Intretech NCTG3 Gen 3 NC Degree Thermometer Instruction Manual
- June 3, 2024
- Intretech
Table of Contents
- NCTG3 Gen 3 NC Degree Thermometer
- Introduction
- Package Contents
- Indications for Use
- Warnings And Cautions
- Precautions
- Operating Your Thermometer
- Troubleshooting / Error Codes
- Maintaining Your Thermometer
- Product Specifications
- FCC Statement
- Warranty Information
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
NCTG3 Gen 3 NC Degree Thermometer
Instruction Manual
NCTG3 Gen 3 NC Degree Thermometer
NC° THERMOMETER
Instructions For Use
NC° Thermometer (Gen3)
Model number NCTG3
Contents Introduction Warnings And Cautions Operating Your Thermometer
Maintenance
Introduction
Please read and save these instructions Thank you for purchasing the NC°
Thermometer (Gen3), please read these instructions carefully to ensure proper
use and safe operation of your thermometer. If you have any questions or
comments about the Thermometer, contact Customer Support at
help@naturalcycles.com.
Thermometer key parts
Package Contents
- Connected thermometer
- Instructions for Use
- Founder welcome card
The Thermometer is intended to be used on its own or with iOS/Android smartphones using a dedicated App.
Indications for Use
The NC° Thermometer (Gen3) is used orally for the intermittent measurement and monitoring of human body temperature. The device can be used by both adults and children.
Warnings And Cautions
The thermometer may be used by itself or alongside the Natural Cycles app. See Natural Cycles application IFU for information regarding use of the app.
- This product is not intended to diagnose or treat any health problem or disease and should not be used as a substitute for the consultation and advice of a physician or other medical processional
- Thermometers should only be used by children under the supervision of an adult · This thermometer should be held during use; the user should be seated or lying down during measurements.
- Keep the device away from sources of high levels of power line magnetic fields to reduce the likelihood of interference.
- The device is suitable for use in all environments listed in these instructions for use, including domestic environments.
- The use of the device may be limited in the presence of electromagnetic disturbances. This could result in issues such as error messages or the failure of the display/device.
- Don’t use near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
- Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 12 in (30 cm) to any part of the equipment, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Precautions
- This thermometer uses Bluetooth Low Energy 5.0 and Near Field Communication (NFC) to interact with a dedicated iOS/Android smartphone App.
- Clean the thermometer probe before and after use. Thermometer is water resistant, not waterproof. Never dip the thermometer into water or other liquids. Do not boil the probe. For cleaning and disinfecting, please see the Care and Cleaning section.
- Keep out of reach of young children and pets. Do not bend or bite the probe as this may damage the probe.
- Do not use the thermometer if there are signs of damage or after a drop/shock that might have caused damage. If damaged, do not attempt to repair. Please contact Customer Support at help.naturalcycles.com.
- Do not store in direct sunlight or at high temperatures- see Product Specifications section.
- Avoid exercise and drinking hot or cold beverages before taking a temperature as these activities may affect the accuracy of the measurement. If the thermometer has been stored in cold and hot conditions, it should be left for 30 minutes to stabilize to room temperature before attempting the measurement.
- The use of heat and cold producing devices, such as electric heating blankets, heating pads or ice packs, may impair the performance of device and increase the risk of injury to the patient.
About Temperature Measurements
Normal oral temperature readings vary throughout the day. Level of activity,
time of day, beverages, smoking and other factors may also affect body
temperature. This device is calibrated at the time of manufacturing. No
recalibration is required. Contact Customer Support at help.naturalcycles.com/
for any accuracy issues.
Operating Your Thermometer
The Thermometer can either be used by itself or the user can sync the
temperature temperature data to the Natural Cycles app on their smartphone.
Before using the thermometer for the first time, pull the plastic tab sticking
out of the battery hatch to activate the battery.
To Set Up Your NC° Thermometer (Gen3) With A New Mobile Device
- Press the settings button on the thermometer. Pressing the settings button will make the thermometer look for the app to establish connection for 60 seconds, if the app is not found, the thermometer shuts itself off.
- Open iOS/Android Natural Cycles smartphone app for additional pairing instructions if using thermometer with smartphone. When the thermometer finds the smartphone app, a connection is established.
- Once the thermometer connects with the phone/device, you should see the setup complete screen.
Fahrenheit / Celsius Switchable Feature
The thermometer takes measurements in both Fahrenheit and Celsius. When the
thermometer is off, press and hold the settings button for 5 or more seconds
to switch the measurements from Fahrenheit to Celsius or Celsius to
Fahrenheit. The thermometer screen will confirm the switch to Fahrenheit or
Celsius. You can change thermometer settings in the app as well.
History Mode
The thermometer includes a recall of the past ten temperatures taken. Press
the history button to see the last temperature. Continue pressing the history
button to see previous temperatures one at a time up to ten temperatures.
Oral Use
-
Press the power button and then place the thermometer in your mouth to begin measuring.
-
Place the probe under the tongue as near as possible to a heat pocket at the back of the mouth, as noted in the diagram below.
-
Hold the thermometer in place during the measurement; do not bite down on the thermometer. The mouth must remain closed to ensure accurate results. The reading should take approximately 20 seconds.
-
When the measurement is complete the thermometer will provide feedback. The feedback can be light and/or sound based on user settings. The final temperature value will be displayed on the thermometer screen. 5. After a measurement is completed, simply press the power button to turn the thermometer off or the thermometer will shut off automatically.
Troubleshooting / Error Codes
Performance may be affected should one or more of the following occur:
- Operation outside stated temperature and humidity range
- Storage outside the stated temperature and humidity range
- Mechanical shock
- User temperature is below ambient temperature
Error | Problem/Cause | Solution |
---|---|---|
Temperature too low, try again | Starting temperature value is below60.8°F | |
(16°C) or final temperature value is too low, below 89.6°F (32.0°C) | Try |
measuring the temperature again. Make sure that your environment is in the
operating
range and that the thermometer is placed as described in the diagram in the
‘Oral Use’ section.
Temperature too high, try again| Starting temperature value is above 104°F
(40°C) 40 or final temperaturevalue is too high, above 109.04°F
(42.0°C)| Try measuring the temperature again. Make sure that your environment
is in the operating range and that the thermometer is placed as described in
the diagram in the ‘Oral Use’ section.
No measurement detected| No placement of thermometer in mouth detected| Try
measuring the temperature again. Make sure that the thermometer is placed as
described in the diagram in the ‘Oral Use’ section.
Battery low| Battery is depleted| Change the battery as described in the
section ‘Replacing The Battery’.
Critically low battery life| Battery is depleted| Change the battery as
described in the section ‘Replacing The Battery’.
Measuring error| User does not take a stable measurement (moving thermometer
around mouth, taking temperature in and out of mouth)
| Wait 10 seconds before trying again. The thermometer screen will display a count-down from 10. When the thermometer reaches 0 the thermometer will return to screen “Start measuring”
Syncing temperatures with NC App
You can sync the temperature to your phone right after you measure or later by
following the instructions in the app. There are two ways to sync temperatures
with the NC app.
- After a successful measurement is completed, the thermometer will advertise the value via NFC or Bluetooth.*
- When the thermometer is in History mode it also advertises unsynced values via NFC or Bluetooth.
*NFC/Bluetooth depends on user settings which can be changed in the NC app.
Syncing temperatures troubleshooting If you have issues syncing your
temperature from the thermometer into the app, please visit
https://help.naturalcycles.com/ or scan the QR code below and find the
instructions for different types of devices.
https://help.naturalcycles.com/
Maintaining Your Thermometer
Care And Cleaning Instructions
Store the unit in the provided protective case when not in use. Do not store
the thermometer where it will be exposed to direct sunlight, dust or humidity.
Avoid extreme temperatures. Do not attempt to disassemble the thermometer.
Clean the thermometer’s tip by wiping with rubbing alcohol (70% isopropyl
alcohol) or mild soap and water. Do not submerge in any liquid or autoclave.
Wipe the thermometer with a dry cloth and allow it to air dry.
Replacing The Battery
If the thermometer battery is low, battery low’ or
critically low battery
life’ messages will appear on the thermometer screen. Use a tool to open the
hatch in the back of the thermometer to remove the existing battery, and
replace it with a new battery.
Product Specifications
Battery | CR2032 |
---|---|
Battery voltage | 3V |
Battery Life | 6 Months |
Automatic Shut-Off | 1 minute |
Measurement Range | 32-42℃ (89.6-107.6°F) |
Measurement Time | 20 Seconds Nominally |
Accuracy | ±0.3℃ (Temperature less than 35.8℃) |
±0.2℃ (Temperature 35.8℃ to less than 37℃)
±0.1℃ (Temperature 37.0℃ to less than 39.0℃)
±0.2℃ (Temperature 39.0℃ to less than 41.0℃)
±0.3℃ (Temperature greater than 41.0℃)
Operating Temperature| 15-40℃ (59-104°F)
Storage Temperature| -25-50℃ (-13-122°F)
Humidity| ≤ 95%R.H.
Atmospheric Pressure| 70 kPa to 106 kPa
Display| OLED Screen
Display Resolution| 0.01℃ or 0.02°F
Dimensions| 136.8mm x 11.0mm x 29.0mm
Weight| 25g
Memory Function| 10 logs
Service Life| 5 Years
RF Interface| Bluetooth LE (2400 – 2483.5 MHz): 2402 – 2480 MHz; 0 dBm
NFC (13.56 MHz): 13.553~13.567MHz
FCC Statement
IMPORTANT INFORMATION REQUIRED BY THE FCC
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions: this device may not cause harmful interference,
and this device must accept any interference received, including interference
that may cause undesired operation.
The equipment is suitable for use in domestic establishments and is tested to
CISPR emissions Class B Group 1 as well as home healthcare immunity criteria
found in IEC 60601-1-2 Table 4, 6, and Table 9. During the IEC 60601-1-2
immunity tests performed the thermometer will accurately measure temperature
or display an error. NC° Thermometer (Gen3) utilizes Bluetooth Low Energy
(BLE), which uses the 2.4 GHz ISM band, BLE operates between 2.402 and 2.480
GHz, and Near Field Communication (NFC), which uses the 13.56MHz ISM band. NFC
operates between 13.553 and 13.567MHz. NC° Thermometer (Gen3) transmits less
than-2 dBm effective radiated power. To protect the environment, dispose of
empty batteries at appropriate collection sites according to national or local
regulations. FCC WARNING: Changes or modifications not expressly approved by
Intretech could void the user’s authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures:
— Reorient or relocate the receiving antenna.
— Increase the separation between the equipment and receiver.
— Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
— Consult the dealer or an experienced radio/TV technician for help
FCC Radiation Exposure Statement
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. This transmitter must not be co-located or operating
in conjunction with any other antenna or transmitter.
Guidance and Manufacturer’s Declaration — Electromagnetic Emissions
Emissions test| Compliance
RF emissions CISPR 11| Group 1
RF emissions CISPR 11| Class B
Harmonic emissions IEC 61000-3-2| Not applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3| Not applicable
Guidance and Manufacturer’s Declaration — Electromagnetic Immunity
Immunity Test| IEC 60601-1-2 Test level| Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2| ±8 kV contact
±2 kV, ±41(V, ±8 kV, ±I5 k\ air| -8 kV contact
-2 kV, ±4kV, ±8 kV, ±15 kV air
Electrical fast transient/burst IEC 610004-4| Not applicable| Not applicable
Surge IEC 61000-4-5| Not applicable| Not applicable
Voltage dips, short interruptions and voltagL variations on power supply input
lines IEC 61000-4-11| Not applicable| Not applicable
Power frequency magne. field IEC 61000-4-8| ‘,l) A M 50Hz/60Hz| 30 Alm
50Hz/60Hz
Conduced RF IEC 6100, 4-6| Not applicable| Not applicable
Radiated RF IEC 61000- 4-3| 10 V/m
80 MHz — 2,7 GHz
80 % AM at 1 kHz| 10 V/m
80 MHz — 2,7 GHz
80 % AM at 1 kHz
Guidance and Manufacturer’s Declaration – IMMUNITY to proximity fields from RF
wireless communications equipment
Immunity test| IEC60601 test level| Compliance level
Test frequency| Modulation| Maximu m power| Immunity level
Radiated RF
IEC 610004-
3| 385 MHz| “Pulse Modulation:
18Hz| 1.8W| 27 V/m| 27 V/m
450 MHz| ‘FM +5Hz deviation: I kHz sine| 2 W| 28 V/m| 28 V/m
710 MHz
745 MHz
780 MHz| “Pulse Modulation:
217Hz| 0.2 W| 9 V/m| 9 V/m
810 MHz
870 MHz
930 MHz| “Pulse Modulation:
18Hz| 2 W| 28 V/m| 28 V/m
1720 MHz
1845 MHz
1970 MHz| “Pulse Modulation:
2I7Hz| 2 W| 28 V/m| 28 V/m
2450 MHz| “Pulse Modulation:
217Hz| 2 W| 28 V/m| 28 V/m
5240 MHz
5500 MHz
5785 MHz| “Pulse Modulation:
217Hz| 0.2 W| 9 V/m| 9 V/m
Note’ – As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may
be used because while it does not represent actual modulation, it would be
worst case.
Note” – The carrier shall be modulated using a 50 % duty cycle square wave
signal.
Statement:
NC° Thermometer (Gen3) Conforms to ASTM E1112
Warranty Information
The manufacturer provides the warranty in accordance with the legislation of
the customer’s own country of residence, with a minimum of 1 year (Germany: 2
years), starting from the date on which the appliance is sold to the end user.
The warranty only covers defects in material or workmanship.
The repairs under warranty may only be carried out by an authorized service
center. When making a claim under the warranty, the original bill of purchase
(with purchase date) must be submitted. The warranty will not apply in cases
of:
– Normal wear and tear – Incorrect use, e.g. overloading of the appliance, or
use of non-approved accessories
– Use of force, damage caused by external influences
– Damage caused by non-observance of the user manual, e.g. connection to an
unsuitable mains supply or non-compliance with the installation instructions
– Partially or completely dismantled appliances
Proposition 65 Warning
WARNING: This product may contain chemicals known to the State of California
to cause birth defects or other reproductive harm. For more information on
Proposition 65 chemicals, please visit: https://www.p65warnings.ca.gov/
Explanation of Symbols
| Technical equipment and batteries do not belong in household waste. They
must be
disposed of at appropriate collection and disposal points.
---|---
****| Consult the instructions for use
| Dust tight, Protected against splashing water
| Degree of protection against electric shock TYPE BF
| Manufacturer
| Reference Number
| Lot Number
|
Name of the Manufacturer:
Xiamen Intretech Inc.
Address of the Manufacturer:
No. 100, Dongfu West Road, Haicang District, Xiamen, Fujian, China
Contact Information:
http://www.intretech.com/
86-(0)592-5797666
Device Name: NC° Thermometer (Gen3)
Date of publication: October 2022
Document version: US v1.0