and kit components away from children and pets before and after use. Avoid contact with your skin, eyes, nose, or mouth. Do not ingest any kit components. The reagent solution contains harmful chemicals (see table below). If the solution contacts your skin, eyes nose, or mouth, flush with large amounts of water.
If irritation persists, seek medical advice: https://www.poisonhelp.org or 1- 800-222-1222
Chemical N ame| Harms (GHS Code) for each ingredient|
Concentration
---|---|---
ProClin 300| Causes skin irritation (H315)
Causes eye irritation (H320)| 0.05%
- For more information on EUAs please visit: https://www.fda.gov/emergency-preparedness-and-response/mcm-legal-regulatory-and-policy-framework/emergency-use-authorization.
- For the most up-to-date information on COVID-19, please visit: www.cdc.gov/COVID19.
STORAGE AND STABILITY
- Store the test kit between 36-86°F (2-30°C) in a place out of direct sunlight
- Reagents and devices must be used at room temperature (59-86°F/15-30°C)
- The unsealed cassette is valid for 1 hour. It is recommended to use the test kit immediately after opening. The expiration date is on the package.
QUALITY CONTROL
Each WELLlife™ COVID-19 / Influenza A&B Home Test has a built-in internal
procedural control. The red line appearing at the “C” position verifies proper
assembly and capillary flow of the test strip. A distinct red Control Line
should always appear if the test has been performed correctly. If the Control
Line does not appear, the test result is invalid, and a new test should be
performed using a new swab and new test kit.
PREPARING FOR THE TEST
NOTE: Do not open the test materials until ready for use. If the test
cassette is open for an hour or longer, invalid test results may occur.
-
Check the expiration date of the test printed on the outer box. For information about current expiration dates for at-home OTC COVID-19 diagnostic tests, visit http://www.fda.gov/covid-tests.
-
Wash your hands with soap and water for 20 seconds and dry them thoroughly, or use hand sanitizer.
-
Locate the tube holder on the box (look for the red circle on the kit’s box).
-
Insert the buffer tube into the tube holder. Ensure that the buffer tube is stable and upright.
- Remove the large cap from the buffer tube and set it aside for later use.
-
Remove test cassette from sealed pouch and lay it on a flat surface. SAMPLE COLLECTION
-
Remove the swab from the pouch. Be careful not to touch the swab tip (soft end) with hand.
-
Carefully insert the swab no more than 3/4 inch (1.5 cm) into the nostril. Slowly rotate the swab at least 5 times against the nostril wall.
Do not insert the swab any further if you feel any resistance.- Remove the swab and repeat in the other nostril using the same swab. Check: Did you swab BOTH nostrils?
Note: If you are swabbing others, please wear a face mask. With children, the maximum depth of insertion into the nostril may be less than ½ to ¾ of an inch, and you may require another adult to hold the child’s head while swabbing.
RUNNING THE TEST
- Remove the swab and repeat in the other nostril using the same swab. Check: Did you swab BOTH nostrils?
-
Immerse the swab into the buffer tube and swirl the swab in the buffer. Ensure the sample is mixed thoroughly by making at least 10 circles. Sample must be adequately mixed into the buffer, otherwise, incorrect results may occur.
-
Leave the swab in the buffer tube for 1 minute. A timer is recommended for this step.
-
After 1 minute, pinch the tip of the swab from the outside of the tube to remove any excess liquid from the swab.
Remove and discard the swab. -
Hold the buffer tube upright and screw the large cap back onto the tube. Ensure a tight fit to prevent leaking.
- Twist to open the small cap at the top of the tube.
-
Invert the buffer tube and squeeze 4 drops of test sample into the sample well on the test cassette. Then discard the buffer tube.
Note: Incorrect results may be observed if <4 drops of sample are added. Sample must be applied to the test cassette within one hour of completing step -
Start timer. Read results at 10 minutes.
Do not interpret results before 10 minutes or after 20 minutes. Inaccurate test interpretations may occur.
INTERPRETATION OF RESULTS
Look for lines next to ‘C’(Control), ‘F-A’, ‘F-B’ and ‘CoV’.
- C = Control Line
- F-A = Flu A Test Line
- F-B = Flu B Test Line
- CoV = COVID-19 Test Line
A red line should always appear at the ‘C’ position; this is a control line and singals that the test is working properly.
INVALID RESULT
Check to see if a pink to red line is visible at the control line ‘C’ in the results window. If a line is not visible at “C”, even if any other line is visible in the results window, the result is considered invalid.
If you do not see a C line, DO NOT CONTINUE reading the results. It means your test is invalid. Repeat the test with a new sample and new test kit materials.
NEGATIVE RESULT
- If a control ‘C’ line is visible and you do not see a line at ‘F-A’ , ‘F-B’ or ‘CoV’, it means the test is negative. The Flu A, Flu B or COVID-19 virus have not been detected.
- To increase the chance that the negative result for COVID-19 in accurate, you should test again in 48 hours after the first day of testing.
If respiratory symptoms persist, you seek follow-up care with healthcare provider.
If the control line at “C” is visible and any other line or multiple lines on ‘F-A’, ‘F-B’ and/or ‘CoV’ are visible, the test is positive for that virus.
NOTE: Any pink to red test line, no matter how faint, should be considered a positive result when the control line is also present.
SERIAL TESTING
Repeat Testing is needed for all samples that are negative for SARS-CoV- 2 on
the first day of testing, even if they are positive for influenza A and/or B.
Repeat testing is needed to improve test accuracy for SARS-CoV-2. Please
follow the table below when interpreting test results. Serial (repeat) testing
does not need to be performed if patients have a positive SARS-CoV-2 result on
the first day of testing.
Status on First Day of Testing| Day 0 (First Test)|
Serial Testing?| Day 2 (Second Test)| Interpretation
---|---|---|---|---
With Symptoms| SARS-CoV-2 (+)
Influenza A and B (-)| NO| Not needed| Positive for COVID-19
Presumptive negative for Influenza
SARS-CoV-2 (+)
Influenza A and/or B (+)| NO| Not needed| Positive for COVID-19
Positive for Influenza A and/or B
SARS-CoV-2 (-)
Influenza A and/or B(-)| YES| SARS-CoV-2 (+)
Influenza A and/ or B (-)| Positive for COVID-19
Presumptive Negative for Influenza
SARS-CoV-2 (-)
Influenza A and/or B (+)| YES| SARS-CoV-2 (+)
Influenza A and/ or B (+)| Positive for COVID-19
Positive for Influenza A and/or B
SARS-CoV-2 (-)
Influenza A and/or B (-)| YES| SARS-CoV-2 (-)
Influenza A and/ or B (+)| Presumptive Negative for COVID-19
Positive for Influenza A and/or B
SARS-CoV-2 (-)
Influenza A and/or B (-)| YES| SARS-CoV-2 (-)
Influenza A and/ or B (-)| Presumptive Negative for COVID-19
Presumptive Negative for Influenza
SARS-CoV-2 (-)
Influenza A and/or B (-)| YES| SARS-CoV-2 (+)
Influenza A and/ or B (+)| Positive for COVID-19
Positive for Influenza A and/or B
SARS-CoV-2 (-)
Influenza A and/or B (+)| YES| SARS-CoV-2 (-)
Influenza A and/ or B (-)| Presumptive Negative for COVID-19
Positive for Influenza A and/or B
Status on First Day of Testing| Day 0 (First Test)|
Serial Testing?| Day 2 (Second Test)| Interpretation
---|---|---|---|---
| SARS-CoV-2 (-)
Influenza A and/or B (+)| YES| SARS-CoV-2 (-)
Influenza A and/ or B (+)| Presumptive Negative for COVID-19
Positive for Influenza A and/or B
SARS-CoV-2 (-)
Influenza A and/or B (+)| YES| SARS-CoV-2 (+)
Influenza A and/ or B (+)| Positive for COVID-19
Positive for Influenza A and/or B
UNDERSTANDING YOUR RESULTS
Invalid Result : The test could not tell whether or not you have COVID-19, influenza A (Flu A), or influenza B (Flu B). The test needs to be repeated with a new kit and sample.
Negative Result : The virus from COVID-19, Flu A, and/or Flu B were not detected in the sample. A negative result does not mean it is certain that you do not have COVID-19, Flu A and/or Flu B infection. There is a higher chance of false negative results with antigen tests compared to laboratory-based molecular tests. If you tested negative and continue to experience COVID-19, Flu A and/or Flu B-like symptoms, you should seek follow-up care with your healthcare provider.
All negative results should be treated as presumptive and confirmation with a molecular assay may be necessary if there is a high likelihood of SARS-CoV-2 infection, such as in an individual with a close contact with COVID-19 or with suspected exposure to COVID-19 or in communities with high prevalence of infection. Negative results do not rule out SARS-CoV-2 infection and should not be used as the sole basis for treatment or patient management decisions, including infection control decisions.
Positive Result: The SARS-CoV-2, Flu A and/or Flu B virus(es) were detected in your sample. It is very likely that you have the respective infection(s) and are contagious. Please contact your healthcare provider or your local health authorities and follow local guidelines for self-isolation. There is a small chance that this test can give you a positive result that is incorrect (a false positive).
Positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. Individuals who test positive with the WELLlife™ COVID-19 / Influenza A&B Home Test should self-isolate and seek follow up care with their physician or healthcare provider as additional confirmatory testing with a molecular test for positive results may also be necessary, if there is a low likelihood of COVID-19, such as in individuals without known exposures to COVID-19 or residing in communities with low prevalence of infection.
RESULTS REPORTING
Report your test result(s) at MakeMyTestCount.Org–this voluntary and anonymous
reporting helps public health teams understand COVID-19 spread in your area
and across the country and informs public health decisions.
LIMITATIONS
- The performance of this test was established based on the evaluation of a limited number of clinical specimens collected between December 2023 and March 2024. The clinical performance has not been established for all circulating variants but is anticipated to be reflective of the prevalent variants in circulation at the time and location of the clinical evaluation. Performance at the time of testing may vary depending on the variants circulating, including newly emerging strains of SARS-CoV-2 and their prevalence, which change over time.
- There is a higher chance of false negative results with antigen tests than with laboratory-based molecular tests due to the sensitivity of the test technology. This means that there is a higher chance this test will give a false negative result in an individual with COVID-19 as compared to a molecular test, especially in samples with low viral load.
- All negative results SARS-CoV-2 or influenza are presumptive and confirmation with a molecular assay may be necessary.
- If you continue to have symptoms of COVID-19 or influenza and both your first and second tests are negative, you may not have COVID-19 or influenza, however, you should follow up with a healthcare provider.
- If the test is positive, then proteins from the virus that causes COVID-19 have been found in the sample and you likely have COVID-19 and the individual likely has respiratory infection with COVID-19 or influenza.
- Incorrect test results may occur if a specimen is incorrectly collected or handled.
- This test is read visually and has not been validated for use by those with impaired vision or color-impaired vision.
- Individuals who recently received nasally administered influenza A or influenza B vaccine may have false positive test results after vaccination.
- Based on sequence and epitope analyses, a potential for cross-reactivity between the SARS-CoV-2 test and HKU1 exist. Wet testing with HKU1 coronavirus was not conducted and therefore, cross-reactivity between SARS-CoV-2 and HKU1 coronavirus cannot be ruled out.
PERFORMANCE CHARACTERISTICS
ANALYTICAL PERFORMANCE
Limit of Detection (Analytical Sensitivity)
Limit of detection (LoD) for SARS-CoV-2 and influenza A and B in WELLlife™
COVID-19 / Influenza A&B Home Test was determined by evaluating different
concentrations of UV-inactivated SARS-CoV- and live influenza A and B viruses.
The viruses were diluted in pooled negative swab matrix (PNSM) to generate
virus dilutions for testing. Anterior nasal swab samples were prepared by
adding 50μL of each virus dilution onto the sterile swab. The swab samples
were tested according to the test procedure in package insert. Range-finding
testing was conducted with three replicates at various dilutions and
confirmatory testing was conducted with 20 replicates. The lowest
concentration that generated ≥95% positive detection rate was set as the LoD
concentration.
Virus Strains| Stock Concentration
(TCID 50 /mL)
| LoD (TCID 50 /mL)| LoD (TCID 50 /
Swab)
|
#Positiv e/ #Total
| Percent Detected
(%)
---|---|---|---|---|---
SARS-CoV-2
UV inactivated, USA-WA1/2020
| 3.16×106| 7.90 x102| 39.5| 20/20| 100%
Influenza A A/Victoria/4897/
2022(H1N1)
| 2.02×105| 1.01 x102| 5.05| 20/20| 100%
Influenza A A/Darwin/6/202
1(H3N2)
| 4.17×105| 2.09 x102| 10.45| 20/20| 100%
Influenza B B/Washington/0
2/2019(Victoria)
| 3.16×106| 3.16 x103| 158| 20/20| 100%
Analytical Reactivity
The analytical reactivity of the antibodies targeting Influenza A, influenza
B, and SARS-CoV-2 in WELLlife™ COVID-19 / Influenza A&B Home Test was
evaluated with the currently available strains.
Influenza Virus
(Type/Subtype)
| Virus Strain Name| Analytical Reactivity| Positive/
Replicates
---|---|---|---
SARS-CoV-2(X
BB.1.5)
| hCoV-19/USA/MD-HP40900/
2022
| 7.8E+01 TCID50/mL| 10/10
A(H1N1)pdm09| A/California/04/2009| 2.80E+03 TCID50/mL| 3/3
A/Brisbane/02/18| 1.51E+02 TCID50/mL| 3/3
A/Michigan/45/15| 1.86E+01 TCID50/mL| 3/3
A/Guangdong- Maonan/SWL
1536/19
| 2.09E+02 TCID50/mL| 3/3
A/NY/03/09| 2.29E+04 TCID50/mL| 3/3
A/Indiana/02/2020| 9.70E+06 CEID50/mL| 3/3
A/Wisconsin/588/2019| 7.00E+03 FFU/mL| 3/3
A/Sydney/5/2021| 4.80E+03 TCID50/mL| 3/3
A/Hawaii/66/2019| 1.85E+07 CEID50/mL| 3/3
A/Wisconsin/67/22| 4.21E+02 TCID50/mL| 3/3
A(H3N2)| A/Tasmania/503/2020| 1.30E+05 FFU/mL| 3/3
A/New York/21/2020| 2.60E+05 FFU/mL| 3/3
A/Alaska/01/2021| 3.75E+04 FFU/mL| 3/3
A/Hong Kong/45/2019| 1.50E+04 FFU/mL| 3/3
A/Hong Kong/2671/19| 1.05E+03 TCID50/mL| 3/3
A(H3N2)v| A/Indiana/08/2011| 8.10E+02 TCID50/mL| 3/3
A(H1N1)v| A/Ohio/09/2015| 7.00E+05 CEID50/mL| 3/3
A(H1N2)v| A/Minnesota/19/2011| 4.00E+06 CEID50/mL| 3/3
A(H5N1)| A/mallard/Wisconsin/2576/20 09| 4.00E+05 CEID50/mL| 3/3
A(H7N3)| A/northern pintail/Illinois/10OS3959/201| 7.00E+05 CEID50/mL| 3/3
Influenza Virus
(Type/Subtype)
| Virus Strain Name| Analytical Reactivity| Positive/
Replicates
---|---|---|---
| 0| |
B(Non Victoria and Non
Yamagata)
| B/Maryland/1/59| 3.38E+03 CEID50/mL| 3/3
B(Victoria lineage)| B/Brisbane/60/2008| 1.29E+00 TCID50/mL| 3/3
B/Colorado/06/17| 5.85E+01 TCID50/mL| 3/3
B/Texas/02/2013| 2.45E+01 TCID50/mL| 3/3
B/Michigan/01/2021| 1.43E+04 TCID50/mL| 3/3
B(Yamagata lineage)| Yamagata – B/Texas/06/2011| 7.55E+02 TCID50/mL| 3/3
Yamagata – B/Utah/09/2014| 1.26E+03 TCID50/mL| 3/3
B/Wisconsin/01/2010| 1.78E+02 TCID50/mL| 3/3
Analytical Specificity: Cross-Reactivity and Microbial Interference
Cross-reactivity of the WELLlife™ COVID-19 / Influenza A&B Home Test was
evaluated by testing a panel of related pathogens, high prevalence disease
agents, and normal or pathogenic flora that are reasonably likely to be
encountered in clinical specimens and could potentially cross-react with the
WELLlife™ COVID-19 / Influenza A&B Home Test including twenty (20) bacteria,
twenty (20) viruses and one (1) negative matrix. Each organism and virus were
tested in triplicate the absence (cross-reactivity) or presence (interference)
of co-spiked UV-inactivated SARS-CoV-2, influenza A, and B at 3 x LoD. No
cross-reactivity was observed with the listed microorganisms when tested at
the concentration presented in the table below. No interference was observed
with the listed microorganisms when tested at the concentration presented in
the table below in the presence of the target analytes.
Potential Cross-Reactant | Concentration Tested |
---|---|
SARS-CoV-1 | 1.25X105 PFU/ml |
MERS-coronavirus | 1.47X105 TCID50/mL |
Human coronavirus OC43 | 7.00X105 TCID50/mL |
Potential Cross-Reactant | Concentration Tested |
--- | --- |
Human coronavirus 229E | 1.58X105 TCID50/mL |
Human coronavirus NL63 | 7.05X104 TCID50/mL* |
Adenovirus Type 1 | 2.23X105 TCID50/mL |
Adenovirus Type 7 | 1.58X105 TCID50/mL |
Cytomegalovirus | 7.05X104 TCID50/mL* |
Epstein Barr Virus | 1.83X106 CP/mL |
Human Metapneumovirus | 3.50X105 TCID50/mL |
Parainfluenza virus 1 | 2.00X105 TCID50/mL |
Parainfluenza virus 2 | 1.75X105 TCID50/mL |
Parainfluenza virus 3 | 7.00X105 TCID50/mL |
Parainfluenza virus 4 | 2.39X105 TCID50/mL |
Enterovirus Type 68 | 2.23X105 TCID50/mL |
Respiratory syncytial virus A | 3.50X105 TCID50/mL |
Respiratory syncytial virus B | 2.29X105 TCID50/mL |
Rhinovirus 1A | 7.05X104 TCID50/mL |
Bordetella pertussis | 2.90X108 CFU/mL |
Candida albicans | 1.21X107 CFU/mL |
Chlamydia pneumoniae | 4.33X106 IFU/mL |
Corynebacterium xerosis | 2.30X107 CFU/mL |
Escherichia coli | 1.79X108 CFU/mL |
Hemophilus influenzae | 9.68X106 CFU/mL |
Lactobacillus Acidophilus | 1.21X107 CFU/mL |
Legionella spp pneumophila | 6.50X106 CFU/mL |
Moraxella catarrhalis | 2.50X108 CFU/mL |
Mycoplasma pneumoniae | 2.50X107 CFU/mL |
Mycobacterium tuberculosis avirulent | 3.03X106 CFU/mL |
Neisseria meningitidis | 3.43X106 CFU/mL |
Neisseria sp. Elongata | 2.68X108 CFU/mL |
Pneumocystis jirovecii | 1.30X107 CFU/mL |
Pseudomonas aeruginosa | 3.45X108 CFU/mL |
Potential Cross-Reactant | Concentration Tested |
--- | --- |
Staphylococcus aureus subsp. aureus | 2.60X108 CFU/mL |
Staphylococcus epidermidis | 9.00X107 CFU/mL |
Streptococcus salivarius | 1.01X106 CFU/mL |
Streptococcus pneumoniae | 1.81X107 CFU/mL |
Streptococcus pyogenes | 7.50X107 CFU/mL |
Measles | 8.48X105 TCID50/mL |
Mumps | 8.48X105 TCID50/mL |
Pooled Negative Nasal Wash | NA |
Endogenous Interfering Substances
The potential interference of endogenous substances with the antibodies used
for the detection of SARS-CoV-2, influenza A and B was examined by testing
nineteen (19) substances in a negative clinical matrix in triplicate, in the
absence or presence of each virus at 3 x LOD concentrations for SARS-CoV-2,
influenza A(H1N1), and influenza B(Yamagata). The interference study was
conducted using medically relevant concentrations of the potentially
interfering substances listed below to assess the potential interference of
the substances on the performance of the WELLlife™ COVID-19 / Influenza A&B
Home Test.
At 15% (v/v) and when diluted down to 0.75% (v/v), FluMist Quadrivalent Live
Intranasal Influenza Virus Vaccine yielded false positive results for
Influenza A and Influenza B. At a dilution of 0.375% (v/v), the results were
negative (0/3 positive results). Hand sanitizer containing 80% ethanol yielded
false positive results for SARS-CoV-2 and Influenza B at a dilution of 15%
(v/v) and when diluted down to 3.75% (v/v). At a dilution of 1.875% (v/v), the
results were negative. Two interferents produced false-negative results for
Influenza B: hand sanitizer cream lotion (15% v/v) and hand soap liquid gel
(10% w/v). All Influenza B results were positive when tested with 7.5% (v/v)
hand sanitizer cream lotion and 0.05% (w/v) hand soap liquid gel.
No interference was observed with the other listed substances when tested at
the concentration presented in the table below in the presence or absence of
the target analytes.
Results of Endogenous Interfering Substances
Potential Interferent| Concentration| Cross-reactivity
(no analyte)
(# pos reps / total reps)
| Interference
(3x co-spike analyte LoD)
(# pos reps / total reps)
---|---|---|---
SARS-
CoV- 2
| Flu
A
| Flu
B
| SARS-
CoV- 2
| Flu
A
| Flu
B
Human Whole Blood
(EDTA tube)
| 4% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Mucin| 0.50%| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Chloraseptic
(Menthol/Benzocaine)
| 1.5 mg/mL| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Naso GEL (NeilMed)| 5% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Nasal Drops
(Phenylephrine)
| 15% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Nasal Spray
(Oxymetazoline)
| 15% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Nasal Spray (Cromolyn)| 15% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Zicam| 5% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Homeopathic (Alkalol)| 10% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Sore Throat Phenol Spray| 15% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Tobramycin| 4 μg/mL| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Mupirocin| 10 mg/mL| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Fluticasone Propionate| 5% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Tamiflu (Oseltamivir
Phosphate)
| 5 mg/mL| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
FluMist/ FluMist Quadrivalent Live intranasal
influenza virus
vaccine
| 15% v/v| 0/3| 3/3| 3/3| 3/3| 3/3| 3/3
0.375% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Zanamivir| 282 ng/mL| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Biotin| 3,500 ng/mL| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Body & Hand Lotion| 0.5% w/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Body Lotion, with 1.2%
dimethicone
| 0.5% w/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Hand Lotion| 5% w/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Hand Sanitizer with Aloe,
62% ethyl alcohol
| 5% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 3/3
Hand Sanitizer cream
lotion
| 15% v/v| 0/3| 0/3| 0/3| 3/3| 3/3| 0/3
7.5% v/v| –| –| –| 3/3| 3/3| 3/3
Hand Sanitizer, 80%
ethanol, fast drying
| 15% v/v| 3/3| 0/3| 2/3| 3/3| 3/3| 3/3
1.875% v/v| 0/3| 0/3| 0/3| –| –| –
Hand soap liquid gel| 10% w/v| 0/3| 0/3| 0/3| 3/3| 3/3| 0/3
0.5% w/v| –| –| –| 3/3| 3/3| 3/3
High Dose Hook Effect
A high-dose hook effect was not observed in WELLlife™ COVID-19 / Influenza A&B
Home Test for the SARS-CoV-2, influenza A and B viral strains at the
concentration listed below.
Virus Type | Virus Strain | Concentration Tested |
---|---|---|
SARS-CoV-2 | UV inactivated, USA-WA1/2020 | 3.16×106 TCID50/mL |
Influenza A (H1N1) | A/Victoria/4897/2022 | 2.02×105 TCID50/mL |
Influenza A(H3N2) | A/Darwin/6/2021 | 4.17×105 TCID50/mL |
Influenza B (Victoria lineage) | B/Washington/02/2019 | 3.16×106 TCID50/mL |
Influenza B (Yamagata
lineage)
| B/Florida/4/2006| 1.17×105 TCID50/mL
Competitive Interference
For co-infection, SARS-CoV-2 at levels near LoD was tested in the presence of
high levels of influenza A or influenza B and influenza A and influenza B at
levels near LOD were tested in the presence of high levels of SARS-CoV-2. No
competitive interference was seen between high levels of SARS-CoV-2 and low
levels of Influenza A and B and between high levels of Influenza A and low
levels of SARS-CoV-2 and influenza A in this testing at the concentration
listed in the tables below. Competitive inhibition were observed between high
levels of influenza B(Yamagata Lineage) and low levels of Influenza A in this
testing at the concentration listed in the tables below.
SARS-CoV-2 &Influenza A &Influenza B Virus(Yamagata Lineage)
SARS-CoV-2 USA-WA1/2020| Influenza A Virus(H1N1pdm09)
A/Victoria/4897/2022
| Influenza B Virus(Yamagata Lineage)
B/Florida/4/2006
---|---|---
Concentration
(TCID 50 /mL)
| Percent
Agreement
| Concentration
(TCID 50 /mL)
| Percent
Agreement
| Concentration
(TCID 50 /mL)
| Percent
Agreement
Negative| 100%| 6.73×104| 100%| 1.76 x102| 100%
2.37×103| 100%| 6.73×104| 100%| Negative| 100%
2.37×103| 100%| 6.73×104| 100%| 1.76 x102| 100%
Negative| 100%| 3.03×102| 0| 3.90 x104| 100%
Negative| 100%| 3.03×102| 0| 1.95 x104| 100%
SARS-CoV-2 USA-WA1/2020| Influenza A
Virus(H1N1pdm09) A/Victoria/4897/2022
| Influenza B Virus(Yamagata Lineage)
B/Florida/4/2006
---|---|---
Concentration
(TCID 50 /mL)
| Percent
Agreement
| Concentration
(TCID 50 /mL)
| Percent
Agreement
| Concentration
(TCID 50 /mL)
| Percent
Agreement
Negative| 100%| 3.03×102| 100%| 7.80 x103| 100%
Negative| 100%| 3.03×102| 100%| 3.90 x103| 100%
2.37×103| 100%| Negative| 100%| 3.90 x104| 100%
2.37×103| 100%| 3.03×102| 0| 3.90 x104| 100%
2.37×103| 100%| 3.03×102| 0| 1.95E+04| 100%
2.37×103| 100%| 3.03×102| 100%| 7.80 x103| 100%
2.37×103| 100%| 3.03×102| 100%| 3.90 x103| 100%
1.05×106| 100%| 3.03×102| 100%| Negative| 100%
1.05×106| 100%| Negative| 100%| 1.76 x102| 100%
1.05×106| 100%| 3.03×102| 100%| 1.76 x102| 100%
SARS-CoV-2 &Influenza A &Influenza B Virus(Victoria Lineage)
SARS-CoV-2 USA-WA1/2020| Influenza A Virus(H1N1pdm09)
A/Victoria/4897/2022
| Influenza B Virus(Victoria Lineage)
B/Washington/02/19
---|---|---
Concentration
(TCID 50 /mL)
| Percent
Agreement
| Concentration
(TCID 50 /mL)
| Percent
Agreement
| Concentration
(TCID 50 /mL)
| Percent
Agreement
Negative| 100%| 6.73 x104| 100%| 3.51 x102| 100%
2.37 x103| 100%| 6.73 x104| 100%| Negative| 100%
2.37 x103| 100%| 6.73 x104| 100%| 3.51 x102| 100%
Negative| 100%| 3.03 x102| 100%| 1.05 x106| 100%
2.37 x103| 100%| Negative| 100%| 1.05 x106| 100%
2.37 x103| 100%| 3.03 x102| 100%| 1.05 x106| 100%
1.05 x106| 100%| 3.03 x102| 100%| Negative| 100%
1.05 x106| 100%| Negative| 100%| 3.51 x102| 100%
1.05 x106| 100%| 3.03 x102| 100%| 3.51 x102| 100%
CLINICAL PERFORMANCE
A prospective study was performed in which seven hundred eighty-seven (787)
direct anterior nasal swab specimens were sequentially enrolled (between
December 2023 and March 2024) and tested fresh. The samples were collected
from symptomatic patients suspected of infection with respiratory symptoms, at
nine (9) clinical sites. Subjects performed testing on self-collected swab
samples in age groups 14 and older, and adult collected samples for age groups
2-13, in a simulated at-home environment. To be enrolled in the study,
patients had to present at the participating study site within five (5) days
of symptom onset with signs and symptoms of respiratory infection generally
observed from SARS-CoV-2, influenza A and/or influenza B, during the study
period. Two anterior nasal swab specimens were collected from each patient:
one swab was collected by a healthcare professional and sent for testing using
a FDA-cleared molecular comparator method, and the other swab was self-
collected and tested immediately with the WELLlife™ COVID-19 / Influenza A&B
Home Test per the test procedure. Out of 787 enrolled subjects, there were 769
evaluable subjects.
SUBJECTS DEMOGRAPHICS Subjects Demographics.
Characteristic| Lay-user/ Tester Collection
N=174
| Self-Collecting N=595| Overall N=769
---|---|---|---
Age
Mean (SD)| 8.7 (4.5)| 46.3 (18.8)| 37.8 (22.9)
Median[Min, Max]| 8[2, 32]| 43[14, 94]| 35[2, 94]
Age Group
<14| 149 (85.6%)| 0| 149 (19.4%)
14-24| 24 (13.8%)| 85 (14.3%)| 109 (14.2%)
25-64| 1 (0.6%)| 380 (63.9%)| 381 (49.5%)
64| 0| 130 (21.8%)| 130 (16.9%)
Six at Birth
Female| 98 (56.3%)| 370 (62.2%)| 468 (60.9%)
Male| 76 (43.7%)| 225 (37.8%)| 301 (39.1%)
Ethnicity
Hispanic/Latino| 91 (52.3%)| 250 (42.0%)| 341 (44.3%)
Not Hispanic/Latino| 83 (47.7%)| 330 (55.5%)| 413 (53.7%)
Unknown/Prefer not to answer| 0| 15 (2.5%)| 15 (2.0%)
Race
Characteristic| Lay-user/ Tester Collection
N=174
| Self-Collecting N=595| Overall N=769
---|---|---|---
American Indian or Alaskan Native| 1 (0.6%)| 2 (0.3%)| 3 (0.4%)
Asian| 19 (10.9%)| 162 (27.2%)| 181 (23.5%)
Black or African American| 44 (25.3%)| 95 (16.0%)| 139 (18.1%)
White| 106 (60.9%)| 294 (49.4%)| 400 (52.0%)
Native Hawaiian or Other Pacific Islander| 0| 0| 0
More than one race| 1 (0.6%)| 6 (1.0%)| 7 (0.9%)
Unknown/Prefer not to answer| 2 (1.1%)| 30 (5.0%)| 32 (4.2%)
Other| 1 (0.6%)| 6 (1.0%)| 7 (0.9%)
SARS-COV-2 PERFORMANCE
WELLlife™ COVID-19 / Influenza A&B Home Test performance compared to reference
PCR: SARS-CoV-2
SARS-CoV- 2|
Comparator Positives
| Comparator
Negatives
|
Total
---|---|---|---
Candidate Positives| 112| 2| 114
Candidate Negatives| 16| 639| 655
Total| 128| 641| 769
Positive Percent Agreement (PPA) = 87.5% (112/128) 95% CI: 80.7 –
92.2%
Negative Percent Agreement (NPA) = 99.7% (639/641) 95% CI: 98.9 –
99.9%
SARS-CoV-2 Clinical Performance in Subjects on Days Post Symptoms Onset
Days of COVID-19
Symptoms
| Number of Subject samples tested| WELLlife™ COVID-19 /
Influenza
A &B Test Positives
| Comparator Positives| % Positive Rate (by Comparator)| PPA
(95%CI)
---|---|---|---|---|---
Day 0| 39| 3| 5| 12.80%| 60.0%
(23.1%, 88.2%)
Day 1| 168| 26| 28| 16.70%| 92.9%
(77.4%, 98.0%)
Day 2| 236| 30| 36| 15.25%| 83.3%
(68.1%, 92.1%)
Day 3| 156| 27| 27| 17.30%| 96.3%
Days of COVID-19
Symptoms
| Number of Subject samples tested| WELLlife™ COVID-19 /
Influenza A &B Test
Positives
| Comparator Positives| % Positive Rate (by Comparator)| PPA
(95%CI)
---|---|---|---|---|---
| | | | | (81.7%, 99.3%)
Day 4| 106| 16| 19| 17.90%| 84.2%
(62.4%, 94.5%)
Day 5| 64| 12| 13| 20.30%| 84.6%
(57.8%, 95.7%)
Total| 769| 114| 128| 16.64%| 87.5%
(80.7%, 92.2%)
INFLUENZA A PERFORMANCE
WELLlife™ COVID-19 / Influenza A&B Home Test performance compared to reference
PCR: Influenza A
Influenza A| Comparator
Positives
| Comparator
Negatives
| ****
Total
---|---|---|---
Candidate Positives| 79| 2| 81
Candidate Negatives| 13| 675| 688
Total| 92| 677| 769
Positive Percent Agreement (PPA) = 85.9% (79/92) 95% CI:
77.3 – 91.6%
Negative Percent Agreement (NPA) = 99.7% (675/677) 95% CI:
98.9 – 99.9%
INFLUENZA B PERFORMANCE
WELLlife™ COVID-19 / Influenza A&B Home Test performance compared to reference
PCR: Influenza B
Influenza B| Comparator
Positives
| Comparator
Negatives
|
Total
---|---|---|---
Candidate Positives| 33| 2| 35
Candidate Negatives| 5| 729| 734
Total| 38| 731| 769
Positive Percent Agreement (PPA) = 86.8% (33/38) 95% CI:
72.7 – 94.2%
Negative Percent Agreement (NPA) = 99.7% (729/731) 95% CI:
99.0 – 99.9%
SERIAL TESTING
- A prospective clinical study was conducted between January 2021 and May 2022 as a component of the Rapid Acceleration of Diagnostics (RADx) initiative from the National Institutes of Health (NIH). A total of 7,361 individuals were enrolled via a decentralized clinical study design, with a broad geographical representation of the United States. Per inclusion criteria, all individuals were asymptomatic upon enrollment in the study and at least 14 days prior to it and did not have a SARS-CoV-2 infection in the three months prior to enrollment. Participants were assigned to one of three EUA authorized SARS-CoV-2 OTC rapid antigen tests to conduct serial testing (every 48 hours) for 15 days. If an antigen test was positive, the serial-antigen testing result is considered positive.
- At each rapid antigen testing time point, study subjects also collected a nasal swab for comparator testing using a home collection kit (using a 15-minute normalization window between swabs). SARS-CoV-2 infection status was determined by a composite comparator method on the day of the first antigen test, using at least two highly sensitive EUA RT-PCRs. If results of the first two molecular tests were discordant, a third highly sensitive EUA RT-PCR test was performed, and the final test result was based upon the majority rule. Study participants reported symptom status throughout the study using the MyDataHelps app. Two-day serial antigen testing is defined as performing two antigen tests 36 – 48 hours apart. Three-day serial antigen testing is defined as performing three antigen tests over five days with at least 48 hours between each test.
- Out of the 7,361 participants enrolled in the study, 5,609 were eligible for analysis. Among eligible participants, 154 tested positive for SARS-CoV-2 infection based on RTPCR, of which 97 (62%) were asymptomatic on the first day of their infection, whereas 57 (39%) reported symptoms on the first day of infection.
- Performance of the antigen test with serial testing in symptomatic individuals is described in the table below. Data establishing PPA of COVID-19 antigen serial testing compared to the molecular comparator single day testing throughout the course of infection with serial testing. Data is from all antigen tests in study combined.
Data establishing PPA of COVID-19 antigen serial testing compared to the molecular comparator single day testing throughout the course of infection with serial testing. Data is from all antigen tests in study combined.
DAYS AFTER FIRST PCR POSITIVE TEST RESULT| SYMPTOMATIC ON FIRST DAY OF
TESTING
---|---
Ag Positive / PCR Positive (Antigen Test Performance %
PPA)
1Test| 2 Test| 3Test
0| 34/57(59.6%)| 47/51(92.2%)| 44/47(93.6%)
2| 58/62(93.5%)| 59/60(98.3%)| 43/43(100.0%)
4| 55/58(94.8%)| 53/54(98.1%)| 39/40(97.5%)
6| 27/34(79.4%)| 26/33(78.8%)| 22/27(81.5%)
8| 12/17(70.6%)| 12/17(70.6%)| 7/11(63.6%)
10| 4/9(44.4%)| 3/7(42.9%)| NA
- Test = one (1) test performance on the noted days after first PCR positive test result. Day 0 is the first day of documented infection with SARS-CoV-2.
- Tests = two (2) tests performance an average of 48 hours apart. The first test performed on the indicated day and the second test performed 48 hours later.
- Tests = three (3) tests performance an average of 48 hours apart. The first test performed on the indicated day, the second test performed 48 hours later, and a final test performed 48 hours after the second test.
ASSISTANCE
If the test does not perform as expected, please email at
wondfo@wondfousa.com or call
1-888-444-3657.
REFERENCES
- US Department of Health and Human Services. National Institutes of Health. Influenza [Fact Sheet]. January 2011.
- Montalto N, Byrd R. An Office-Based Approach to Influenza: Clinical Diagnosis and Laboratory Testing. American Family Physician. January 2003; 67:11-118.
- Cui J, Li F, Shi ZL. Origin and evolution of pathogenic coronaviruses. Nat Rev Microbiol 2019; 17: 181-192.
- Lauer S, et al. The Incubation Period of Coronavirus Disease 2019 (COVID-19) From Publicly Reported Confirmed Cases: Estimation and Application. Ann Intern Med 2020 May 5;172(9):577-582.
INDEX OF SYMBOLS
Wondfo USA Co., Ltd. 6720 Cobra Way, San Diego, CA 92121
-
Manufacturing site
Guangzhou Wondfo Biotech Co., Ltd.
No. 8 Lizhishan Road, Science City Huangpu District, 510663 Guangzhou, P.R. China
Made in China -
Distributed by Wondfo USA Co., Ltd. 545 Willowbrook Centre Parkway, Unit B Willowbrook, IL 60527
+1 630-468-2199
www.wondfousa.com -
Rev.A2 Rel.: 2024/05/21